28 May 2021 Objective-lens-free confocal endomicroscope using Lissajous scanning lensed-fiber
Author Affiliations +
Abstract

We report an objective-lens-free endomicroscopic catheter for compact Lissajous scanned confocal endomicroscopic system using a scanning lensed-fiber. The fiber scanner comprises a single-mode fiber (SMF) spliced with a coreless silica fiber segment and quadrupole piezoelectric tube. An objective lens was directly formed at the distal end of an SMF using a fusion splicer. The lensed fiber eliminates the need for conventional beam focusing elements such as a gradient index rod lens and precise optical alignment, allowing low cost and facile fabrication of an exceptionally short endomicroscopic catheter. The endomicroscopic scanner was fully packaged within a stainless tube of 2.6 mm in outer diameter and 20 mm in length, which can be easily inserted through the working channel of conventional laparoscope or colonoscope. The microscopic images of a stained kidney section and ear of a mouse were successfully obtained with the Lissajous scanning confocal endomicroscope. The compact Lissajous scanning lensed fiber can provide a new route for diverse in vivo endomicroscopic applications.

1.

Introduction

Endomicroscopes are actively combined with advanced imaging techniques such as multiphoton microscopy,14 optical coherence tomography (OCT),57 and confocal microscopy,810 and they allow minimally invasive endoscopic diagnosis and in vivo imaging. Despite the great potential, the integration of micro-optical modules inside an endomicroscope is still challenging due to their bulk size and lack of conventional optical components. A fiber-optic scanning probe plays an important role for developing compact endomicroscopic imaging probes due to many technical benefits such as compactness, flexibility, lightweight, cost effectiveness, and biocompatibility. In particular, compact fiber-optic probes facilitate the direction insertion through the working channel of an endoscope for on-demand detection and real-time diagnosis.

In recent years, assorted micro-optic probes have been extensively incorporated for endomicroscopic imaging applications. Conventional endomicroscopic probes consist of microscanner, optical fiber, gradient index (GRIN) lens, and additional optical elements to scan a laser beam into the region of interest. For instance, microelectromechanical system (MEMS) lens scanning probe,6 scanning MEMS mirror-based optical probe,1113 or a tubular piezoelectric-based scanning optical fiber1418 allow the compact integration of endomicroscope for clinical applications. In addition, fiber optic probes have been even successfully integrated within a single medical needle.19 However, conventional probes still require additional optical elements for beam focusing or beam deflection such as prism, mirror, or objective lens. The optical elements require precise optical alignments and large packaging size. A GRIN rod lens often serves as an objective lens of the probe to minimize the catheter size as well as to increase the image resolution; however, it exhibits some disadvantages such as long rigid length and high manufacturing cost. Moreover, the GRIN lens restricts the numerical aperture (NA) and the field-of-view of an endomicroscopic probe. As a result, a compact and alignment-free packaging of an imaging probe is still in need for clinical endomicroscopic applications.

A scanning lensed fiber can provide a simple solution for compact packaging of alignment-free laser scanning applications.2022 Lissajous scanning or biaxial resonant scanning has been employed for fiber-scanning endomicroscopes to improve the mechanical stability and the scanning uniformity.23,24 Recently, the selection rule for scanning frequencies realizes high-definition and high frame rate (HDHF) for Lissajous scanned imaging.25 However, compact optical packaging is still challenging in the integration of scanning fiber-based endomicroscopic probes, which can be physically applicable for conventional clinical endoscopes.

