28 October 2021 Flat-panel conebeam CT in the clinic: history and current state
Author Affiliations +
Abstract

Research into conebeam CT concepts began as soon as the first clinical single-slice CT scanner was conceived. Early implementations of conebeam CT in the 1980s focused on high-contrast applications where concurrent high resolution (<200  μm), for visualization of small contrast-filled vessels, bones, or teeth, was an imaging requirement that could not be met by the contemporaneous CT scanners. However, the use of nonlinear imagers, e.g., x-ray image intensifiers, limited the clinical utility of the earliest diagnostic conebeam CT systems. The development of consumer-electronics large-area displays provided a technical foundation that was leveraged in the 1990s to first produce large-area digital x-ray detectors for use in radiography and then compact flat panels suitable for high-resolution and high-frame-rate conebeam CT. In this review, we show the concurrent evolution of digital flat panel (DFP) technology and clinical conebeam CT. We give a brief summary of conebeam CT reconstruction, followed by a brief review of the correction approaches for DFP-specific artifacts. The historical development and current status of flat-panel conebeam CT in four clinical areas—breast, fixed C-arm, image-guided radiation therapy, and extremity/head—is presented. Advances in DFP technology over the past two decades have led to improved visualization of high-contrast, high-resolution clinical tasks, and image quality now approaches the soft-tissue contrast resolution that is the standard in clinical CT. Future technical developments in DFPs will enable an even broader range of clinical applications; research in the arena of flat-panel CT shows no signs of slowing down.

1.

From Conebeam CT to Flat-Panel CT

The dynamic spatial reconstructor (Mayo Clinic, 1980 to 199013) was one of the first clinical attempts to reconstruct a volume of three-dimensional (3D) data from large-area projection x-ray images. This ambitious vision involved obtaining real-time 3D reconstructions of contrast-enhanced coronary vessels from data acquired in a few milliseconds. Such a feat would push the imaging requirements far beyond what was achievable at the time with state-of-the art single-slice clinical CT scanners. In addition to the need for subsecond imaging of a large volume, this clinical application also required sub-200-μm spatial resolution. The system suffered from a high level of complexity, with multiple x-ray tubes and corresponding x-ray image intensifiers (XRIIs) acquiring images, whereas the whole gantry rotated through partial arcs in 140  ms. The XRII itself had several enabling capabilities, including a large area, high resolution, and high frame rates (60 fps and above). Unfortunately, the advantages of the XRII were offset by significant limitations, including the spatial and signal nonlinearities inherent in the electron acceleration technology, in addition to a limited dynamic range (somewhat offset by multiple gain factors) and optical scatter.

Ultimately, the success of the system was hampered by the size and complexity of the system and by the nonidealities of the imaging chain. These same challenges were faced throughout the early- to-mid 1990s by innovators who explored conebeam CT opportunities using XRIIs mounted on radiation therapy simulators,4,5 CT gantries,6,7 dedicated dental scanners,8 and angiographic C-arms.913

Dynamic, flat-panel, solid-state, x-ray image detectors (DPFs) used in digital fluoroscopy and radiography emerged in the late 1990s. This new generation of dynamic detectors uses a thin layer of x-ray absorptive material deposited on either (1) an electronic active-matrix array fabricated in a film of hydrogenated amorphous silicon (a-Si:H) or (2) an array of active elements integrated on a complementary metal–oxide–semiconductor (CMOS) panel. The nature of the x-ray absorptive material determines if a detector is one of two categories: indirect-conversion (x-ray scintillator-based) or direct-conversion (x-ray photoconductor-based) (see Fig. 6 for an interesting evaluation of commercial flat panels and impact of converter material on modulation transfer function and detective quantum efficiency). As compared with XRIIs, DFPs have no geometrical distortion or vignetting, immunity from blooming in the periphery of high-exposure regions and negligible contrast loss due to internal scatter. DFPs exhibit a wider dynamic range, as compared with (medical imaging) XRIIs, which have limited dynamic range within each electronic gain stage. DFPs provide high resolution over a large area, whereas XRIIs can do so only for small fields of view. Although the earliest flat panels suffered from high electronic noise, by the mid-2000s, the detective quantum efficiency of flat panels exceeded that of image intensifier systems for all but the lowest doses (i.e., <1  μR per frame at the detector).14,15 Finally, manufacturing costs have dropped significantly over the two decades since the introduction of DFPs, and many fabrication facilities can now easily switch their production lines between flat panels for medical imaging and for consumer electronics.

DFPs have replaced XRIIs for fluoroscopy-guided diagnosis and intervention and have been integrated into on-line patient position verification systems in radiation therapy using alternative, higher-absorptivity conversion materials for use with MV photons. They have also enabled dedicated conebeam CT imaging systems for breast, dental, head, and extremity imaging. An indicative (and incomplete) timeline highlighting technical developments of digital flat panels (DFPs) is presented in Fig. 1. Development of new converter-material combinations, new readout technologies, and new direct-conversion technologies has led to lower noise floors, higher frame rates, and expanded dynamic range, all while maintaining the large-area format and high resolution.

Fig. 1

Incomplete but indicative timeline of the development of DFP technology over the last two decades. The evolution in detector size, frame rate, pixel size, and bit-depth is captured. Manufacturers of both aSi:H and CMOS first provided small(er) panels suitable for cardiology and have then evolved their technology to cover large areas, faster readout rates, and higher bit depth. All product names should be considered registered trademarks of the respective companies. V, Varian; GE, General Electric; C, Canon; D, Dexela; T, Trixell; TD, Teledyne DALSA; PE, Perkin Elmer; -b, binned; *dual gain and dynamic gain switching modes.

JMI_8_5_052115_f001.png

The flat-panel detector allows for wide z-axis coverage, resulting in the ability to image entire organs, such as the brain, heart, liver, breast, or kidneys, in a single axial rotational acquisition. In addition, the detector provides ultrahigh spatial resolution in both the radial and axial directions. In general, flat-panel CTs (FPCTs) have continued to focus on clinical applications that require a combined very high spatial and contrast resolution, such that the dose-to-voxel volume trade-off is deemed acceptable in the context of the ALARA principle. While early DFP technology did not meet all of the imaging requirements for all clinical applications (especially readout-rate), the geometric fidelity and high resolution of the detectors allowed for the implementation of flat-panel CT for “semistatic” imaging tasks, such as neurovascular imaging; 3D radiation oncology treatment verification; 3D mammographic CT; and extremity, dental, and ear, nose, and throat (ENT) imaging.

In this historical and technical review, we present the development of flat-panel CT with a focus on four clinical applications: breast FPCT, fixed C-arm FPCT, on-board FPCT, and extremity/head FPCT. Table 1 summarizes parameters for current “state-of-the-art” systems that are available today. This is a single snap-shot in time and as the continuous development of DF detector technology in Fig. 1 demonstrates, we expect further improvements in capabilities and in the resulting image quality for an ever-expanding range of imaging applications.

Table 1

Summary of system parameters for the clinical applications described in this historical review.

Detector specificationsSystem geo.Acq. geometryImage chain specs.Notes
Breast CTMAT: 1024 × 655–b
PS: 384
FP: 40 × 30
FR: 30
TFT
SDD: 92.3
Mag: 1.4
Proj: 300 to 500
AR: 360
ST: 10 to 16
PW: 10
kV: 49 to 65
FIL: 1.6 mm Al to 0.2 mm Cu
FS: 0.1 to 0.3
or
MAT: 1944 × 1536*
PS: 150
FP: 29 × 23
FR: 30
CMOS

SDD: 69.9
Mag: 1.4
Fixed C-arm CTMAT: 1024 × 1024
MAT: 1240 × 960 and smaller
PS: 100 to 200*
FP: 20 × 20, 40 × 40
FR: 15, 30, 60
SDD: 120
Mag: 1.4 to 1.7
Proj: 130 to 600
AR: 200 to 220 (260) single sweep**; New traj #
ST: 3.5 to 20
PW: up to 100
kV: 50 to 125
FIL: Cu
FS 0.3 to 1.0
* 75  μm available
** Robotic trajectory with moving center of rotation to increase FOV # wobble/sinusoidal trajectory to reduce conebeam artifact aSi:H and CMOS
Cone-beam CT for image-guided radiotherapyMAT: 1024 × 1024 –b2 × 2
PS: 200 –b 2 × 2
FP: 41 × 41

SAD: 100*
Mag: 1.2 to 1.6
Proj: e.g., 600
AR: 190 to 360 offset detector geometry to increase FOV up to 50 cm in diameter
ST: 60**
PW: up to 100
kV: 40 to 130
FIL: Bow-tie + var. apertures
FS: 0.5 to 1.2
** Rotation speed limited to 1 RPM for C-arm accelerators
Cone-beam CT now also available on circular gantry designs without speed restriction.
Support for 4D cone-beam CT.
* Allows radiographic simulation of MV beam geometry
Both with CsI:Tl
or
MAT: 1024 × 768 –b 2 × 2
PS: 194 –b 2 × 2
FP: 40 × 30
HeadMAT: var.
PS: var.
FP: 5 × 5 to 25 × 25*
FR: var.
CMOS and aSi
Variable
SDD: 50–70 range
Mag: 1.2 to 2
Proj: var, 180 to 360 typical
AR: 1890 to 360
ST: 5 to 30
PW: 1 typical
kV: 60 to 120
FIL: Al 2 to 14 mm
FS 0.4 to 0.5
VS: 75 to 400  μm
Highly variable
Most systems plug into wall power
ExtremityAT: 1500 × 2000 to 1800 × 2200 typical, –b 2 × 2 typical
PS: 130 to 300
FP: 25 × 30
FR: 15 to 25
SDD: 55
Mag: 1.3 to 1.8
Proj: 200 to 600
AR: 210 to 360
ST: 18 to 60
PW: 1
kV: 60 to 120, 80 to 110 typical
FIL: Al
FS: 0.6
VS: 150 to 350  μm
MAT, matrix; –b, binned; PS, pixel size in μm; FP, flat panel dimensions in cm; VS, voxel size of the 3D reconstructions in μm; Mag, magnification; SDD, source-to-detector distance in cm; SAD, source to axis distance in cm; TFT, thin-film transistor; CMOS, complementary metal–oxide semiconductor; Proj, number of projections; AR, angular range in degrees; ST, scan time in s; FIL, filtration; kV, kilovoltage range; FS, focal spot size in mm.

The systems available in the 2010s had limited detector readout speed, truncation of the trans-axial field of view, limited dynamic range, increased scatter due to large-volume irradiation, and required reconstruction using incomplete data due to the conebeam acquisition geometry. Some of the algorithms developed to mitigate the artifacts arising from the detector limitations are described briefly here. Further developments of the detector technology, including increased frame rates and dynamic range (including dual-gain readout) and improved noise floor (e.g., due to the use of CMOS instead of aSi:H), have enabled “close to true CT” image quality, with low-contrast resolution approaching 5 HU for voxel sizes of 150  μm. Nonetheless, all FPCT systems continue to require dedicated correction algorithms. New opportunities, especially driven recently by the development of AI-based algorithms for image processing and reconstruction, are opening interesting avenues for further improvements.

2.

Reconstruction and Artifact Correction

In the following, knowledge of two-dimensional (2D) CT reconstruction is assumed. An excellent review of CT reconstruction algorithm history is provided in Ref. 16. The artifacts that are common to both traditional and flat-panel CT, such as beam hardening, scatter correction, and spectral correction, are not addressed here since solutions are, in general, similar between standard diagnostic CT and flat-panel-based CT systems.

A literature search in PubMed using the terms [(conebeam CT reconstruction) AND (flat panel)] returns a total of 207 publications between 1999 and 2021, indicating that the topic continues to be an active area of research. In diagnostic flat-panel CT, such as for applications in breast and dental imaging, longer times between image acquisition and reconstructed volume viewing may be acceptable. However, in both radiation therapy and interventional image guidance, image data must be provided during the therapeutic session and therefore fast reconstruction is an absolute necessity. An “acceptable” reconstruction time is <1  min, and a “barely adequate” reconstruction time would be within 3 min. In addition, some level of soft-tissue visualization (e.g., fat/tissue separation) requires that at least 100 (preferably 600 or more) views be acquired.

For these reasons and others, most clinical systems use direct (noniterative) reconstruction with algorithms that permit independent processing of each projection (or of at most two or three acquired in sequence) followed by backprojection in 3D. A few iterations may be used to correct for nonidealities, such as scatter and beam hardening. In general, the standard fanbeam convolution-backprojection algorithms produce unacceptable artifacts for cone angles greater than 8  deg,17 and therefore reconstruction algorithms targeted to large-cone-angle FPCT were developed.

The following “completeness condition” must be true if an analytic solution to the 3D reconstruction problem is desired: “If on every plane that intersects the object there lies a vertex, then one has complete information about the object.”18,19 Some mechanical motions that meet this condition have the apex of the cone moving on one of the following trajectories: a complete sphere, a hemisphere, two orthogonal circles, a helix, a circle and an orthogonal line, or the pattern made by the stitching on a baseball.2022 To keep mechanical implementation simple, earliest FPCT systems used a (close to) circular trajectory, and almost exclusively used some version of the Feldkamp–Davis–Kress (FDK) algorithm,23 a straightforward (but approximate) extension of fanbeam filtered-backprojection. We start here with a description of the mathematically exact algorithm first described by Grangeat, which can be applied to arbitrary trajectories for which complete data are available.2426 This algorithm reduces to the FDK algorithm for the standard (incomplete) perfect-circle trajectory still used in many FPCT imaging systems today.

2.1.

Exact Reconstruction Using Conebeam Backprojection

Inversion of the 2D Radon transform for parallel-beam CT reconstruction using convolution backprojection is described by La Rivière and Crawford.16In analogy with the 2D Radon transform, inversion of the 3D Radon transform is also possible. This property of the Radon transform was exploited by Grangeat in his description of an exact reconstruction algorithm first published in 1987.24

To understand this approach, we link the 2D Radon transform, the 3D Radon transform, and the x-ray transform. The x-ray transform and the 2D Radon transform are identical in the 2D case since an nD construct is projected onto an (n1)-D dimensional construct, i.e., a plane is projected onto line integrals. In 3D, this is different as Fig. 2 shows. The 3D x-ray transform projects the object’s information onto one-dimensional integrals, i.e., line integrals. The 3D Radon transform, however, integrates the 3D object onto 2D plane integrals. Hence, the functions are different, yet related. In fact, part of the 3D Radon transform can be determined from 2D x-ray views, by computing line integrals of the projection (see Fig. 2). Each line on the detector forms a plane with the respective x-ray source. Therefore, computing the line integral on the pixelwise x-ray line integrals will form a plane integral. As Grangeat showed, this is also true, up to a weighting for cone-beam geometries.2426

Fig. 2

Modified from Ref. 27 under the terms of the Creative Commons Attribution 3.0 License. Schematic showing the relationship between the 2D x-ray projection and the 3D Radon transform. The conventional x-ray projection image is formed by the integral through the object along the direction perpendicular to the detector. By integrating along a direction parallel to the detector (denoted by t in the following derivation), an integral over one plane through the object is performed and hence provides the prerequisite data for the 3D Radon transform.

JMI_8_5_052115_f002.png

Based on these observations, Defrise and Clack have proposed an algorithm that can be applied to a multitude of continuous acquisition trajectories, yet reduces to the FDK-method for circular scans; it is presented here for a flat-panel geometry.28

Since each cone-beam projection is filtered independently, we define for each focal spot at location λ a 3D coordinate system attached to the detector, such that the origin O is the orthogonal projection of the orbit point a(λ) onto the detector plane. Orthonormal vectors u and v are chosen within the detector plane (see Fig. 3).

Fig. 3

Schematic showing the geometry of FPCT data acquisition by cone shaped x-ray beam and a flat panel detector, and the relationship among detector, x-ray transform (e.g., the line integral along a(λ)A), Radon plane, and Radon shell. The focal spot trajectory is affected by bearing jitter and gantry sag, causing deviations from a perfect circular trajectory. See Ref. 29 for a standard illustration.

JMI_8_5_052115_f003.png

Referring to Fig. 3, we describe in general terms the outline of the algorithm, which can be clustered into the main steps of cosine-weighting, filtering, and cone-beam backprojection, as is typical for fanbeam reconstruction in CT.

  • Step 1: Cosine weighting of cone-beam projection (u,v,λ), where u and v are the 2D detector coordinates, projection angle λ, and source–detector distance D distance between a(λ) and O]:

    p1(u,v,λ)=Du2+v2+D2p(u,v,λ).

  • Step 2: Radon-based filtering (equivalent to rowwise ramp filter for circular orbits).

  • Step 2.1: Compute the 2D parallel Radon transform (cf. Fig. 3), where s is the perpendicular distance from the origin to the virtual detector row (corresponding to the intersection of the Radon plane with the detector) and μ is the angle of rotation of the virtual detector row. Integrate over t= to , or equivalently from D1 to D2 for a finite-sized detector:

    R1(s,μ,λ)=R{f1(u,v,λ)}=t=top1(scosμ+tsinμ,ssinμtcosμ,λ)dt.

  • Step 2.2: Differentiate with respect to s

    R2(s,μ,λ)=R1(s,μ,λ)s.

  • Step 2.3: Multiply with the trajectory-dependent weighting factor M(s,μ,λ)

    R3(s,μ,λ)=14π2M(s,μ,λ)·|cosμ|·R2(s,μ,λ),
    with
    M(s,μ,λ)=|cosμ|m2|cosμ|m+2max(1(sDcosλcosμsinλ)2,0)m/2,
    and m>2. Here, max (a,b) will return a if a>b and b if b>a. A common choice is m=4.24

    Note that for s, μ, and λ, the actual parameters of the trajectory are required.

  • Step 2.4: Differentiate again with respect to s

    R4(s,μ,λ)=R3(s,μ,λ)s.

  • Step 2.5: Backproject into the original projection

    p2(u,v,λ)=(u2+v2+D2)·R1{f1(u,v,λ)}=(u2+v2+D2)μR4(t,μ,λ)|t=ucosμ+vsinμdμ.

  • Step 3: Backprojection of the cone-beam projections

    f(x)=λ1C2p2(P(x),λ)dλ,
    where C=|xa(λ)| is the distance between the voxel x under consideration and the source position and P(x) is the cone-beam projection of x onto the detector coordinates u and v.

2.2.

Feldkamp–Davis–Kress for Circular Trajectory

For step 1 of the above algorithm on a circular trajectory, factor Du2+v2+D2 is applied prior to convolution with an appropriate apodized filter, and the factor corrects for the difference in path length (or difference in number of voxels to which a particular value in the projection is applied) for rays detected at the edge of the detector compared to rays detected at the center of the detector. The factor 1/C2, applied to the convolved projections during backprojection, accounts for the dependence of ray density at the point (s,μ) on the distance of the point from the x-ray source. For those systems that acquire 2D projection images over only 180 deg plus the fan angle, Parker’s “short scan” weighting scheme30 can be used to smoothly scale the data that are acquired twice, prior to reconstruction.

A comparison of reconstruction image quality for a circular trajectory acquisition reconstructed using FDK and using an exact reconstruction based on Grangeat’s work is shown in Fig. 4. An extension of the circular trajectory by including x-ray source motion in a smooth sinusoidal pattern above and below the plane, reconstructed using Grangeat’s work, is also shown. This trajectory has been proposed as a universal coupled source–detector trajectory for reducing artifact from high-contrast regions.31

Fig. 4

Comparison of reconstruction algorithms for FPCT reconstructions of the head. (a) FDK-type reconstruction; (b) same data reconstructed using a Grangeat-type reconstruction algorithm; (c) and (e) coronal and axial slices through a volume reconstruction of a phantom for data acquired on a “mostly circular” trajectory (circular but with deviations due to gantry sag, bearing jitter, and other C-arm gantry nonidealities), Grangeat-type reconstruction; (d) and (f) coronal and axial slices through a volume reconstruction of a phantom for data acquired using syngo DynaCT Sine Spin™ (a sinusoidal trajectory) showing improved soft tissue image quality especially in the posterior fossa and skull base; (g) clinical reconstruction using the same acquisition geometry and reconstruction as for (d) and (f).

JMI_8_5_052115_f004.png

2.3.

Iterative Reconstruction and Deep-Learning

Although iterative algebraic reconstruction was the first approach used to reconstruct tomographic data from x-ray projections, it was quickly supplanted by the more computationally efficient (from both speed and data access requirements) filtered backprojection algorithms. As CPU and GPU hardware have continued to evolve and associated computation power has increased, investigations into the value of iterative approaches have resurfaced. In iterative reconstruction, the image acquisition/reconstruction process is modeled as a system of linear equations with the number of observations (x-ray line projection measurements) usually far exceeding the number of voxels in the volume to be reconstructed. Given the discretized version of the x-ray projection,

Ax=p,
where A is a system matrix that describes the projection of a linearized volume x to a linearized vector p that contains all observed detector measurements. The above system of equations can be solved using Radon’s ideas to yield
x=(ATA)+ATp,
where AT is a discrete version of the backprojection operation and (ATA)+ is a circulant Moore–Penrose matrix that takes the role of a filtering step. In general, the system of equations represented by the above is quite large, making the matrix inversion too computationally expensive for direct evaluation. Instead, an iterative approach is typically adopted, successively minimizing an error term that compares the measurements with simulated projections. This formulation is flexible and can be used both for overdetermined systems (more measurements than voxels) and underdetermined systems (e.g., sparse data acquisitions) and may include regularization terms that include a priori knowledge to improve the conditioning of the problem, control noise, and enforce desirable image properties in the reconstructed image. The challenge with iterative approaches continues to be the need for instance-specific tuning of several control parameters, which have a significant impact on trade-offs between spatial resolution and noise in the “final” reconstruction. Interestingly, first FPCT-based products that use iterative reconstruction have recently been brought to the market, e.g., iCBCT, currently a standard offering on Varian’s Halcyon™ and Ethos™ platforms.

Research combining iterative reconstruction and correction approaches with AI/deep learning has exploded in the last 5 years, indicating that AI-based iterative reconstruction may become mainstream again.32 Pure machine learning methods suffer from the problem that they will only be applicable to observations that match the training distribution. The wide range of variation in medical imaging and human anatomy poses a challenge to postprocessing and purely data-driven methods. Using the above matrix notation for image formation and reconstruction, individual operations can be mapped onto a neural network. For example, you can have a parameterized solution of the following form:

x=ATFCp.

For example, C could be an optimized trainable projection-domain weighting with (more optimal) redundancy weights, and F could be a learned filtering optimization that applies to a specific acquisition geometry (i.e., limited angle) while maintaining the known physics and geometry of backprojection in the system matrix A. Similarly, fan-beam limited-angle reconstruction can be tackled by encoding 3D geometries in A.33 A neural-net representation of the equation is shown in Fig. 5. Radon-based filtering as in the Defrise–Clack algorithm could also be integrated into such hybrid deep neural networks. In contrast to completely data-driven approaches, these hybrids come with the guarantee of a reduced error in terms of maximal error bounds and increased generalization ability.34 In other approaches, deep learning has also been used to reduce cone-angle artifact in circular cone-beam CT35 and to generate an image-based geometry correction.36

Fig. 5

Visual representation of the matrix inversion equation as a neural network.

JMI_8_5_052115_f005.png

2.4.

FP-Related Artifact Management in FPCT

A recent review article by Tang et al.29 provides an excellent review of cone beam artifacts and methods to correct for those artifacts in axial conebeam CT. We therefore provide a historical view of earlier work done to mitigate those artifacts that arise specifically due to use of a DFP, covering lag and dynamic range, and truncation in the trans-axial direction.

a-Si:H. The amorphous silicon flat panel detector uses indirect technology, with a scintillator layer (often columnar CsI:Tl) that converts x-ray photons to light, and a second layer of photodiodes and active-matrix thin-film transistors made in amorphous silicon. The photodiodes convert photons to electrons and the transistors switch out the current that is collected on the capacitance of the diodes.

Complementary Metal Oxide Semiconductor (CMOS). CMOS DFPs also use a scintillator such as CsI:Tl to convert incident x-rays into visible photons and a photodiode to convert visible photons into electrons for subsequent readout. CMOS technology offers a variety of potential advantages in material characteristics over a-Si:H, including lower dark current (reduced electronic noise), higher charge mobility (lower image lag and higher frame rate), and fabrication with finer pixel pitch compared with a-Si:H active matrix DFPs. New transistor implementations have been developed that mitigate the sensitivity to radiation damage that was seen in early CMOS FPs.

An interesting visualization of the current status of commercial off-the-shelf flat panels is shown in Fig. 6. Both CMOS and a-Si:H panels fall into the lower right quandrant (especially when comparing the 2  lp/mm point on the curve) since they use similar scintillator converter technology. In spite of the attractive noise performance of CMOS FPs, a-Si:H panels continue to be more widely used due to CMOS panel size limitations and cost. An extensive body of work describing modeling, measurement, and calibration techniques to correct for the nonidealities of a-Si:H panels exists. A pictorial summary of the image-correction process implemented for a typical a-Si:H detector is shown in Fig. 8.

Fig. 6

Detective quantum efficiency (DQE) versus modulation transfer function for 17 commercial off the shelf products. The DQE(0) for CsI and Gadox FPs is for a dose of 1  μGy. Modified from Ref. 37.

JMI_8_5_052115_f006.png

Correcting for lag. The use of amorphous silicon leads to slow discharge rates of the transistor elements, and therefore some signal accumulated in one frame is read out in subsequent frames. A second source of lag is scintillator afterglow, although this effect is much smaller than the former. A PubMed search on “flat panel lag” produces 39 results for the time 1999 to 2021.

Both software38,39 and hardware methods to deal with lag in flat-panel imagers (FPIs) were developed early on. One effective hardware method used LEDs built into the panels to backlight the photodiodes, which saturates the traps.40 Forward biasing to push current through the diode can be used to achieve the same goal.4042 These hardware approaches do not correct for scintillator afterglow.

In software correction, an impulse response function (IRF) is generated by fitting a suitable model, such as a multiexponential43 or power function,44 to the lag decay. A single IRF can be used to describe the entire panel, or an independent function can be fit for each individual pixel, which allows for variation in lag across the detector. A basic assumption of the IRF model is that the panel acts in a linear and time-invariant way. The correction is then provided by a temporal deconvolution of the detector output with the modeled IRF. The accuracy of the linear, time-invariant model depends on the size of the object (i.e., on the dynamic range seen by individual pixels) and on the exposure level at which the calibration of lag coefficients and lag rates is carried out. This implies that large signal variations across the panel, such as those near the boundary of an object, show nonlinear behavior.45,46

Implementation of software algorithms for lag correction continues to be an area of research, although most industry users of a-Si:H technology have chosen to implement corrections in reconstructed image space rather than implement a lengthy system-specific calibration process (see Fig. 8).

Dynamic range limitations. The first generation of a-Si:H DFPs provided first 12-bit and then 14-bit analog-to-digital converters (ADCs), and CT scanners of the same era already had detectors with 18-bit ADCs.49 More recent generations of a-Si:H provide 16-bit standard. The limited bit-depth of the DFP projections caused degradation of conebeam CT image quality: (1) large relative errors in beam intensity values measured through dense anatomy, resulting from too large quantization steps, cause streak artifacts off dense objects, (2) insufficient resolution of pixel gain and offset normalization caused ring artifacts, and (3) truncation of low density anatomical detail at patient boundaries, due to insufficient dynamic range, results in shadow artifacts and incorrect Hounsfield numbers. A comparison of 14-bit versus 16-bit FPCT images is shown in Fig. 7 in addition to an FPCT acquired using a dynamic gain readout amplifier that extends the bit-depth to 16.4, demonstrating excellent reconstruction fidelity out to the body boundary.50

Fig. 7

(a) FPCT reconstruction of a head phantom using the dual-gain mode of a Paxscan 4030CB 16-bit flat panel; (b) the same phantom imaged using a 14-bit flat panel—note increase in streaks and ring artifacts. (c) FPCT reconstruction of a human pelvis also using the dual-gain mode. Prior to reconstruction, the projection images were processed using a severe unsharp masking algorithm to enhance any possible boundary artifacts resulting from gain switching—no such boundary artifacts are visible. Images courtesy Ed Shapiro and Rick Colbeth, Varex Imaging.

JMI_8_5_052115_f007.png

Truncation in the trans-axial direction. The maximum dimension of a-Si:H detectors has so far been limited to 40  cm, leading to truncation in the trans-axial direction for 3D imaging of lung fields, abdomen, and hips (e.g., during surgical pin placement guided using mobile C-arms). For such cases, truncation of projection images leads to severe artifacts around the edges of the reconstructed FOV and also to inaccuracies in the reconstructed HU values due to the associated cupping effect in the image.