Here, we report an objective-free optical fiber scanner for confocal endomicroscope using Lissajous scanning lensed fiber. Figure 1(a) shows the schematic illustration of confocal endomicroscopic probe, which comprises a quadrupole piezoelectric (PZT) fiber scanner and a single-mode fiber (SMF) serially concatenated with a coreless silica fiber (CSF) segment. The lens was made by melting at the end of CSF using a fusion splicer, which focused incident light source to the target. The length of the CSF and the radius of curvature (ROC) of the lens determined the imaging characteristics such as working distance and beam spot size. The ROC of the lensed fiber was controlled by filament power and duration time. This all-fiber scanner allows no need for additional optical elements such as GRIN lens for focusing light and precise alignment. The Lissajous lensed fiber scanner was fully packaged with 2.6 mm diameter as an endomicroscopic probe and coupled with a confocal imaging system.

Fig. 1

(a) Illustration of the Lissajous scanning lensed fiber for confocal endomicroscope. The compact endomicroscopic probe comprises a PZT fiber scanner and a lensed fiber. (b) Design parameters of the lensed fiber. The lensed fiber consists of SMF, CSF, and a built-in lens at the end of CSF. The incident beam from the SMF is expanded through the CSF region and focused by a fiber lens. (c), (d) Numerical simulation results of lensed fiber. (c) The beam spot size and (d) the working distance of the lensed fiber depend on the length of CSF and the ROC of the fiber optic lens.

JOM_1_3_034501_f001.png

Figure 1(b) shows the design schematics of lensed fiber. The lensed fiber consists of a hybrid optical fiber comprising an SMF serially fusion spliced with a CSF segment and a built-in lens fabricated at the end of CSF. The CSF as a beam expanding region was applied to achieve a sufficiently large beam size to the lens. The curved surface of the lens focuses incident light to the sample and collects photons from the sample for confocal imaging. The optical performance of the lensed fiber was analyzed using the ray matrix transformation, based on the ABCD matrix analysis at 488-nm wavelength.26,27 Light propagation was considered as a Gaussian beam in this theoretical model. The working distance d and the beam waist size ωf of a lensed fiber are described as

Eq. (1)

d=(n1n2n2R)(λn1πωi2)2Lc(n1n2+Lc(n2n1)n2R)(n2n1n2R)2+(λn1πωi2)2(n1n2+Lc(n2n1)n2R)2  ,

Eq. (2)

ωf=ωi(1+n2n1n2Rd)2+(λn1πωi2)2(n1n2d+Lc(1+n2n1n2Rd))2.
The terms of n1, n2, R, λ, ωi, and Lc denote the refractive index of CSF (1.4630 at 488 nm in wavelength), the refractive index of the air, the ROC of a lens, a light wavelength (488 nm), the beam radius of SMF (ωi=3.7  μm), the length of CSF, respectively. The length of the CSF and the ROC serve as the critical design factors for the working distance and the beam waist size. The commercial SMF and CSF were employed in the simulations and the input Gaussian beam was 488 nm wavelength. Figures 1(c) and 1(d) show the calculated results for the beam spot size and the working distance depending on the CSF length and the ROC of a lens. A wide range of imaging characteristics can be determined by properly selecting the design parameters. Note that, a small beam spot size was desirable for obtaining high-resolution confocal imaging. The calculated results clearly indicate that the minimum beam spot at focus occurs when the CSF length becomes longer with the small ROC of a fiber lens.

2.