Building on early work in diagnostic CT imaging, several different techniques were proposed to estimate the missing data based on either a priori knowledge about the data or using redundancies in the data including: symmetric mirroring,51 use of redundant rays,52 polynomial extrapolation,5355 use of a priori information,56 iterative techniques as well as non-FDK reconstruction approaches such as projection onto PI-lines,52 and the truncated Hilbert transform.57 In radiation therapy where focal-spot-to-detector distances are even longer and thus trans-axial truncation is even more severe, use of an off-set detector and a full 360-deg acquisition is the standard practice; the same geometry is often used in dental CT, where cost considerations mandate use of as small a detector as possible. Algorithms to correct for sampling and filtering discontinuities in the overlap region were developed (Fig. 8).5861

Fig. 8

Illustration of the application of standard correction steps to data acquired using a 30×40  cm a-Si:H 16-bit flat panel. Some corrections such as “overexposure” and “truncation” have only a minor effect on this rather small volume. Original data courtesy of Drs. Arnd Doerfler, Stefan Lang, and Philip Hoelter, University Hospital Erlangen; corrections courtesy Dr. Michael Manhart. All images include standard I0, log, and water correction. (a) Window width 100. (b) Window width 200. + indicates application of a correction for the listed nonideality. LFD, low-frequency drop.47,48

JMI_8_5_052115_f008.png

3.

FPCT (kV and MV) in Radiation Therapy

3.1.

Flat-Panel CT for Image-Guided Radiotherapy

Radiotherapy is an essential component of effective cancer control, and clinical evidence supports its use in over 50% of cancer patients at some point in their care. As a local therapy, it is critical to provide accurate and precise placement of the radiation dose within the body. Failure to cover the desired target tissues will undermine the likelihood of success while covering excess tissues increases the likelihood of normal tissue toxicity. In the 1990s, advances in computing for personalized treatment planning and the development of electromechanical radiation beam shaping devices such as multileaf collimators enabled the creation of highly conformal dose distributions and escalated the need for greater precision and accuracy in dose placement at each treatment fraction.62 The traditional use of portal films or portal images based on the megavoltage energy treatment beams did not provide the capacity to visualize and localize soft-tissues, which are well-documented to move relative to bony landmarks.63 The integration of a soft-tissue imaging system at the exact time of treatment to localize soft-tissue structures including the tumor and organs-at-risk was needed. Advances in volumetric reconstruction and detectors promised to make cone-beam CT a potential solution (see Fig. 9).

Fig. 9

Artist’s rendition of FPCT implemented on a radiation therapy treatment system. The FPCT system is installed on the same gantry as the treatment head, with an offset of 90 deg.

JMI_8_5_052115_f009.png

3.2.

Image-Guided Radiotherapy Requirements

The image-guidance problem in radiotherapy had several constraints if it was to be quickly and broadly impactful. From the perspective of quality, images would need to be capable of capturing the contrast between soft-tissues (10 to 30 HUs) within the human body and be capable of providing sufficient resolution to localize these and higher-contrast boundaries to within a millimeter in three dimensions. The radiation dose associated with the imaging activity needed to be low enough (∼ few cGy) as to allow multiple imaging events per treatment session and accommodate up to 40 fractions associated with some treatment protocols [e.g., prostate intensity-modulated radiation treatment (IMRT)]. The image needed to be captured with the patient in the position for treatment, such that the image could inform minor corrections in position to assure accurate targeting was achieved before dose delivery. In addition, the field-of-view of the imaging system needed to accommodate the anatomy of interest (up to 50 cm axially and tens of centimeters superior-inferior), including the surrounding organs-at-risk that were to be avoided during the irradiation. Of course, the images needed to be generated in a time interval that would provide confidence that that patient’s anatomy had not moved significantly between imaging, correction, and treatment (<60 to 120 s). This included the capability to take back-to-back images within minutes for verification purposes. In addition to rapid acquisition, the imaging system needed to provide absolute positional information relative to the treatment isocenter and coordinate framework of the treatment machined (sub-mm tolerance for systematic guidance errors). Finally, the imaging system needed to integrate into the existing treatment machines that were increasingly able to generate highly conformal, IMRT plans. Without image-guidance, the improvements in conformality could not be capitalized upon and without image-guidance could put patients at risk. Of course, the current economic paradigm required that the imaging and guidance process could not substantially reduce the overall treatment capacity of the treatment machine if it was to be clinically and operationally viable. Taken together, these constraints represented a substantial challenge that were fortunately well-matched to on-going developments in cone-beam CT reconstruction,58,64 remarkable increases in computational capacity, and progress in the development of large-area, FPIs.65

3.3.

Development of Integrated Cone-Beam CT for Image-Guided RT

The integration of a kilovoltage x-ray sources into a megavoltage radiation therapy treatment machine dates back to the 1950s. These systems were first used to either overcome the poor contrast of the bony anatomy at megavoltage energies66 or to overcome the large source size in Co60 treatment machines.67 The integration of digital detector systems (key for subsequent volumetric image reconstruction) was first explored by Sephton using a floor-mounted detector in combination with a gantry-mounted x-ray tube.68

The first fully integrated kilovoltage radiographic and volumetric CT capable radiation treatment device was described in 1997 at the ICCR in Salt Lake City.69 This system was built upon a modified Philips SL-20 accelerator (see Fig. 10) that had been adapted to include a retractable x-ray tube and two CCD-based fluoroscopic imaging systems for the orthogonal megavoltage and kilovoltage beams. While clinical utility was promising, many issues related to image quality and geometric accuracy persisted.70

Fig. 10

(a) Installation of kV cone-beam CT system on Elekta SL-20 accelerator drum gantry in 1997 (images courtesy of M. Moreau, D. Drake, and D. Jaffray from William Beaumont Hospital and R. Cooke from Elekta). (b) Cone-beam CT sagittal slice through anthropomorphic head phantom. (c) Image of the pelvis of a canine volunteer. (d)–(g) IRB-approved patient study showing kV, MV, and combined radiographic capabilities.

JMI_8_5_052115_f010.png

Advances in flat-panel imaging technology were critical for the advancement of geometric stability and for improvements in image quality. Fluoroscopic systems suffered from substantial losses in the optical chain despite the use of high efficiency optics (front surface mirrors, f/0.7 lenses), high efficiency conversion plates (Lanex Fast-Back or CsI:Tl scintillators), and cooled back-thinned CCD detectors.70 The development of large area amorphous silicon-based detectors promised to address the coupling issues (Fig. 11) but still suffered from kTC noise, lag, ghosting, and artifacts associated with electronics-related instabilities.7174 These instabilities were a major draw-back for imaging processes that required highly stable gain and offset corrections for reconstructing volumetric images from projections acquired over 60 to 100 s (the rotational gantry speed of a medical linear accelerator).

The geometric characteristics of the DFPs were outstanding compared with the lens-mirror-phosphor coupling of the earlier prototypes, making geometric corrections for mechanical instabilities in the overall gantry assembly a tractable problem requiring only routine quality assurance in the clinical setting.75 Integrating these corrections into the reconstruction process minimized the need for additional preprocessing.76 The resulting images were capable of transferring substantial detail for high contrast objects including fine structure in bony anatomy (see Figs. 12 and 13).

Fig. 11

(a) Early work in the evaluation of amorphous silicon FPIs for KV cone-beam CT system in controlled conditions. (b), (c) Axial images of a euthanized rat comparing image quality between flat-panel CBCT and a traditional Philips SR-7000 CT scanner (effective 1 mm slice thickness) suggested soft-tissue imaging was possible with FPIs.

JMI_8_5_052115_f011.png

Fig. 12

(a) Adaptation of the 20.5-cm FPI to the linear accelerator gantry for studies of image quality and geometric stability. (a)–(c) Comparison of soft-tissue performance between the clinical CT scanner (Philips SR-7000), bench-top, and gantry-mounted systems (1-mm effective slice thickness).

JMI_8_5_052115_f012.png

Fig. 13

Installation of the large-area 41-cm FPI (Perkin-Elmer RID1640) on the Elekta SL-20 accelerator gantry in 2000. (a)–(d) Reconstruction of anthropomorphic head phantom demonstrating reconstruction in 3D and (d)–(f) remarkable resolution in the progressively zoomed-in sagittal slice set. Reconstructed at 300  μm resolution using modified FDK filtered-backprojection method.

JMI_8_5_052115_f013.png

Efforts to adopt the use of megavoltage beams for cone-beam CT image-guidance were also pursued with reasonable success.77,78 Figure 14 shows the substantial progress made in the development of efficient detectors and flat panel technology. However, the fundamental issues of detector efficiency, treatment collimator limiting field-of-view, and dose to achieve acceptable contract-to-noise prevented commercial success.79

Fig. 14

(a) 6-MV conebeam CT using Varian’s a-Si:H portal vision system adapted to a Varian 600CD. Axial and coronal slices demonstrate the advantage of conebeam CT for guidance. The isocenter volume covered is a 20-cm cube, which included nearly the entire head. The total number of monitor units (MU) employed in generating the data set was prohibitive at 1500. (b) Axial, coronal, and sagittal CT slices of a head phantom acquired at 6 MV using a 16-cGy irradiation over a 1-min rotation on a Varian 21EX clinac. This was the result of a high efficiency 1-cm-thick pixelated CsI flat panel detector prototype built and tested at MSKCC. It was concluded that this dose is still a factor of 3× too high, and the thick scintillator was cost prohibitive for a commercial product.77

JMI_8_5_052115_f014.png

Concerns related to the accuracy of simple circular trajectories were also a factor in accepting the utility of cone-beam CT imaging approaches.80 More complete analyses of the implications of incomplete data acquisition allowed the community to understand the implications for clinical utility.81

Image-quality issues related to the largely unmanaged presence of x-ray scatter generated in the patient reaching the detector continue to be an issue.82 Empirical approaches drawn from the traditional CT scanning systems83 and measurement-based approaches have been explored.84,85 Several groups continue to develop computational solutions based on analytical and Monte Carlo approaches.8689 Ongoing efforts on fluence-field modulation approaches offer promise to further mitigate the impact of scatter and better deploy the dose budget for a desired imaging task.90,91 These approaches are applicable to all applications of cone-beam CT.

3.4.

Clinical Adoption and Ongoing Challenges

The clinical utility of these devices depended critically on the development of rapid image reconstruction methods and integrated image-guidance tools to allow on-line images to be used for corrections in position.92 Further improvements included the development of four-dimensional (4D) cone-beam reconstruction techniques that enabled projection-based sorting for multiphase reconstructions of the thorax.93 These technological advances have allowed image-guidance technologies to be broadly deployed with multiple commercial solution introduced to the market and thousands of these systems deployed in cancer centers around the world (Figs. 15 and 16).

Fig. 15

(a) Early prototype of Elekta Synergy at Princess Margaret Cancer Center, Toronto, Canada. (b) Coronal slice showing prostate anatomy. (c) Coronal slice demonstrating lesion in the lung prior to SBRT. (d)–(f) Thoracic spine reconstructed at a resolution of 120  μm. All images acquired with 50-cm diameter FOV using an offset detector, 120 kVp, and <5  cGy imaging dose.

JMI_8_5_052115_f015.png

Fig. 16

Varian’s on-board imaging system. (a) Geometry on 2100c with robotic arms with imaging phantom. (b) First patient head images dated September 11, 2004, at MSKCC (125 kVp, 80 mA, 8  ms×600 projections, scan diameter 26 cm, axial image 512×512  pixels at 0.5 mm and slice thickness 2.5 mm from five slice average). All images courtesy of Ed Shapiro, Varian.

JMI_8_5_052115_f016.png

4.

C-Arm FPCT in Angiography and Image-Guided Interventions

4.1.

From XRIIs to DFPs: C-Arm Conebeam CT in the Interventional Suite

Endovascular therapy options for the treatment of various vascular diseases have exploded since the first interventional coronary treatments. These advances have been made possible mainly through the development of interventional devices and of new techniques for their deployment. In general, image guidance during such procedures is provided by 2D projection images: real-time x-ray fluoroscopy and digital subtraction angiography (DSA). Conebeam CT images provided during an intervention (see Fig. 17) provide benefits for diagnosis, treatment planning, therapy guidance, and outcomes prediction, augmenting the 2D projection images with 3D quantitative information. Conebeam CT is always an adjunct to the main goal of providing real-time, very high-resolution angiograms during subselective injections of contrast. In addition to endovascular therapies, conebeam CT has also proven to be valuable in percutaneous needle-based interventions, such as ablation, biopsies, and vertebroplasty by providing some soft-tissue contrast combined with accurate 3D device and bone visualization to augment real-time 2D x-ray fluoroscopy. 3D imaging capability must be provided without interfering with patient access and standard workflows on C-arm gantries that are used across a wide range of applications from surgical guidance, to neuro interventions, to body and peripheral interventional radiology.

Fig. 17

Artist’s sketch of trajectories for conebeam CT acquisitions using a flat-panel C-arm. (a) Propellor mode is used primarily for neurointerventions; (b) patient-side mode is used primarily for body interventional radiology and surgical interventions.

JMI_8_5_052115_f017.png

A standard interventional C-arm system has several different operating modes. Real-time low-exposure, high frame-rate fluoroscopy (1  μR per frame at the detector, up to 60 fps in cardiology) is used to guide the positioning of endovascular or percutaneous devices. When higher quality diagnostics are needed, a series of high-exposure images (typically two to three frames per second, at 100 to 500  μR per frame at the detector) is obtained. For endovascular interventions, an injection of contrast material (5 to 15 ml) enables vessel visualization. For DSA imaging, a few images without contrast are obtained prior to the injection, and digital subtraction of a “mask”—or noncontrasted—image from the contrast-filled images provides clear visualization of vessels without the presence of background clutter such as bone, airways, or teeth. Fluoroscopy is then used again to guide the placement of devices, embolics, etc., and further high-dose or DSA series are obtained to verify placement of devices and to verify treatment success at the completion of the procedure.

Proof of principle for C-arm-based conebeam CT imaging was first demonstrated in the 1990s by several different academic-industry partnership teams, using clinically available C-arm systems mounted with XRIIs.912 Data were acquired over limited angular range (<60  deg), and algebraic iterative techniques were used to reconstruct 3D vascular trees from few-view (<20) sparse vascular data sets. In London, Ontario, rotational motion of a clinical XRII-based C-arm was tracked using a physical rotary encoder attached to the gantry, angle-dependent XRII distortion was corrected using a 2D grid of beads placed on the XRII,94,95 and the imperfect acquisition geometry (i.e., noncircular) trajectory of the x-ray focal spot was measured (and found to be nominally reproducible) using a single bead fixed in 3D lab space.96 Reconstruction using the FDK algorithm with Parker weighting provided some of the first in vivo C-arm conebeam CT images of low-contrast structures and contrast-filled vessels, as shown in Fig. 18.13

Fig. 18

(a) Anteroposterior and (b) lateral maximum intensity projection views through a reconstructed volume of pig cerebral vasculature. Vessel labels: 1, common carotid; 2, ascending pharyngeal; 3, rete; 4, anterior branch of internal carotid; 5, inferior dental; 6, lingual; 7, wire in the endotracheal tube; 8, dense bone surrounding the inner ear. (c) Slice through the 3D volume at the height of the rete; note star pattern surrounding the common carotid due to pulsatile variation of opacification. (d) Surface-rendered image of a 3D reconstruction of a surgically created side-wall aneurysm. (e) In the same aneurysm, a lower injection rate led to incomplete filling of the aneurysm and therefore to underestimation of the aneurysm cross-section.

JMI_8_5_052115_f018.png

Replacement of the XRII by flat panels enabled acquisition of geometrically accurate, angle-independent projection images with a monotonic signal response that could be corrected to linear. The remaining somewhat unique challenge for C-arm-based CBCT was accurate characterization of the acquisition geometry. 2D images acquired during clinical interventional procedures often require steep angulations, and a C-arm gantry must provide highly flexible positioning. Therefore, the C-gantry is slim and retains some flexibility. However, the acquisition geometry of the mechanical system must be determined to an accuracy that is considerably better than the expected resolution of the final image. An alternative is to compute the trajectory of the source and detector system from a set of image-based 3D-2D correspondences. Using the reconstruction algorithm (and variants) described above, any trajectory that provides sufficient data can then be reconstructed exactly, and approximations can be applied for data acquired using incomplete trajectories. Clinical fixed C-arm systems today rely on reproducible trajectories and use a combination of image-based, external tracker-based (e.g., laser, depth camera) and physics-modeling-based approaches to characterize the intrinsic and extrinsic parameters of the projection matrix for each projection (see Faugeras95 for a complete description of projection matrix formalism). Effects, such as wobble due to bearings, and movement of the central ray on the detector due to gravity-induced sag, can significantly degrade the resolution of the CBCT reconstructed volume if not properly characterized or if the system calibration is not up to date. Accurate system calibration is also required to place the reconstructed volume accurately in the angiographic lab space, enabling registration with pre-interventional data sets and accurate comparison of 3D image content pre-, intra-, and postintervention.

Figure 18 shows first in-vivo images acquired using the clinical XRII C-arm system in London, Ontario. Image quality is surprisingly good, although several artifacts can be seen, such as streak artifact due to limited number of projection images (only 130) and nonlinear partial volume artifact (further exacerbated by C-arm trajectory wobble, bearing jitter, and sag), beam hardening and time-varying opacification of vessels since imaging was carried during an arterial injection of contrast agent over several cardiac cycles. Other more subtle artifacts include residual veiling glare, and ring artifact due to incomplete relinearization of projection data acquired with a beam-shaping filter (bathtub shape) used to reduce the dynamic range of the images at the detector.

This effect of variable contrast opacification continues to be one of the challenges for FPCT in angiography; the result is two bright streaks per cardiac cycle, radiating out from the center of the vessel and diametrically opposed to each other.97 Streaks lead to inaccurate vessel boundaries and may obscure subtle detail and smaller vessels in the tissue surrounding the larger vessel. Nonuniform mixing within an aneurysm may also lead to uncertainty in the aneurysm volume and diameter as shown in Fig. 18.

4.2.

4D Digital Subtraction Angiography

4D-DSA is an extension of 3D DSA imaging that permits time-resolved visualization of a bolus of contrast through 3D vascular structures. Two different approaches have been proposed. The first from Schmitt et al.98100 reconstructs a conventional FPCT volume, followed by an additional 2D-digital subtraction angiogram whose dynamic properties are then mapped to the 3D volume. A second approach proposed by Mistretta and coauthors101,102 acquires two “extended arc” mask and fill runs, and a standard 3D DSA volume is reconstructed. A constraint volume is formed by segmenting the vessel tree. Backprojecting each fill-run projection onto the 3D vessel tree results in a volumetric, time-resolved 4D-DSA volume. Example images are shown in Fig. 19. The technique has been successfully applied to visualization of complex vascular disease such as arteriovenous malformations; the ability to see any desired viewing angle at any desired time of bolus passage assists with planning injections of embolic material and has also enabled accurate planning of stereotactic radiosurgery.104 Further extensions of the technique, including tracking the time-varying contrast density due to variable dilution over the cardiac cycle, are being explored to provide quantitative estimates of vascular flow.105107

Fig. 19

(a) 4D-DSA reconstruction of an arteriovenous malformation. (Top row) Seven 4D-DSA volumes reconstructed from a single extended-arc DSA acquisition viewed from antero-posterior direction; (bottom row) the same volume at the same seven time points as top, but rotated 20 deg between each time point; reconstructed using temporal consistency constraints.103 (b) Results from a prototype reconstruction algorithm with estimated bolus arrival time coded in color. All images courtesy of Prof. Charbel Mounayer, Dupuytren University Hospital, Limoges; reconstructions provided by Dr. Annette Birkhold, Siemens Healthcare GmbH.

JMI_8_5_052115_f019.png

4.3.

Motion-Compensated Reconstruction from Single-Sweep Acquisitions

4.3.1.

Rigid-motion correction

Especially for image-guided neurovascular interventions, correction for rigid motion of the head is key to achieving consistent “close to clinical CT” soft tissue image quality. The challenge is exacerbated by the low frame rates of flat panels, and therefore imaging times can be as long as 10 to 15 s in order to acquire 600 projections. Interestingly, system geometric calibration (discussed above) and motion compensation are closely related, with use of an “objective image quality” metric between simulated and measured projections or iterative correction approaches using image-based metrics such as gradients or entropy to measure artifact levels. An alternative approach exploits redundancies in the projection images, identified via application of epipolar geometry,26,108,109 and has been applied with success to C-arm CT images of the head, as shown in Fig. 20.

Fig. 20

(a) and (c) Motion-corrupted images; (b) and (d) motion-corrected images using epipolar consistency approaches. Clinical images courtesy of Profs. Arnd Doerfler, Stefan Lang, and Philip Hoelter, University Hospital Erlangen and FAU, Erlangen; reconstructions provided by Dr. Michael Manhart, Siemens Healthcare GmbH.

JMI_8_5_052115_f020.png

4.3.2.

Cardiac motion correction

Single-sweep C-arm CT images can provide good image quality in 3D volumes of the left atrium and pulmonary vein anatomy, as well as the left ventricle.110 In addition, adenosine-induced asystole and rapid ventricular pacing have been investigated to improve the image quality of single-sweep acquisitions.111113 Significant effort has been, and continues to be, dedicated to CT reconstruction of the coronary arteries for guidance of percutaneous coronary interventions as summarized in an excellent review by Cimen et al.114 For the coronaries, continuous opacification must be combined with accurate centering of the heart (especially when using a smaller 20  cm×20  cm flat panel) and maintenance of breathhold to limit blurring due to respiratory motion. Tomographic reconstructions methods then apply one of (1) gating, (2) motion compensation, or (3) gating and motion compensation while being adapted to the specific problem of high-contrast moving objects.

4.3.3.

Breathing motion correction

FPCT is used during interventional hepatic vascular procedures such as transarterial chemoembolization (TACE) or selective internal radiation therapy.115 Degradation of image quality due to breathing motion has been shown to be a problem especially given the extended breathholds (image acquisition times of 5 to 10 s for 300 projections, corrupt image quality in up to 50% of cases116) leading to reduced diagnostic confidence. Using approaches similar to those for reconstruction of moving coronary vessels, reconstruction steps include: (1) reconstruction of a motion-blurred volume; (2) processing of the motion-blurred volume via thresholding, segmentation, landmark identification, etc.; (3) calculation, using a robust cost function, of a nonperiodic, smooth elastic 3D motion vector field describing the nonrigid deformation map between the projection images and the motion-blurred volume; and (4) re-reconstruction of the volume using the motion-corrected projection images. Iterate over steps 2 to 4 using an appropriate stopping criterion117120 (see Fig. 21 for an example of achievable image quality improvement). Excellent depiction of feeder vessels and tumor vasculature is especially important when image guidance tools such as virtual parenchymal perfusion (VPP) estimation121,122 are used during interventional procedures [see Fig. 21(c)].

Fig. 21

(a) Uncorrected and (b) motion-corrected FPCT reconstruction of an arterial injection of contrast in the liver vasculature including a hepatocellular tumor. Image courtesy of Armeen Mahvash, M.D., MD Anderson Cancer Center, Houston; prototype correction provided by Dr. Guenter Lauritsch, Siemens Healthcare GmbH and Dr. Christopher Rohkohl. (c) Combined use of automated tumor feeder detection and VPP in superselective conventional transcatheter arterial chemoembolization. The virtual embolized area, shown in orange, changed according to the catheter position (an orange marker). The colored area also extended to the right kidney in the axial, coronal, and sagittal directions. (c) Bottom right: common hepatic arteriogram showed a tumor stain (arrow). Reproduced with permission from Ref. 121.

JMI_8_5_052115_f021.png

4.4.

Summary of C-Arm FPCT

C-arm-based FPCT continues to evolve beyond the original application of 3D reconstruction of contrast-filled vascular structures (see Fig. 22). As DFP technology and the C-arm gantry hardware evolve, new applications such as multisweep acquisitions for quantitative perfusion imaging in the brain, and multiphase cardiac chamber reconstruction will enter routine clinical use. Although the underlying image quality of C-arm FPCT is approaching that of clinical CT (see Fig. 22), to date no C-arm manufacturer has obtained the FDA label of “computed tomography” for C-arm-based FPCT, preferring to use terms such as “CT-like” to describe the 3D reconstructed data. One can easily predict that as trajectory flexibility, DFP frame rate, noise floor, and stability improve, and new AI-based correction algorithms become standard, it will be possible to achieve the Hounsfield unit stability, uniformity, and reproducibility to merit the label.

Fig. 22

(a) Use of a dual-volume rendering technique to visualize a flow diverter across the neck of an aneurysm; (b) multiplanar reformat image showing the residual contrast inflow in the aneurysm through the flow diverter. Images courtesy David Niemann, MD, University of Wisconsin Hospitals and Clinics Authority. Multiplanar reformat images of a human brain (c) before and (d) after application of a noise-reduction algorithm. Image quality is sufficient to distinguish gray and white matter in the brain, a Hounsfield unit difference of only 10  HU.

JMI_8_5_052115_f022.png

5.

FPCT in Mammography

5.1.

Breast Cancer Imaging

The main motivation for CT imaging of the breast (see Fig. 23) is to eliminate the effect of tissue superposition. In mammography, non contrast-enhanced breast tumors are not differentiated from normal fibroglandular tissue in the breast due to differences in photon attenuation. Rather, x-ray breast imaging is a purely morphological imaging modality. That is, breast tumors are detected because their shape is different from that of normal breast tissue. Therefore, in a 2D modality such as mammography, the overlap of normal tissue projected on top of a pathologic lesion results in a loss of sensitivity.123 Furthermore, the overlap of various separate normal fibroglandular strands onto the same location may, randomly, form a shape that mimics a lesion of concern, resulting in a loss of specificity.124 To ameliorate this effect, other imaging modalities that decrease the superposition effect have been sought, such as stereoscopic mammography125 and digital breast tomosynthesis.126,127 Although the former also showed good performance, the latter has been widely introduced for every-day clinical breast imaging.128 However, even though tomosynthesis does provide some information in the third dimension, its resolution in this direction is still quite limited.

Fig. 23

Artist’s sketch of patient setup and tube-detector trajectory for a dedicated breast CT imaging system.

JMI_8_5_052115_f023.png

5.2.

Breast Imaging Requirements

Breast cancer detection and diagnosis is an especially demanding clinical application for imaging technology due to the need to adequately image two very different types of lesions: soft tissue-based and microcalcifications.130 The former usually presents as masses with or without spiculations, in addition to some other forms, such as architectural distortions. These lesion types are of low contrast and in the mm to cm scale (Fig. 24). Therefore, low energy x-ray spectra are needed to increase subject contrast, whereas spatial resolution is not at a premium (with the caveat that mass edge characterization and detection of spiculations, important for diagnosis, do benefit from high spatial resolution) compared with body imaging. However, the calcifications can be as small as 200  μm, resulting in the need for high spatial resolution. Their small size also makes calcification detection quantum-noise limited,131 resulting in the need for higher doses.

Fig. 24

(a) Mediolateral oblique view mammogram depicting a ductal invasive carcinoma as a suspicious mass in the lower part of the breast (arrow). (b) Axial, (c) sagittal, and (d) coronal cone-beam breast CT images with a highly suspicious mass (arrows). Reprinted from Wienbeck et al.129 with permission from Elsevier.

JMI_8_5_052115_f024.png

Therefore, a conebeam CT system for breast imaging needs to balance various competing requirements. The use of softer x-ray spectra to increase contrast, for improved mass visibility, results in reduced transmission, decreasing dose at the detector, increasing noise, and hence decreasing calcification visibility. To compensate for this, increases in tube current and/or x-ray pulse length are needed, but of course these increase the dose to the breast. Furthermore, increasing the pulse length also results in an increase in the effective focal spot size, reducing the spatial resolution of the system.

In addition, as opposed to all other flat-panel conebeam CT uses, the main application in breast imaging would ideally be screening of the general population. This, of course, makes radiation dose a real concern, realistically limiting it to a level similar to that currently used for 2D mammography-based screening.

Although the use of iodinated contrast enhancement in breast imaging results in substantially improved performance,132 its use in screening of the general population and for first-line diagnostic work-up is not clinically feasible. Therefore, the contrast demands of this clinical application cannot be ameliorated by the use of iodinated contrast.

On the other hand, except for specific cases, which right now are only in the research realm, breast imaging does not require fast acquisitions. In fact, conebeam breast CT acquisitions are currently 10 to 16 s long.133,134 In a study involving 40 patient conebeam breast CT acquisitions with a system with a 10-s acquisition time, only two of the cases (5%) were found to suffer of motion artifacts due to the relatively long acquisition time.133

5.3.