Results and Discussion

Figure 2(a) outlines the fabrication process for the lensed fiber probe includes three main steps. A 125-μm diameter SMF (S405XP, Thorlabs, Inc.) and CSF (FG125LA, 125-μm cladding diameter, Thorlabs, Inc.) were selected for the fabrication. All protective coatings of fibers were removed in isopropanol alcohol solution prior to the fusion splicing process. The CSF was first fusion-spliced to the distal end of SMF using a fiber-optic fusion splicer (GPX 3400, Vytran) with filament (V2 filament for 125-μm diameter optical fiber). An optical fiber cleaver (LDC-400, Vytran) was then employed to obtain desire length of CSF under the precise control of cleave position. Afterward, the SMF spliced with certain length of CSF was placed in fusion splicer to fabricate a lens at the end of CSF. The incident light source exits from the SMF core expands through CSF region. Note that, the CSF was employed to achieve a sufficiently large beam size to lens. On the other hand, concentric interference patterns generate after a certain length of CSF, which are due to reflection of light from the CSF-air boundary. Therefore, CSF length was first optimized considering the NA of SMF, refractive index of CSF, and incident wavelength. The optimal length of CSF was selected about 650  μm by measuring the beam propagation using confocal laser scanning microscopy (LSM 510/CF26, Carl Zeiss). The lens for beam focusing was fabricated by melting the end of CSF after forming the beam expanding region. The ROC of the lensed fiber was controlled by filament power with fixed duration time of fusion splicer. The ROC at 62.5  μm away from the axial axis can be achieved from 75 to 180  μm with different filament power as shown in Fig. 2(b). A low electric discharge power may cause an incompletely melted CSF, which results in the large value and the variation of ROC. However, a longer melting time under high power results in low ROC and low variation. Under applied higher power, the ROC variation along the lateral direction increases due to the gravity and then the melted CSF further increases the ROC. The minimum ROC of 75  μm with 650  μm CSF length was fabricated to achieve a high-resolution confocal imaging probe as shown Fig. 2(c). The measured ROC of upper and lower hemisphere shows only 0.4% difference and the maximum ROC difference along the axial axis is measured by 9  μm. The working distance and focused beam spot size were evaluated using a confocal laser scanning microscope and 488-nm fiber coupled laser source. The incident beam propagating through the SMF was expanded in the CSF region and then focused by the fiber lens. The calculated NA and the working distance of the lensed fiber are 0.16 and 290  μm, respectively. In addition, the full-width half-maximum (FWHM) of the focal spot is 2.3  μm.

Fig. 2

(a) Fabrication procedure of lensed fiber. (b) The filament power of fusion splicer was the parameter to control the ROC of the lens. (c) Optical image of a fabricated lensed fiber.

JOM_1_3_034501_f002.png

A fully packaged confocal endomicroscope was developed using a Lissajous lensed fiber scanner. A lensed fiber was assembled with PZT-driven optical fiber scanners for Lissajous scanning as shown in Fig. 3(a). The length and the diameter of the PZT were 13 and 1.5 mm, respectively. Note that the PZT tube-driven resonant Lissajous fiber scanner is widely used in endomicroscopic catheters, which has many advantages such as high scan amplitude with low operating voltage, compact, and robust packaging. The optical fiber scanner consists of piezoelectric tube, a lensed fiber, and a micromachined tethered silicon oscillator (MTSO). The MTSO consists of a silicon spring and additional mass28 and mounted on the lensed fiber with UV curable polymer (NOA 63, Norland Optics) to modulate the stiffness and resonant scanning frequency. The lensed fiber was forced to oscillate in resonant frequency by Lissajous fiber scanner. The fiber scanner operates in first-mode vibration and the resonant frequency of x-axis and y-axis are 1002 and 1210 Hz, respectively.29 The Lissajous scanning lensed fiber was then fully packaged with a stainless-steel tube housing of 2.6 mm in diameter and 20 mm in length and an optical window of 2.5 mm diameter as shown in Figs. 3(b) and 3(c). Note that such a short length of catheter stainless-steel tube allows the direction insertion through the working channel of conventional clinical flexible endoscope. The optical characterization of the probe was also performed by the optical sectioning along the z-axis using a conventional confocal laser scanning microscope. The working distance was about 190  μm from the optical window and the FWHM of the intensity field at the focal spot is 2.6  μm as shown in Figs. 3(d) and 3(e). A slight expansion in FWHM is due to the misalignment during the packaging process. The compactly packaged confocal endomicroscope was easily inserted into the instruments channel of a conventional endoscope such as flexible and rigid endoscope, which can also provide a compact handheld probe for in vivo tissue imaging or optical biopsy applications as shown in Figs. 3(f) and 3(g).