Flat-Panel Conebeam Breast CT Systems

The most obvious aspect that differentiates conebeam breast CT from the rest is its geometry. Strictly speaking, breast CT uses a half-cone beam geometry, with the central ray located at one edge of the x-ray beam, to graze the chest wall of the patient being imaged. To maximize tissue coverage, the size of the inactive area of the flat-panel detector edge on the chest wall side needs to be minimized.

This geometry also results in breast CT involving a maximum cone-beam angle that is larger than those of other applications. To improve on the resulting limitations in undersampling, alternative acquisition geometries have been investigated. In one breast CT implementation, a saddle trajectory was proposed, resulting in improved coverage and sampling of the frequency space.135 In more recent work, implementations involving multiple x-ray sources have been investigated using computer simulations.136 Other image acquisition trajectories, such a full revolution plus a linear scan, have been previously proposed.137 However, to date, all current flat-panel-based breast CT implementations that have reached the stage of patient imaging, either as commercial or research systems, use the traditional half-cone beam geometry with a single circular scan with one x-ray source (Fig. 25).

Fig. 25

Photo of a flat panel cone beam dedicated breast CT system showing the main components of the imaging chain. The plastic cup located surrounding the space for the pendant breast isolates the moving parts of the system from the breast and the rest of the patient.

JMI_8_5_052115_f025.png

Two types of DFPs have been used in flat-panel conebeam breast CT systems. The initial breast CT systems rely on a-Si with a columnar Csi:Tl scintillator-based TFT DFP. The usual size of these DFPs is 40  cm×30  cm, which are large enough, considering the magnification factors of breast CT systems, to allow for the imaging of a whole breast in one projection.133135,138,129 Some more recent system designs have incorporated the use of CMOS-based DFPs, which tend to be smaller in size, in the order of 29  cm×23  cm. The use of this new DFP technology has allowed for an increase in spatial resolution due to their smaller pixel pitch (Table 1).

Some early implementations of breast CT systems did not involve pulsed x-ray sources, of course impacting spatial resolution. However, current systems all use pulsed sources, which, together with the transition to CMOS-based detectors, have resulted in substantially improved spatial resolution in breast CT.139 As is shown by Gazi et al., increasing tube current does not seem to result in a substantial impact in the focal spot size.139 This is important since increasing the tube current allows for maintaining a short pulse length, crucial to minimizing the effective focal spot size in the direction of tube travel.

It should be noted that a completely different breast CT system design involves spiral imaging with an x-ray fan-beam and the use of a strip photon-counting detector.140 However, since this is not a cone-beam CT system, it will not be discussed further.

5.4.

Scatter Correction

Various different approaches for x-ray scatter correction have been proposed for flat-panel conebeam breast CT. Of course, one approach, which is, strictly speaking, not correcting for scatter, is to correct for the resulting cupping artifact postreconstruction.141,142 One method that has been proposed to estimate the scatter signal specific for the imaged breast is the beam pass array method (Fig. 26).143145 One important benefit of this method is that although it requires the acquisition of additional images of the breast, the presence of a beam pass array between the source and breast substantially reduces the dose of these extraprojections. The inverse of this approach, the beam stop array method, has also been proposed for breast CT.146 To limit the increase in dose involved due to the acquisition of extraprojections for the beam stop array method, only a few projections are acquired and the rest of the information is estimated via interpolation.

Fig. 26

(a) Schematic of the beam pass array method, which results in a sampling of the primary-only signal in the FPCB dedicated breast CT projection. (b) Correction of the acquired projections results in improved contrast and increased homogeneity throughout the reconstructed image. The projection-domain correction using information acquired for the specific imaged breast yields superior results compared to the image-domain correction performed by the system. Reprinted from Sechopoulos126 with permission from AAPM, and from Ramamurthy et al.143 © Institute of Physics and Engineering in Medicine, reproduced by permission of IOP Publishing, all rights reserved.

JMI_8_5_052115_f026.png

Other computational approaches, which do not require the acquisition of additional projections of the breast in question, have been also proposed.147,148 Currently, the single commercial FDA-approved conebeam breast CT system incorporates a cupping artifact correction postreconstruction only. Therefore, the commercial suitability of all the other methods that involve scatter correction is not yet clear.

5.5.

Contrast-Enhanced Breast Imaging

The use of iodinated contrast for breast cancer imaging allows for the interrogation of the vascular status of the breast and, specifically, that of suspicious lesions. The rapid angiogenesis promoted by signaling from the tumor to allow it to grow results in malformed, leaky, blood vessels. When iodinated contrast is injected intravenously, the iodinated blood mixture circulates and leaks out of these tumor vessels. Contrast-enhanced breast imaging, therefore, aims to image this extravasated iodinated blood, resulting in images in which malignancies have considerably higher contrast (Fig. 27).

Fig. 27

(a) Craniocaudal (CC) view mammogram showing suspicious clustered microcalcifications (arrow) in the inner part of the right breast. The density of the tumor is superimposed by normal breast tissue. (b) The microcalcification cluster is visible in the dedicated noncontrast cone-beam breast CT (arrow). (c) Addition of contrast for contrast-enhanced cone-beam breast CT shows a clearly delineated, intensely homogeneous contrast enhanced tumor mass (arrow). (d) Axial, (e) coronal, and (f) sagittal views of the subtraction images show the enhancement in the tumor mass, which was determined to be a ductal invasive carcinoma. Reprinted from Wienbeck et al.129 with permission from Elsevier.

JMI_8_5_052115_f027.png

The x-ray spectra used in conebeam breast CT systems, of considerably lower tube voltages than those in other CT applications, are ideally suited for contrast-enhanced imaging, given the closeness of the energy of the x-rays to the k-edge of iodine.149 Therefore, there are multiple studies that have shown the improved performance of contrast-enhanced imaging for this application.138,150,151

One particular advantage that makes breast CT amenable to contrast-enhanced imaging is that, as opposed to in mammography and tomosynthesis, no breast compression is used. Therefore, there is no restriction of blood flow into the breast during acquisition, as with those modalities. This makes image acquisition with short delay after injection and/or of multiple images postinjection much easier to implement.

5.6.

Dynamic Perfusion Breast Imaging

Due to this possibility of performing dynamic imaging, among other advantages, it has been proposed recently that flat-panel conebeam breast CT may be ideally suited for dynamic perfusion imaging. An optimized implementation of dynamic CT perfusion for breast cancer imaging would result not only in iodine-only images in which lesions are better depicted but also would result in the acquisition of enhanced functional information. Incorporation of high-speed added filtration switching,152 spectral reconstruction,153,154 and motion correction155 could yield dynamic images with very high combined spatiotemporal resolution. These additional capabilities could provide valuable information to, especially, determine tumor status and help predict tumor response to neoadjuvant therapy. As opposed to current flat-panel conebeam breast CT systems, such a system should be aimed to have a reduced acquisition time, in addition to the ability to acquire multiple images with minimal or no delay in between acquisitions.

6.

Dedicated Head and Extremities Flat-Panel CBCT

6.1.

Introduction to Dedicated Head and Extremities FPCT

Flat-panel-based CBCT for dedicated head and extremities systems was born out of contemporaneous developments in imaging for radiotherapy and interventional imaging (Fig. 28). Like interventional CBCT, the first 3D systems for dental and maxillofacial imaging were based on image intensifiers.8,156158,159Initial FPCT systems for dental applications were reported in the early 2000s by Sukovic et al.160 and Baba et al.161 These advances were quickly followed by a large number of successful commercial FPCT offerings. Farman and Scarf162 and Molteni163 each have a detailed history and summary of FPCT head imaging systems. The success and development of dedicated CBCT continue to broaden and extend to other applications including extremities imaging.

Fig. 28

Artist’s sketch of (a) a dedicated FPCT system for head imaging and (b) an extremity FPCT imaging system in the “weightbearing imaging” position.

JMI_8_5_052115_f028.png

One of the reasons for the success of these dedicated systems is the ability to customize the device to the anatomical and clinical target. For example, by focusing on a single clinical target (e.g., head or extremities) as opposed to general full-body CT, a number of optimizations can be made that can have a distinct advantage over traditional diagnostic CT.

6.2.

Customization of Flat-Panel CBCT for Head Imaging

There are a number of factors that have contributed to the success of flat-panel CT for dedicated head imaging. These factors were particularly well-aligned for dental imaging with dedicated CBCT being described by some as the “most significant advance in dental imaging since the introduction of rotational panoramic radiography.”162 The matching of the flat-panel technology to the particular clinical application and environment contributed to the early adoption of flat-panel CBCT for dental imaging. That success, in turn, has enabled other dedicated head imaging applications such as temporal bone and sinus imaging for otolaryngology.

6.2.1.

High-resolution, 3D capability

The high-resolution capability provided by flat-panel detectors is particularly important in dental, maxillofacial, and temporal bone imaging. Visualization of submillimeter anatomical features—e.g., in and around teeth in the mandible and maxilla, of the inner ear bones and semicircular canals, and air–tissue–bone boundaries in the paranasal sinuses—is critical for the diagnosis and treatment of a range of conditions. At the advent of flat-panel CT, traditional diagnostic multirow detector CT typically produced images with anisotropic resolution with a slice thickness greater than the axial, in-plane resolution. Flat-panel CBCT with its isotropic submillimeter resolution permitted not only better visualization of small features but also the ability to produce high-quality oblique slices and curved planar reformations to highlight specific anatomy (e.g., double oblique reformatting to visualize the stapes in the inner ear164 and curved cross-sections along the dental arch).165 In dental imaging, which was previously largely focused on 2D projection imaging, the capacity for 3D assessment and quantitation was a distinct advantage.

6.2.2.

Compact system design

Traditional multirow CT scanners are large devices with complex siting requirements including significant requirements for electrical power, structural and load-bearing capacity, and radiation shielding. These elements are an obstacle to point-of-care imaging in dental offices (where x-ray imaging was common but with much smaller devices) and private practice physicians’ offices (where imaging was often conducted out-of-office). By focusing on head imaging alone, a number of physical aspects of the device may be optimized leading to a compact design.

Flat-panel detectors are themselves are relatively thin and lightweight. Their large area permits wide coverage in a single circular orbit, covering the mandible and maxilla, paranasal sinuses, or the temporal bones relatively easily. There is much less variability in patient head size as compared with abdominal imaging, and the sizes are smaller permitting smaller source-to-detector distance in a dedicated system. Combined with the focus on relatively high-contrast applications (e.g., bone quality and structure, soft-tissue/bone interfaces in dental imaging, and soft-tissue/bone/air boundaries in sinus and inner ear imaging), there are also much lower power requirements on x-ray tube power. While care must be taken to balance focal spot size effects (important for high resolution scanning) with the short source-to-detector distance,166 the modest x-ray power needed generally allows for the use of compact “monoblock” designs. Such x-ray sources have all of the high voltage electronics within the same structure as the x-ray tube eliminating the need for bulky high-voltage cabling and reducing the relative power requirements. Combined with the modest rotation speeds that require smaller, lower power motors, most dedicated head scanners can be powered off standard electrical outlets. The form factor of head scanners is also optimized—often for scanning in a standing or seated position. This small footprint is a distinct advantage for placement at the point-of-care in a physician’s office. An example dedicated cone-beam CT head scanner is shown in Fig. 29.

Fig. 29

(a) Illustration of a dedicated head CT scanner for otolaryngology applications. The seated design is compact and requires only modest shielding requirements. (b) The high-contrast, high-spatial resolution is excellent for visualizing the temporal bone and inner ear structures. (c) Another sample image from a dedicated dental scanner showing interfaces between teeth and bone as well as position within the mandible and maxilla. Panels (a) and (b) are courtesy Xoran Technologies, Inc.; panel (c) is courtesy Yuxiang Xing, Tsinghua University.

JMI_8_5_052115_f029.png

6.2.3.

Low cost and low dose

Another important element of dedicated head scanners is the relatively low cost as compared with traditional multirow CT. Much of the cost reduction goes hand-in-hand with the compact design, modest gantry requirements (i.e., smaller x-ray tubes, and smaller motors), as well as the relatively inexpensive flat-panel detector. It is widely reported that flat-panel CBCT dose is lower than conventional CT.167169 This is, in part, by design with the focus on high-contrast features and low tube power, which limits the radiation exposures that are available. However, there are also geometric advantages as compared with diagnostic CT (e.g., no overshoot due to neighboring “slices” that one would find in a spiral acquisition). The relatively low exposures contribute additionally to lower cost of these systems with modest siting requirements including simpler shielding plans.

6.3.

Clinical Applications for CBCT Head Imaging

There is a plethora of clinical imaging applications reported in the literature.170 Initial application of dental scanners focused on oral surgery planning158 and implant planning171173 but has grown to include evaluations of implant failure.174 Critical elements of these studies include assessment of bone quality including height and width (e.g., of the mandible)175 and bone mineral density,176 3D assessment of alveolar ridge, identification of vital structures including the inferior alveolar nerve, and modeling for the fabrication of surgical guides.175 (Sample image is shown in Fig. 30.) Other applications include the evaluation of mandibular fractures,176 the temporomandibular joint,177,178 and tooth root fractures.179 Otolaryngological applications include assessments of inner ear disease and the temporal bone180 (see Fig. 30) including assessments of semicircular canal dehiscence,181 otosclerosis,182 and cochlear implant placement.183185 Dedicated head CBCT has also found application in diagnosing and gauging inflammatory sinusitis186,187 as well as in the planning and assessment of sinus188 and maxillofacial surgeries.189

Fig. 30

(a) Illustration of a dedicated extremities CBCT capable of bilateral leg scans. The weight bearing scan shows reduced joint space in a sample knee scan. (Images courtesy CurveBeam). (b) Another extremities CBCT that targets individual extremities with a short scan orbit and detector that passes between the legs. A sample ankle image shows the relation between bones and surfaces in this complex joint. (Images courtesy Wojciech Zbijewski, Johns Hopkins University.).

JMI_8_5_052115_f030.png

6.4.

Customization of Flat-Panel System for Extremities Imaging

Many of the customizations for head scanning apply to extremities scanning as well. In particular, the focus on high-resolution targets including bone structure, morphology, and interfaces is well-suited to flat-panel CBCT. Again, focusing on extremities targets that are smaller than general, full-body, multirow CT requirements permits compact designs and significant cost savings. Many dedicated extremities system integrate self-shielding into the scanner design with lead drapes, etc. to further reduce siting requirements. As with head scanning, dedicated extremities generally operate at lower dose190,191 than multirow CT, further reducing siting requirements.

6.4.1.

Weight-bearing imaging

One element that is distinct for extremities imaging is the customization of scanners for weight bearing192194 imaging. In contrast to multirow CT where patients are lying flat on a scanning table, dedicated extremities CBCT have been designed to permit scanning of standing patients. Imaging of patients in natural poses and of anatomy in the loaded state allows visualization of joint spaces, impingements, erosions, etc. under weight-bearing conditions.

Examples of two extremities scanners are shown in Figs. 29 and 30, where design choices have been made to scan either both legs or single legs, respectively. Note that accessibility is particularly important for patients with joint and mobility issues, and these systems are designed with additional arm supports.

6.5.

Clinical Applications for Extremities Imaging

Dedicated extremities CBCT have been used to diagnose a wide range of issues including arthrography,195 fracture assessment,196,197 and fracture healing.198,199 There are ongoing studies looking into the efficacy of dedicated systems for the evaluations of bone health including measures of bone mineral density200 and assessment of osteoporosis and osteoarthritis.201,202

6.6.

Point-of-Care Dedicated CBCT

One of the main advantages of dedicated CBCT is the ability to perform point-of-care scanning. This has the potential to streamline patient care and have diagnostic imaging, diagnosis, and treatment in one center. The low-cost and compact form factors of the CBCT systems allow these devices to be put in smaller institutions and individual private practices providing much convenience for the patient. However, such installations raise additional challenges including the potential for overuse through self-referral and the need for robust quality assurance. Trained and qualified medical physicists and engineers need to be part of a regular maintenance program to ensure proper quality control in image quality, radiation dose, protocols, shielding, etc.

6.7.

Future of Dedicated CBCT Imaging

Flat-panel CBCT is finding its way into other dedicated applications such as spine imaging,203 which overlaps somewhat with other interventional systems described previously in this paper. Some applications of flat-panel CBCT have been driven by advances in soft-tissue image quality. For example, through rigorous modeling of the many nonidealities that flat-panel systems present (including detector lag, detector glare, and scattered radiation), sophisticated correction schemes have been developed204 that permit good soft tissue visualization. Recent studies205 have shown that dedicated CBCT image quality can be sufficient for intracranial diagnosis and monitoring including hemorrhage and hydrocephalus. Mobile dedicated systems permit point-of-care deployment in critical/intensive care units and can potentially be brought to the patient bedside, eliminating the need to transport critically ill patients between departments in the hospital.

7.

Future Enabled by New DFP Developments

Flat-panel CT has become an integral component of diagnostic imaging, radiation therapy image guidance, and image-guided interventional therapy. As DFPs have evolved, so has the image content of the 3D images and therefore developments in DFP technology point to future opportunities in FPCT. A first area of innovation is the construction of dual-layer dual energy flat panels. Feasibility has been demonstrated using the DL DFP, which consists of a top layer with a 200-μm-thick CsI scintillator coupled to an amorphous silicon (aSi) DFP of 150-μm pixel size and a bottom layer with a 550-μm-thick CsI scintillator coupled to an identical aSi DFP. The two layers are separated by a 1-mm Cu filter to increase spectral separation. Applications such as dual-energy CBCT for contrast enhancement and material decomposition have been demonstrated. The merits of the DL detector include superior spatial and temporal registration between its constituent images and less complicated acquisition sequences as compared with a kV-switching approach.206,207 In the context of the Horizon 2020 EU funded activities, a consortium has come together to investigate the use of organic photodetectors (OPDs) as a transparent substrate, for use in multilayer indirect-conversion DFPs. Visible-light OPDs offer cost-effective fabrication methods using low temperature processes, making them particularly attractive for large area image detectors on lightweight flexible plastic substrates.208,209 Direct-conversion photon-counting energy discrimination detectors have also been investigated in the context of FPCT spectral imaging, and the results are promising.210212 First results from a hybrid aSi:H/direct-conversion strip detector have been evaluated in the context of interventional neurology, demonstrating the possibility for energy-based discrimination between intracranial hemorrhage, calcifications, and iodine staining from periprocedural contrast-enhanced imaging sequences.213

Disclosures

No conflicts of interest, financial or otherwise, are declared by the authors.

Acknowledgments

The authors would like to acknowledge the following individuals for assistance and information: Robert Bennett, Guenter Lauritsch, Michael Manhart, Annette Birkhold, Thomas Brunner, Andreas Maier, Bernard Sandkamp, Remy Klausz, Jeff Siewerdsen, John Wong, Ed Shapiro, and Rick Colbeth.

References

1. 

E. L. Ritman et al., “Computed tomographic imaging of the heart: the dynamic spatial reconstructor,” Radiol. Clin. North Am., 18 (3), 547 –555 (1980). Google Scholar

2. 

R. A. Robb, “The dynamic spatial reconstructor: an x-ray video-fluoroscopic CT scanner for dynamic volume imaging of moving organs,” IEEE Trans. Med. Imaging, 1 (1), 22 –33 (1982). https://doi.org/10.1109/TMI.1982.4307545 ITMID4 0278-0062 Google Scholar

3. 

R. A. Robb et al., “High-speed three-dimensional x-ray computed tomography: the dynamic spatial reconstructor,” Proc. IEEE, 71 (3), 308 –319 (1983). https://doi.org/10.1109/PROC.1983.12589 IEEPAD 0018-9219 Google Scholar

4. 

M. D. Silver et al., “CT imaging with an image intensifier: using a radiation therapy simulator as a CT scanner,” Proc. SPIE, 1443 (1991). https://doi.org/10.1117/12.43447 PSISDG 0277-786X Google Scholar

5. 

M. F. Kijewski, P. F. Judy and G. K. Svensson, “Image quality of an analog radiation therapy simulator-based tomographic scanner,” Med. Phys., 11 (4), 502 –507 (1984). https://doi.org/10.1118/1.595519 MPHYA6 0094-2405 Google Scholar

6. 

D. Saint-Felix et al., “In vivo evaluation of a new system for 3D computerized angiography,” Phys. Med. Biol., 39 (3), 583 –595 (1994). https://doi.org/10.1088/0031-9155/39/3/020 PHMBA7 0031-9155 Google Scholar

7. 

R. Ning and R. A. Kruger, “Image intensifier-based computed tomography volume scanner for angiography,” Acad. Radiol., 3 (4), 344 –350 (1996). https://doi.org/10.1016/S1076-6332(96)80255-8 Google Scholar

8. 

P. Mozzo et al., “A new volumetric CT machine for dental imaging based on the cone-beam technique: preliminary results,” Eur. Radiol., 8 (9), 1558 –1564 (1998). https://doi.org/10.1007/s003300050586 Google Scholar

9. 

B. A. Schueler and X. Hu, “Correction of image intensifier distortion for three-dimensional x-ray angiography,” Proc. SPIE, 2432 272 –279 (1995). https://doi.org/10.1117/12.208345 PSISDG 0277-786X Google Scholar

10. 

B. A. Schueler et al., “Three-dimensional vascular reconstruction with a clinical x-ray angiography system,” Acad. Radiol., 4 (10), 693 –699 (1997). https://doi.org/10.1016/S1076-6332(97)80141-9 Google Scholar

11. 

R. Koppe et al., 3D Reconstruction of Cerebral Vessel Malformations Based on Rotational Angiography (RA), Elsevier, Amsterdam, The Netherlands (1998). Google Scholar

12. 

R. Koppe et al., Digital Stereotaxy/Stereotactic Procedures with C-Arm Based Rotation-Angiography (RA), Elsevier, Amsterdam, The Netherlands (1996). Google Scholar

13. 

R. Fahrig et al., “Use of a C-arm system to generate true three-dimensional computed rotational angiograms: preliminary in vitro and in vivo results,” Am. J. Neuroradiol., 18 (8), 1507 –1514 (1997). Google Scholar

14. 

R. Baba et al., “Comparison of flat-panel detector and image-intensifier detector for cone-beam CT,” Comput. Med. Imaging Graph., 26 (3), 153 –158 (2002). https://doi.org/10.1016/S0895-6111(02)00008-3 Google Scholar

15. 

A. R. Cowen, A. G. Davies and M. U. Sivananthan, “The design and imaging characteristics of dynamic, solid-state, flat-panel x-ray image detectors for digital fluoroscopy and fluorography,” Clin. Radiol., 63 (10), 1073 –1085 (2008). https://doi.org/10.1016/j.crad.2008.06.002 Google Scholar

16. 

P. La Rivière and C. Crawford, “From EMI to AI: a brief history of commercial CT reconstruction algorithms,” J. Med. Imaging, 8 (5), 052111 (2021). https://doi.org/10.1117/1.JMI.8.5.052111 Google Scholar

17. 

D. W. Holdsworth, M. Drangova and A. Fenster, “A high-resolution XRII-based quantitative volume CT scanner,” Med. Phys., 20 (2 Pt. 1), 449 –462 (1993). https://doi.org/10.1118/1.597038 MPHYA6 0094-2405 Google Scholar

18. 

B. D. Smith, “Image reconstruction from cone-beam projections: necessary and sufficient conditions and reconstruction methods,” IEEE Trans. Med. Imaging, 4 (1), 14 –25 (1985). https://doi.org/10.1109/TMI.1985.4307689 ITMID4 0278-0062 Google Scholar

19. 

H. K. Tuy, “An inversion formula for cone-beam reconstruction,” SIAM J. Appl. Math., 43 (3), 546 –552 (1983). https://doi.org/10.1137/0143035 SMJMAP 0036-1399 Google Scholar

20. 

B. D. Smith, “Cone-beam tomography: recent advances and a tutorial,” Opt. Eng., 29 (5), 524 –534 (1990). https://doi.org/10.1117/12.55621 Google Scholar

21. 

H. Kudo and T. Saito, “Feasible cone beam scanning methods for exact reconstruction in three-dimensional tomography,” J. Opt. Soc. Am. A, 7 (12), 2169 –2183 (1990). https://doi.org/10.1364/JOSAA.7.002169 JOAOD6 0740-3232 Google Scholar

22. 

N. J. Pelc and D. A. Chesler, “Utilization of cross-plane rays for three-dimensional reconstruction by filtered back-projection,” J. Comput. Assist. Tomogr., 3 (3), 385 –395 (1979). https://doi.org/10.1097/00004728-197906000-00015 JCATD5 0363-8715 Google Scholar

23. 

L. A. Feldkamp, L. C. Davis and J. W. Kress, “Practical cone-beam algorithm,” J. Opt. Soc. Am. A, 1 (6), 612 –619 (1984). https://doi.org/10.1364/JOSAA.1.000612 JOAOD6 0740-3232 Google Scholar

24. 

P. Grangeat, “Analyse d’un système d’imagerie 3D par reconstruction à partir de radiographies X en géométrie conique,” ENST, Paris (1987). Google Scholar

25. 

P. Grangeat, “Mathematical framework of cone beam 3D reconstruction via the first derivative of the radon transform,” Lect. Notes Math., 1497 66 –97 (1991). https://doi.org/10.1007/BFb0084509 LNMAA2 0075-8434 Google Scholar

26. 

A. Aichert et al., “Epipolar consistency in transmission imaging,” IEEE Trans. Med. Imaging, 34 (11), 2205 –2219 (2015). https://doi.org/10.1109/TMI.2015.2426417 ITMID4 0278-0062 Google Scholar

27. 

M. Vassholz et al., “New x-ray tomography method based on the 3D radon transform compatible with anisotropic sources,” Phys. Rev. Lett., 116 (8), 088101 (2016). https://doi.org/10.1103/PhysRevLett.116.088101 PRLTAO 0031-9007 Google Scholar

28. 

M. Defrise and R. Clack, “A cone-beam reconstruction algorithm using shift-variant filtering and cone-beam backprojection,” IEEE Trans. Med. Imaging, 13 (1), 186 –195 (1994). https://doi.org/10.1109/42.276157 ITMID4 0278-0062 Google Scholar

29. 

X. Tang et al., ““On the data acquisition, image reconstruction, cone beam artifacts, and their suppression in axial MDCT and CBCT: a review,” Med. Phys., 45 (9), e761 –e782 (2018). https://doi.org/10.1002/mp.13095 MPHYA6 0094-2405 Google Scholar

30. 

D.L. Parker, “Optimal short scan convolution reconstruction for fan beam CT,” Med. Phys., 9 (2), 254 –257 (1982). https://doi.org/10.1118/1.595078 MPHYA6 0094-2405 Google Scholar

31. 

J. W. Stayman et al., “Task-driven source-detector trajectories in cone-beam computed tomography: I. theory and methods,” J. Med. Imaging, 6 (2), 025002 (2019). https://doi.org/10.1117/1.JMI.6.2.025002 Google Scholar

32. 

G. Wang et al., “Image reconstruction is a new Frontier of machine learning,” IEEE Trans. Med. Imaging, 37 (6), 1289 –1296 (2018). https://doi.org/10.1109/TMI.2018.2833635 ITMID4 0278-0062 Google Scholar

33. 

T. Würfl et al., “Deep learning computed tomography: learning projection-domain weights from image domain in limited angle problems,” IEEE Trans. Med. Imaging, 37 (6), 1454 –1463 (2018). https://doi.org/10.1109/TMI.2018.2833499 ITMID4 0278-0062 Google Scholar

34. 

A. K. Maier et al., “Learning with known operators reduces maximum error bounds,” Nat. Mach. Intell., 1 (8), 373 –380 (2019). https://doi.org/10.1038/s42256-019-0077-5 Google Scholar

35. 

M. Jordi et al., “Efficient high cone-angle artifact reduction in circular cone-beam CT using deep learning with geometry-aware dimension reduction,” Phys. Med. Biol., 66 (13), 5015 (2021). https://doi.org/10.1088/1361-6560/ac09a1 PHMBA7 0031-9155 Google Scholar

36. 

K. Xiao et al., “X-ray cone-beam computed tomography geometric artefact reduction based on a data-driven strategy,” Appl. Opt., 58 (17), 4771 –4780 (2019). https://doi.org/10.1364/AO.58.004771 APOPAI 0003-6935 Google Scholar

37. 

A. Datta, Z. Zhong and S. Motakef, “A new generation of direct x-ray detectors for medical and synchrotron imaging applications,” Sci. Rep., 10 (1), 20097 (2020). https://doi.org/10.1038/s41598-020-76647-5 SRCEC3 2045-2322 Google Scholar

38. 