Fig. 3

(a) Lissajous fiber scanner was demonstrated based on a lensed fiber probe with a piezoelectric driven actuator. (b), (c) The Lissajous lensed fiber scanner was fully packaged with a stainless-steel tube (2.6-mm outer diameter and 20-mm total length) and a laser cut cover glass (0101030, Marienfeld) as optical window (2.5-mm outer diameter and 0.14-mm thickness) for developing endomicroscopic probe. (d) Optical characterization of confocal endomicroscopic probe. The beam profile was measured by optical sectioning along the z-axis with a conventional confocal laser scanning microscope. (e) The beam profile of probe before (dotted line) and after (solid line) packaging at the focal plane. The working distance is about 190  μm from the optical window. The FWHM of the intensity field at the focal spot before and after packaging are 2.3 and 2.6  μm, respectively. (f), (g) The compact confocal endomicroscopic probe was easily incorporated into an instrument channel of conventional endoscope such as (f) colonoscope and (g) laparoscope.

JOM_1_3_034501_f003.png

The fully packaged probe was finally connected to a confocal endomicrscope system. Figure 4(a) shows a schematic illustration for confocal endomicroscope system. The 488-nm wavelength fiber coupled laser source (LDM series 488, LASOS) was coupled to FC/APC collimator (PAFA-4-B, Thorlabs, Inc.) for fluorescein isothiocyanate (FITC) imaging. The excitation light was reflected to the dichroic mirror (DM, FF497-Di01, Semrock) and coupled to the Lissajous scanning lensed fiber probe. The fluorescence signal transmits though the DM and a band pass filter (FF02-525/40-25, Semrock). The photomultiplier tube (PMT: 9110, Hamamatsu) detects fluorescence signals, which was transmitted to an image signal processor (ISP). The ISP consists of a function generator, analog-to-digital convertors (ADC), and a video memory. The ADC converts the analog PMT and driving signals to digital signal with 2.5-MHz sampling rate. The phase delay between the driving signal and the fiber scanner motion was matched manually, and adjusted digital signals were transferred to the video memory. The ISP reconstructs image coordinates by mapping the PMT signal based on the calibrated driving signals in the confocal images of 256×256  pixels on the field programmable gate array board. The reconstructed image is finally displayed on the monitor. The system provides 10-Hz frame rate and over 80% fill factor by operating the fiber scanner with 1000- and 1210-Hz resonant frequencies. Using the selection rule of scanning frequencies, both axis scanning frequencies are selected to have high greatest common devisor (GCD) and high total lobe number for Lissajous pattern, i.e., the sum of scanning frequencies divided by the GCD of biaxial scanning frequencies.25 The confocal fluorescence images were obtained from a stained mouse kidney section and mouse ear skin to evaluate the imaging performance of Lissajous scanning lensed fiber. Figure 4(b) shows the reflectance imaging result of groups 6 and 7 in aluminum pattern of US Air Force (USAF) target. The minimum width of about 2.19  μm in group 7 was clearly resolved by confocal endomicroscope. The total FOV was 217  μm×217  μm at 30Vpp. An ex vivo imaging of the mouse kidney section stained with Alexa Fluor 488 wheat germ agglutinin was performed to compare confocal images using endomicroscope system and conventional confocal laser scanning microscope as shown in Fig. 4(c). The convoluted tubules of the mouse kidney were clearly observed in confocal images, which was marked with white arrows. The ex vivo confocal fluorescence imaging while in contact with mouse ear blood vessel was also performed as shown in Fig. 4(d). The confocal images are comparable to those from GRIN lens-based fiber scanning endomicroscope,29 although the lensed fiber endomicroscopic probe is fully packaged with 30% shorter length. The confocal fluorescence imaging results of biological specimen show that the developed compact Lissajous scanning lensed fiber probe can be replaced conventional bulk-optics-based imaging probe.

Fig. 4

(a) A schematic illustration for confocal endomicroscope system. (b) Confocal reflectance images of groups 6 and 7 from high-resolution USAF test target acquired using the Lissajous lensed fiber scanner. (c) The ex vivo image of mouse kidney section stained with Alexa Fluor 488 (F-24630, Invitrogen) using confocal endomicroscope system. The lumen of the convoluted tubule is marked with white arrows. (d) An ex vivo confocal fluorescence image of mouse ear blood vessel after injecting 1% FITC to a wild-type mouse. The total FOV and applied voltage of each figures were (b) 217  μm×217  μm at 30Vpp, (c) 194  μm×194  μm at 30Vpp, and (d) 428  μm×428  μm at 50Vpp.

JOM_1_3_034501_f004.png

3.

Conclusions

In summary, this work has successfully developed the objective lens-free imaging probe for confocal laser endomicroscope. The Lissajous scanning lensed fiber was developed to eliminate bulk beam focusing elements and demonstrated confocal imaging of a biological specimen. The ray tracing analysis using ABCD matrix was performed to optimize the imaging characteristics of the lensed fiber. The lensed fiber scanner was fully packaged with a stainless-steel tube housing of 2.6 mm in diameter and 20 mm in length. The compact scanning probe was easily incorporated into a conventional endoscope such as colonoscope and laparoscope. The reflection and ex vivo confocal images were successfully obtained using confocal endomicroscope without conventional GRIN lens-based optical system. This compact Lissajous scanning lensed fiber can provide a new direction for developing miniaturized endoscopic imaging probe such as OCT, confocal laser microscopy, or two-photon microscopy.

Acknowledgments

This work was supported by the National Research Foundation of Korea (NRF) funded by the Ministry of Science ICT & Future Planning (2016R1A2B301306115) and Ministry of Trade, Industry & Energy, Republic of Korea (P0013915). Disclosures: The authors declare no conflicts of interest.

References

1. 

W. Denk, J. Strickler and W. Webb, “Two-photon laser scanning fluorescence microscopy,” Science, 248 73 –76 (1990). https://doi.org/10.1126/science.2321027 SCIEAS 0036-8075 Google Scholar

2. 

H. Bao et al., “Fast handheld two-photon fluorescence microendoscope with a 475  μm×475  μm field of view for in vivo imaging,” Opt. Lett., 33 1333 –1335 (2008). https://doi.org/10.1364/OL.33.001333 OPLEDP 0146-9592 Google Scholar

3. 

D. Y. Kim et al., “Lissajous scanning two-photon endomicroscope for in vivo tissue imaging,” Sci. Rep., 9 (1), 3560 (2019). https://doi.org/10.1038/s41598-019-38762-w SRCEC3 2045-2322 Google Scholar

4. 

A. Dilipkumar et al., “Label-free multiphoton endomicroscopy for minimally invasive in vivo imaging,” Adv. Sci., 6 (8), 1801735 (2019). https://doi.org/10.1002/advs.201801735 Google Scholar

5. 

D. Huang et al., “Optical coherence tomography,” Science, 254 1178 –1181 (1991). https://doi.org/10.1126/science.1957169 SCIEAS 0036-8075 Google Scholar

6. 

H.-C. Park et al., “Forward imaging OCT endoscopic catheter based on MEMS lens scanning,” Opt. Lett., 37 2673 –2675 (2012). https://doi.org/10.1364/OL.37.002673 OPLEDP 0146-9592 Google Scholar

7. 

B. J. Vakoc et al., “Cancer imaging by optical coherence tomography: preclinical progress and clinical potential,” Nat. Rev. Cancer, 12 363 (2012). https://doi.org/10.1038/nrc3235 NRCAC4 1474-175X Google Scholar

8. 

B. R. Masters, “Book review: Handbook of biological confocal microscopy, Second Edition, Edited by J. B. Pawley,” Opt. Eng., 35 1 (1996). https://doi.org/10.1117/1.600871 Google Scholar

9. 