J. Hsieh, O. E. Gurmen and K. F. King, “Recursive correction algorithm for detector decay characteristics in CT,” Proc. SPIE, 3977 298 –305 (2000). https://doi.org/10.1117/12.384505 PSISDG 0277-786X Google Scholar

39. 

N. Mail et al., “An empirical method for lag correction in cone-beam CT,” Med. Phys., 35 (11), 5187 –5196 (2008). https://doi.org/10.1118/1.2977759 MPHYA6 0094-2405 Google Scholar

40. 

M. Overdick, T. Solf and H. A. Wischmann, “Temporal artefacts in flat dynamic x-ray detectors,” Proc. SPIE, 4320 47 –58 (2001). https://doi.org/10.1117/12.430913 PSISDG 0277-786X Google Scholar

41. 

I. Mollov, C. Tognina and R. Colbeth, “Photodiode forward bias to reduce temporal effects in a-Si based flat panel detectors,” Proc. SPIE, 6913 69133S (2008). https://doi.org/10.1117/12.769951 PSISDG 0277-786X Google Scholar

42. 

J. Starman et al., “Parameter investigation and first results from a digital flat panel detector with forward bias capability,” Proc. SPIE, 6913 69130L (2008). https://doi.org/10.1117/12.772356 PSISDG 0277-786X Google Scholar

43. 

K. Yang, A. L. Kwan and J. M. Boone, “Computer modeling of the spatial resolution properties of a dedicated breast CT system,” Med. Phys., 34 (6), 2059 –2069 (2007). https://doi.org/10.1118/1.2737263 MPHYA6 0094-2405 Google Scholar

44. 

R. L. Weisfield et al., “High-performance amorphous silicon image sensor for x-ray diagnostic medical imaging applications,” Proc. SPIE, 3659 307 –317 (1999). https://doi.org/10.1117/12.349505 PSISDG 0277-786X Google Scholar

45. 

J. Starman et al., “Investigation into the optimal linear time-invariant lag correction for radar artifact removal,” Med. Phys., 38 (5), 2398 –2411 (2011). https://doi.org/10.1118/1.3574873 MPHYA6 0094-2405 Google Scholar

46. 

J. Starman et al., “A nonlinear lag correction algorithm for a-Si flat-panel x-ray detectors,” Med. Phys., 39 (10), 6035 –6047 (2012). https://doi.org/10.1118/1.4752087 MPHYA6 0094-2405 Google Scholar

47. 

P. Bartl et al., “Measurement and detailed simulation of the modulation transfer function of the Medipix2,” Nucl. Instrum. Methods Phys. Res., Sect. A, 591 (1), 314 –317 (2008). https://doi.org/10.1016/j.nima.2008.03.081 NIMAER 0168-9002 Google Scholar

48. 

H. Illers et al., “Measurement of the modulation transfer function of digital x-ray detectors with an opaque edge-test device,” Radiat. Prot. Dosim., 114 (1–3), 214 –219 (2005). https://doi.org/10.1093/rpd/nch506 RPDODE 0144-8420 Google Scholar

49. 

E. Shefer et al., “State of the art of CT detectors and sources: a literature review,” Curr. Radiol. Rep., 1 (1), 76 –91 (2013). https://doi.org/10.1007/s40134-012-0006-4 Google Scholar

50. 

P. G. Roos et al., “Multiple gain ranging readout method to extend the dynamic range of amorphous silicon flat panel imagers,” Proc. SPIE, 5368 139 –149 (2004). https://doi.org/10.1117/12.535471 PSISDG 0277-786X Google Scholar

51. 

B. Ohnesorge et al., “Efficient correction for CT image artifacts caused by objects extending outside the scan field of view,” Med. Phys., 27 (1), 39 –46 (2000). https://doi.org/10.1118/1.598855 MPHYA6 0094-2405 Google Scholar

52. 

Y. Pan, X. Zou and D. Xia, “Image reconstruction in peripheral and central regions-of-interest and data redundancy,” Med. Phys., 32 (3), 673 –684 (2005). https://doi.org/10.1118/1.1844171 MPHYA6 0094-2405 Google Scholar

53. 

G. Glover and N. Pelc, Method and Apparatus for Compensating CT Images for Truncated Projections, USPTO(1985). Google Scholar

54. 

J. Hsieh et al., “A novel reconstruction algorithm to extend the CT scan field-of-view,” Med. Phys., 31 (9), 2385 –2391 (2004). https://doi.org/10.1118/1.1776673 MPHYA6 0094-2405 Google Scholar

55. 

J. Starman et al., “Estimating 0th and 1st moments in C-arm CT data for extrapolating truncated projections,” Proc. SPIE, 5747 (2005). https://doi.org/10.1117/12.596041 PSISDG 0277-786X Google Scholar

56. 

K. J. Ruchala et al., “Methods for improving limited field-of-view radiotherapy reconstructions using imperfect a priori images,” Med. Phys., 29 (11), 2590 –2605 (2002). https://doi.org/10.1118/1.1513163 MPHYA6 0094-2405 Google Scholar

57. 

Y. Ye et al., “A general local reconstruction approach based on a truncated Hilbert transform,” Int. J. Biomed. Imaging, 2007 063634 (2007). https://doi.org/10.1155/2007/63634 Google Scholar

58. 

P. S. Cho, R. H. Johnson and T. W. Griffin, “Cone-beam CT for radiotherapy applications,” Phys. Med. Biol., 40 (11), 1863 –1883 (1995). https://doi.org/10.1088/0031-9155/40/11/007 PHMBA7 0031-9155 Google Scholar

59. 

P. S. Cho, A. D. Rudd and R. H. Johnson, “Cone-beam CT from width-truncated projections,” Comput. Med. Imaging Graph, 20 (1), 49 –57 (1996). https://doi.org/10.1016/0895-6111(96)00031-6 Google Scholar

60. 

G. Wang, “X-ray micro-CT with a displaced detector array,” Med. Phys., 29 (7), 1634 –1636 (2002). https://doi.org/10.1118/1.1489043 MPHYA6 0094-2405 Google Scholar

61. 

J. Bian et al., “Optimization-based image reconstruction from sparse-view data in offset-detector CBCT,” Phys. Med. Biol., 58 (2), 205 –230 (2013). https://doi.org/10.1088/0031-9155/58/2/205 PHMBA7 0031-9155 Google Scholar

62. 

D. A. Jaffray, D. Yan and J. W. Wong, “Managing geometric uncertainty in conformal intensity-modulated radiation therapy,” Semin. Radiat. Oncol., 9 (1), 4 –19 (1999). https://doi.org/10.1016/S1053-4296(99)80051-4 SRONEO Google Scholar

63. 

M. G. Herman et al., “Clinical use of electronic portal imaging: report of AAPM Radiation Therapy Committee Task Group 58,” Med. Phys., 28 (5), 712 –737 (2001). https://doi.org/10.1118/1.1368128 MPHYA6 0094-2405 Google Scholar

64. 

L. Feldkamp, L. Davis and J. Kress, “Practical cone-beam algorithm,” J. Opt. Soc. Am. A, 1 (6), 612 –619 (1984). https://doi.org/10.1364/JOSAA.1.000612 Google Scholar

65. 

L. E. Antonuk et al., “Large area, flat-panel, amorphous silicon imagers,” Proc. SPIE, 2432 216 (1995). https://doi.org/10.1117/12.208340 SPIECJ 0361-0748 Google Scholar

66. 

P. J. Biggs, M. Goitein and M. D. Russell, “A diagnostic x-ray field verification device for a 10 MV linear accelerator,” Int. J. Radiat. Oncol. Biol. Phys., 11 (3), 635 –643 (1985). https://doi.org/10.1016/0360-3016(85)90201-9 IOBPD3 0360-3016 Google Scholar

67. 

H. E. Johns and J. R. Cunningham, “A precision cobalt 60 unit for fixed field and rotation therapy,” Am. J. Roentgenol. Radium. Ther. Nucl. Med., 81 (1), 4 –12 (1959). AJROAM 0092-5381 Google Scholar

68. 

R. Sephton and J. Hagekyriakou, “A diagnostic-quality electronic portal imaging system,” Radiother. Oncol., 35 (3), 240 –247 (1995). https://doi.org/10.1016/0167-8140(95)01572-X RAONDT 0167-8140 Google Scholar

69. 

D. A. Jaffray and J. W. Wong, “Exploring ‘target of the day’ strategies for a medical linear accelerator with conebeam-CT scanning capability,” in Int. Conf. Use Comput. Radiother., (1997). Google Scholar

70. 

D. A. Jaffray et al., “A radiographic and tomographic imaging system integrated into a medical linear accelerator for localization of bone and soft-tissue targets,” Int. J. Radiat. Onco.l Biol. Phys., 45 (3), 773 –789 (1999). https://doi.org/10.1016/S0360-3016(99)00118-2 Google Scholar

71. 

J. H. Siewerdsen and D. A. Jaffray, “Cone-beam computed tomography with a flat-panel imager: effects of image lag,” Med. Phys., 26 (12), 2635 –2647 (1999). https://doi.org/10.1118/1.598803 MPHYA6 0094-2405 Google Scholar

72. 

J. H. Siewerdsen and D. A. Jaffray, “A ghost story: spatio-temporal response characteristics of an indirect-detection flat-panel imager,” Med. Phys., 26 (8), 1624 –1641 (1999). https://doi.org/10.1118/1.598657 MPHYA6 0094-2405 Google Scholar

73. 

D. A. Jaffray and J. H. Siewerdsen, “Cone-beam computed tomography with a flat-panel imager: initial performance characterization,” Med. Phys., 27 (6), 1311 –1323 (2000). https://doi.org/10.1118/1.599009 MPHYA6 0094-2405 Google Scholar

74. 

J. H. Siewerdsen et al., “Empirical and theoretical investigation of the noise performance of indirect detection, active matrix flat-panel imagers (AMFPIs) for diagnostic radiology,” Med. Phys., 24 (1), 71 –89 (1997). https://doi.org/10.1118/1.597919 MPHYA6 0094-2405 Google Scholar

75. 

M. B. Sharpe et al., “The stability of mechanical calibration for a kV cone beam computed tomography system integrated with linear accelerator,” Med. Phys., 33 (1), 136 –144 (2006). https://doi.org/10.1118/1.2143141 MPHYA6 0094-2405 Google Scholar

76. 

D. A. Jaffray et al., “Flat-panel cone-beam computed tomography for image-guided radiation therapy,” Int. J. Radiat. Oncol. Biol. Phys., 53 (5), 1337 –1349 (2002). https://doi.org/10.1016/S0360-3016(02)02884-5 IOBPD3 0360-3016 Google Scholar

77. 

E. J. Seppi et al., “Megavoltage cone-beam computed tomography using a high-efficiency image receptor,” Int. J. Radiat. Oncol. Biol. Phys., 55 (3), 793 –803 (2003). https://doi.org/10.1016/S0360-3016(02)04155-X IOBPD3 0360-3016 Google Scholar

78. 

M. A. Mosleh-Shirazi et al., “A cone-beam megavoltage CT scanner for treatment verification in conformal radiotherapy,” Radiother. Oncol., 48 (3), 319 –328 (1998). https://doi.org/10.1016/S0167-8140(98)00042-5 RAONDT 0167-8140 Google Scholar

79. 

B. A. Groh et al., “A performance comparison of flat-panel imager-based MV and kV cone-beam CT,” Med. Phys., 29 (6), 967 –975 (2002). https://doi.org/10.1118/1.1477234 MPHYA6 0094-2405 Google Scholar

80. 

S. Bartolac et al., “A local shift-variant Fourier model and experimental validation of circular cone-beam computed tomography artifacts,” Med. Phys., 36 (2), 500 –512 (2009). https://doi.org/10.1118/1.3062875 MPHYA6 0094-2405 Google Scholar

81. 

C. Coolens et al., “Implementation and characterization of a 320-slice volumetric CT scanner for simulation in radiation oncology,” Med. Phys., 36 (11), 5120 –5127 (2009). https://doi.org/10.1118/1.3246352 MPHYA6 0094-2405 Google Scholar

82. 

J. H. Siewerdsen and D. A. Jaffray, “Cone-beam computed tomography with a flat-panel imager: magnitude and effects of x-ray scatter,” Med. Phys., 28 (2), 220 –231 (2001). https://doi.org/10.1118/1.1339879 MPHYA6 0094-2405 Google Scholar

83. 

J. Hsieh, L. W. Goldman and J. B. Fowlkes, Image Artifacts, Causes, and Correction, in Medical CT and Ultrasound: Current Technology and Applications, 487 Advanced Medical Publishing, Madison, Wisconsin (1995). Google Scholar

84. 

J. H. Siewerdsen et al., “The influence of antiscatter grids on soft-tissue detectability in cone- beam computed tomography with flat-panel detectors,” Med. Phys., 31 (12), 3506 –3520 (2004). https://doi.org/10.1118/1.1819789 MPHYA6 0094-2405 Google Scholar

85. 

J. H. Siewerdsen et al., “A simple, direct method for x-ray scatter estimation and correction in digital radiography and cone-beam CT,” Med. Phys., 33 (1), 187 –197 (2006). https://doi.org/10.1118/1.2148916 MPHYA6 0094-2405 Google Scholar

86. 

G. Jarry et al., “Characterization of scattered radiation in kV CBCT images using Monte Carlo simulations,” Med. Phys., 33 (11), 4320 –4329 (2006). https://doi.org/10.1118/1.2358324 MPHYA6 0094-2405 Google Scholar

87. 

G. J. Bootsma, F. Verhaegen and D. A. Jaffray, “The effects of compensator and imaging geometry on the distribution of x-ray scatter in CBCT,” Med. Phys., 38 (2), 897 –914 (2011). https://doi.org/10.1118/1.3539575 MPHYA6 0094-2405 Google Scholar

88. 

G. J. Bootsma, F. Verhaegen and D. A. Jaffray, “Spatial frequency spectrum of the x-ray scatter distribution in CBCT projections,” Med. Phys., 40 (11), 111901 (2013). https://doi.org/10.1118/1.4822484 MPHYA6 0094-2405 Google Scholar

89. 

G. J. Bootsma, F. Verhaegen and D. A. Jaffray, “Efficient scatter distribution estimation and correction in CBCT using concurrent Monte Carlo fitting,” Med. Phys., 42 (1), 54 –68 (2015). https://doi.org/10.1118/1.4903260 MPHYA6 0094-2405 Google Scholar

90. 