P.-L. Hsiung et al., “Detection of colonic dysplasia in vivo using a targeted heptapeptide and confocal microendoscopy,” Nat. Med., 14 454 (2008). https://doi.org/10.1038/nm1692 1078-8956 Google Scholar

10. 

G. Li et al., “Ultra-compact microsystems-based confocal endomicroscope,” IEEE Trans. Med. Imaging, 39 (7), 2406 –2414 (2020). https://doi.org/10.1109/TMI.2020.2971476 ITMID4 0278-0062 Google Scholar

11. 

D. Wang et al., “Correction of image distortions in endoscopic optical coherence tomography based on two-axis scanning MEMS mirrors,” Biomed. Opt. Express, 4 2066 –2077 (2013). https://doi.org/10.1364/BOE.4.002066 BOEICL 2156-7085 Google Scholar

12. 

C. Duan et al., “An endoscopic forward-viewing OCT imaging probe based on a two-axis scanning MEMS mirror,” in IEEE 11th Int. Symp. Biomed. Imaging (ISBI), 1397 –1400 (2014). https://doi.org/10.1109/ISBI.2014.6868139 Google Scholar

13. 

J. Sun et al., “3D In Vivo optical coherence tomography based on a low-voltage, large-scan-range 2D MEMS mirror,” Opt. Express, 18 12065 –12075 (2010). https://doi.org/10.1364/OE.18.012065 OPEXFF 1094-4087 Google Scholar

14. 

W. Liang et al., “Increased illumination uniformity and reduced photodamage offered by the Lissajous scanning in fiber-optic two-photon endomicroscopy,” J. Biomed. Opt., 17 021108 (2012). https://doi.org/10.1117/1.JBO.17.2.021108 JBOPFO 1083-3668 Google Scholar

15. 

D. Do, H. Yoo and D. G. Gweon, “Fiber-optic raster scanning two-photon endomicroscope using a tubular piezoelectric actuator,” J. Biomed. Opt., 19 066010 (2014). https://doi.org/10.1117/1.JBO.19.6.066010 JBOPFO 1083-3668 Google Scholar

16. 

E. J. Seibel and Q. Y. J. Smithwick, “Unique features of optical scanning, single fiber endoscopy,” Lasers Surg. Med., 30 177 –183 (2002). https://doi.org/10.1002/lsm.10029 LSMEDI 0196-8092 Google Scholar

17. 

N. Zhang et al., “Compact piezoelectric transducer fiber scanning probe for optical coherence tomography,” Opt. Lett., 39 186 –188 (2014). https://doi.org/10.1364/OL.39.000186 OPLEDP 0146-9592 Google Scholar

18. 

C. M. Lee et al., “Scanning fiber endoscopy with highly flexible, 1 mm catheterscopes for wide-field, full-color imaging,” J. Biophotonics, 3 385 –407 (2010). https://doi.org/10.1002/jbio.200900087 Google Scholar

19. 

D. Lorenser, R. A. McLaughlin and D. D. Sampson, “Optical coherence tomography in a needle format,” Optical Coherence Tomography in a Needle Format, 2413 –2472 2015). Google Scholar

20. 

D. R. Rivera et al., “Use of a lensed fiber for a large-field-of-view, high-resolution, fiber-scanning microendoscope,” Opt. Lett., 37 (5), 881 –883 (2012). https://doi.org/10.1364/OL.37.000881 OPLEDP 0146-9592 Google Scholar

21. 

S. Kato, S. Chonan and T. Aoki, “High-numerical-aperture microlensed tip on an air-clad optical fiber,” Opt. Lett., 39 (4), 773 –776 (2014). https://doi.org/10.1364/OL.39.000773 OPLEDP 0146-9592 Google Scholar

22. 

E. J. Min et al., “Single-body lensed-fiber scanning probe actuated by magnetic force for optical imaging,” Opt. Lett., 34 (12), 1897 –1899 (2009). https://doi.org/10.1364/OL.34.001897 OPLEDP 0146-9592 Google Scholar

23. 

H. C. Park, Y. H. Seo and K. H. Jeong, “Lissajous fiber scanning for forward viewing optical endomicroscopy using asymmetric stiffness modulation,” Opt. Express, 22 5818 –5858 (2014). https://doi.org/10.1364/OE.22.005818 OPEXFF 1094-4087 Google Scholar

24. 

K. Hwang, Y. H. Seo and K. H. Jeong, “Microscanners for optical endomicroscopic applications,” Micro Nano Syst. Lett., 5 (1), 1 –11 (2017). https://doi.org/10.1186/s40486-016-0036-4 Google Scholar

25. 

K. Hwang et al., “Frequency selection rule for high definition and high frame rate Lissajous scanning,” Sci. Rep., 7 14075 (2017). https://doi.org/10.1038/s41598-017-13634-3 SRCEC3 2045-2322 Google Scholar

26. 

H. Kogelnik, “On the propagation of Gaussian beams of light through lens like media including those with a loss or gain variation,” Appl. Opt., 4 1562 –1569 (1965). https://doi.org/10.1364/AO.4.001562 APOPAI 0003-6935 Google Scholar

27. 

W. Jung et al., Numerical Analysis of Gradient Index Lens-Based Optical Coherence Tomography Imaging Probes, 10 SPIE Press(2010). Google Scholar

28. 

H. C. Park et al., “Micromachined tethered silicon oscillator for an endomicroscopic Lissajous fiber scanner,” Opt. Lett., 39 6675 –6678 (2014). https://doi.org/10.1364/OL.39.006675 OPLEDP 0146-9592 Google Scholar

29. 

K. Hwang et al., “Handheld endomicroscope using a fiber-optic harmonograph enables real-time and in vivo confocal imaging of living cell morphology and capillary perfusion,” Microsyst. Nanoeng., 6 (1), 1 –11 (2020). https://doi.org/10.1038/s41378-020-00182-6 Google Scholar

Biography

Jae-Beom Kim received his PhD in bio and brain engineering a KAIST, Republic of Korea, in 2018. Currently, he is a senior engineer at SK Hynix in the Republic of Korea.

Jaehun Jeon received his BSc degree in bio and brain engineering from KAIST, Republic of Korea, in 2020. Currently, he is an MS student in bio and brain engineering from KAIST.

Kyungmin Hwang received her PhD in bio and brain engineering from KAIST, Republic of Korea, in 2020. Currently, she serves as a CEO at VPIX Medical, Inc. in the Republic of Korea.

Daniel Y. Kim received his MS degree in bio and brain engineering from KAIST, Republic of Korea, in 2018. He is currently studying at the School of Medicine from Cha University, Republic of Korea.

Ki-Hun Jeong received his PhD from the University of California at Berkeley in 2005. Currently, he is a full professor and the department head in the Department of Bio and Brain Engineering at KAIST in the Republic of Korea, as well as a codirector at KAIST Institute for Health Science and Technology.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Jae-Beom Kim, Jaehun Jeon, Kyungmin Hwang, Daniel Y. Kim, and Ki-Hun Jeong "Objective-lens-free confocal endomicroscope using Lissajous scanning lensed-fiber," Journal of Optical Microsystems 1(3), 034501 (28 May 2021). https://doi.org/10.1117/1.JOM.1.3.034501
Received: 8 February 2021; Accepted: 14 May 2021; Published: 28 May 2021
Lens.org Logo
CITATIONS
Cited by 6 scholarly publications.
Advertisement
Advertisement
KEYWORDS
Confocal microscopy

Scanners

Single mode fibers

Optical fibers

Fusion splicing

GRIN lenses

Packaging

RELATED CONTENT


Back to Top