S. Bartolac et al., “Fluence field optimization for noise and dose objectives in CT,” Med. Phys., 38 (Suppl. 1), S2 (2011). https://doi.org/10.1118/1.3574885 MPHYA6 0094-2405 Google Scholar

91. 

S. Bartolac and D. Jaffray, “Compensator models for fluence field modulated computed tomography,” Med. Phys., 40 (12), 121909 (2013). https://doi.org/10.1118/1.4829513 MPHYA6 0094-2405 Google Scholar

92. 

M. H. Smitsmans et al., “Automatic localization of the prostate for on-line or off-line image- guided radiotherapy,” Int. J. Radiat. Oncol. Biol. Phys., 60 (2), 623 –635 (2004). https://doi.org/10.1016/j.ijrobp.2004.05.027 Google Scholar

93. 

J. J. Sonke et al., “Respiratory correlated cone beam CT,” Med. Phys., 32 (4), 1176 –1186 (2005). https://doi.org/10.1118/1.1869074 MPHYA6 0094-2405 Google Scholar

94. 

R. Fahrig, M. Moreau and D.W. Holdsworth, “Three-dimensional computed tomographic reconstruction using a C-arm mounted XRII: correction of image intensifier distortion,” Med. Phys., 24 (7), 1097 –1106 (1997). https://doi.org/10.1118/1.598013 MPHYA6 0094-2405 Google Scholar

95. 

O. D. Faugeras, Three-Dimensional Computer Vision: A Geometric Viewpoint, MIT Press, CambridgeMassachusetts (1993). Google Scholar

96. 

R. Fahrig and D. W. Holdsworth, “Three-dimensional computed tomographic reconstruction using a C-arm mounted XRII: Image-based correction of gantry motion nonidealities,” Med. Phys., 27 (1), 30 –38 (2000). https://doi.org/10.1118/1.598854 MPHYA6 0094-2405 Google Scholar

97. 

A. Fieselmann et al., “A model for filtered backprojection reconstruction artifacts due to time-varying attenuation values in perfusion C-arm CT,” Phys. Med. Biol., 56 (12), 3701 –3717 (2011). https://doi.org/10.1088/0031-9155/56/12/016 PHMBA7 0031-9155 Google Scholar

98. 

M. Grass et al., Improved Flow Reconstruction in 3D Rotational Angiography (3DRA), IEEE, Piscataway, New Jersey (2003). Google Scholar

99. 

H. Schmitt et al., “An x-ray-based method for the determination of the contrast agent propagation in 3-D vessel structures,” IEEE Trans. Med. Imaging, 21 (3), 251 –262 (2002). https://doi.org/10.1109/42.996343 ITMID4 0278-0062 Google Scholar

100. 

H. Schmitt et al., “Reconstruction of blood propagation in three-dimensional rotational x-ray angiography (3D-RA),” Comput. Med. Imaging Graph, 29 (7), 507 –520 (2005). https://doi.org/10.1016/j.compmedimag.2005.03.006 Google Scholar

101. 

B. Davis et al., “4D digital subtraction angiography: implementation and demonstration of feasibility,” Am. J. Neuroradiol., 34 (10), 1914 –1921 (2013). https://doi.org/10.3174/ajnr.A3529 Google Scholar

102. 

K. L. Ruedinger et al., “4D-DSA: development and current neurovascular applications,” Am. J. Neuroradiol., 42 (2), 214 –220 (2021). https://doi.org/10.3174/ajnr.A6860 Google Scholar

103. 

N. Huizinga et al., “4D flat panel conebeam CTA for in vivo imaging of the microvasculature of the human cortex with a novel software prototype,” Am. J. Neuroradiol., 41 (6), 976 –979 (2020). https://doi.org/10.3174/ajnr.A6574 Google Scholar

104. 

K. K. Chen et al., “Application of time-resolved 3D digital subtraction angiography to plan cerebral arteriovenous malformation radiosurgery,” Am. J. Neuroradiol., 38 (4), 740 –746 (2017). https://doi.org/10.3174/ajnr.A5074 Google Scholar

105. 

G. Shaughnessy et al., “Measuring blood velocity using 4D-DSA: a feasibility study,” Med. Phys., 45 (10), 4510 –4518 (2018). https://doi.org/10.1002/mp.13120 MPHYA6 0094-2405 Google Scholar

106. 

Y. Wu et al., “Quantification of blood velocity with 4D digital subtraction angiography using the shifted least-squares method,” Am. J. Neuroradiol., 39 (10), 1871 –1877 (2018). https://doi.org/10.3174/ajnr.A5793 Google Scholar

107. 

N. Kato et al., “4D digital subtraction angiography for the temporal flow visualization of intracranial aneurysms and vascular malformations,” J. Stroke Cerebrovasc. Dis., 29 (12), 105327 (2020). https://doi.org/10.1016/j.jstrokecerebrovasdis.2020.105327 Google Scholar

108. 

M. Berger et al., “Motion compensation for cone-beam CT using Fourier consistency conditions,” Phys. Med. Biol., 62 (17), 7181 –7215 (2017). https://doi.org/10.1088/1361-6560/aa8129 PHMBA7 0031-9155 Google Scholar

109. 

R. Frysch and G. Rose, “Rigid motion compensation in interventional C-arm CT using consistency measure on projection data,” Lect. Notes Comp. Sci., 9349 298 –306 (2015). https://doi.org/10.1007/978-3-319-24553-9_37 Google Scholar

110. 

A. Thiagalingam et al., “Intraprocedural volume imaging of the left atrium and pulmonary veins with rotational x-ray angiography: implications for catheter ablation of atrial fibrillation,” J. Cardiovasc. Electrophysiol., 19 (3), 293 –300 (2008). https://doi.org/10.1111/j.1540-8167.2007.01013.x JCELE2 1045-3873 Google Scholar

111. 

J. Gerds-Li et al., “Rapid ventricular pacing to optimize rotational angiography in atrial fibrillation ablation,” J. Interventional Cardiac Electrophysiol., 26 (2), 101 –107 (2009). https://doi.org/10.1007/s10840-009-9397-4 Google Scholar

112. 

C. Kriatselis et al., “A new approach for contrast-enhanced x-ray imaging of the left atrium and pulmonary veins for atrial fibrillation ablation: rotational angiography during adenosine-induced asystole,” Europace, 11 (1), 35 –41 (2009). https://doi.org/10.1093/europace/eun311 Google Scholar

113. 

M. Tang et al., “Reconstructing and registering three-dimensional rotational angiogram of left atrium during ablation of atrial fibrillation,” Pacing Clin. Electrophysiol., 32 (11), 1407 –1416 (2009). https://doi.org/10.1111/j.1540-8159.2009.02481.x Google Scholar

114. 

S. Cimen et al., “Reconstruction of coronary arteries from x-ray angiography: a review,” Med. Image Anal., 32 46 –68 (2016). https://doi.org/10.1016/j.media.2016.02.007 Google Scholar

115. 

L. Pung et al., “The role of cone-beam CT in transcatheter arterial chemoembolization for hepatocellular carcinoma: a systematic review and meta-analysis,” J. Vasc. Interv. Radiol., 28 (3), 334 –341 (2017). https://doi.org/10.1016/j.jvir.2016.11.037 JVIRE3 1051-0443 Google Scholar

116. 

I. J. Lee et al., “Cone-beam CT hepatic arteriography in chemoembolization for hepatocellular carcinoma: angiographic image quality and its determining factors,” J. Vasc. Interv. Radiol., 25 1369 –1379 (2014). https://doi.org/10.1016/j.jvir.2014.04.011 Google Scholar

117. 

A. Sindel et al., “Respiratory motion compensation for C-arm CT liver imaging,” Bildverarbeitung für die Medizin, 221 –226 Springer, Berlin, Heidelberg (2015). Google Scholar

118. 

A. Klugmann et al., “Deformable respiratory motion correction for hepatic rotational angiography,” Comput. Med. Imaging Graph, 66 82 –89 (2018). https://doi.org/10.1016/j.compmedimag.2018.03.003 Google Scholar

119. 

M. D. Burgio et al., “Clinical impact of a new cone beam CT angiography respiratory motion artifact reduction algorithm during hepatic intra-arterial interventions,” Eur. Radiol., 30 (1), 163 –174 (2020). https://doi.org/10.1007/s00330-019-06355-w Google Scholar

120. 

L. S. Becker et al., “Evaluation of a motion correction algorithm for C-arm computed tomography acquired during transarterial chemoembolization,” Cardiovasc. Intervent. Radiol., 44 (4), 610 –618 (2021). https://doi.org/10.1007/s00270-020-02729-6 CAIRDG 1432-086X Google Scholar

121. 

S. Miyayama et al., “Performance of novel virtual parenchymal perfusion software visualizing embolized areas of transcatheter arterial chemoembolization for hepatocellular carcinoma,” Hepatol. Res., 47 (5), 446 –454 (2017). https://doi.org/10.1111/hepr.12766 Google Scholar

122. 

D. Lesage et al., “Virtual parenchymal perfusion for selective intra-arterial therapy of liver cancer,” in IEEE 13th Int. Symp. Biomed. Imaging, (2016). https://doi.org/10.1109/ISBI.2016.7493469 Google Scholar

123. 

R. L. Birdwell et al., “Mammographic characteristics of 115 missed cancers later detected with screening mammography and the potential utility of computer-aided detection,” Radiology, 219 (1), 192 –202 (2001). https://doi.org/10.1148/radiology.219.1.r01ap16192 RADLAX 0033-8419 Google Scholar

124. 

K. Holland et al., “Influence of breast compression pressure on the performance of population-based mammography screening,” Breast Cancer Res., 19 (1), 126 (2017). https://doi.org/10.1186/s13058-017-0917-3 BCTRD6 Google Scholar

125. 

C. J. D’Orsi et al., “Stereoscopic digital mammography: improved specificity and reduced rate of recall in a prospective clinical trial,” Radiology, 266 (1), 81 –88 (2013). https://doi.org/10.1148/radiol.12120382 RADLAX 0033-8419 Google Scholar

126. 

I. Sechopoulos, “A review of breast tomosynthesis. Part I. The image acquisition process,” Med. Phys., 40 (1), 014301 (2013). https://doi.org/10.1118/1.4770279 MPHYA6 0094-2405 Google Scholar

127. 

I. Sechopoulos, “A review of breast tomosynthesis. Part II. Image reconstruction, processing and analysis, and advanced applications,” Med. Phys., 40 (1), 014302 (2013). https://doi.org/10.1118/1.4770281 MPHYA6 0094-2405 Google Scholar

128. 

S. M. Friedewald et al., “Breast cancer screening using tomosynthesis in combination with digital mammography,” JAMA, 311 (24), 2499 (2014). https://doi.org/10.1001/jama.2014.6095 JAMAAP 0098-7484 Google Scholar

129. 

S. Wienbeck, J. Lotz and U. Fischer, “Review of clinical studies and first clinical experiences with a commercially available cone-beam breast CT in Europe,” Clin. Imaging, 42 50 –59 (2017). https://doi.org/10.1016/j.clinimag.2016.11.011 CLIMEB 0899-7071 Google Scholar

130. 

D. B. Kopans, Breast Imaging, 1114 3rd ed.Lippincott Williams & Wilkins, Baltimore, Maryland (2007). Google Scholar

131. 

A. E. Burgess, F. L. Jacobson and P. F. Judy, “Human observer detection experiments with mammograms and power-law noise,” Med. Phys., 28 (4), 419 –437 (2001). https://doi.org/10.1118/1.1355308 MPHYA6 0094-2405 Google Scholar

132. 

E. M. Fallenberg et al., “Contrast-enhanced spectral mammography vs. mammography and MRI—clinical performance in a multi-reader evaluation,” Eur. Radiol., 27 (7), 2752 –2764 (2017). https://doi.org/10.1007/s00330-016-4650-6 Google Scholar

133. 

A. O’Connell and D. Kawakyu-O’Connor, “Dedicated cone-beam breast computed tomography and diagnostic mammography: comparison of radiation dose, patient comfort, and qualitative review of imaging findings in BI-RADS 4 and 5 lesions,” J. Clin. Imaging Sci., 2 (1), 7 (2012). https://doi.org/10.4103/2156-7514.93274 Google Scholar

134. 

K. K. Lindfors et al., “Dedicated breast CT: initial clinical experience,” Radiology, 246 (3), 725 –733 (2008). https://doi.org/10.1148/radiol.2463070410 RADLAX 0033-8419 Google Scholar

135. 

M. P. Tornai et al., “Design and development of a fully 3D dedicated x-ray computed mammotomography system,” Proc. SPIE, 5745 189 –197 (2005). https://doi.org/10.1117/12.595636 PSISDG 0277-786X Google Scholar

136. 

A. E. Becker et al., “Cone beam CT multisource configurations: evaluating image quality, scatter, and dose using phantom imaging and Monte Carlo simulations,” Phys. Med. Biol., 65 (23), 235032 (2020). https://doi.org/10.1088/1361-6560/abc306 PHMBA7 0031-9155 Google Scholar

137. 

S. Vedantham et al., “Dedicated breast CT: radiation dose for circle-plus-line trajectory,” Med. Phys., 39 (3), 1530 –1541 (2012). https://doi.org/10.1118/1.3688197 MPHYA6 0094-2405 Google Scholar

138. 

S. Wienbeck et al., “Contrast-enhanced cone-beam breast-CT (CBBCT): clinical performance compared to mammography and MRI,” Eur. Radiol., 28 (9), 3731 –3741 (2018). https://doi.org/10.1007/s00330-018-5376-4 Google Scholar

139. 

P. M. Gazi et al., “Evolution of spatial resolution in breast CT at UC Davis,” Med. Phys., 42 (4), 1973 –1981 (2015). https://doi.org/10.1118/1.4915079 MPHYA6 0094-2405 Google Scholar

140. 

N. Berger et al., “Dedicated breast computed tomography with a photon-counting detector: initial results of clinical in vivo imaging,” Invest. Radiol., 54 (7), 409 –418 (2019). https://doi.org/10.1097/RLI.0000000000000552 INVRAV 0020-9996 Google Scholar

141. 

M. C. Altunbas et al., “A post-reconstruction method to correct cupping artifacts in cone beam breast computed tomography,” Med. Phys., 34 (7), 3109 –3118 (2007). https://doi.org/10.1118/1.2748106 MPHYA6 0094-2405 Google Scholar

142. 

X. Yang et al., “Cupping artifact correction and automated classification for high-resolution dedicated breast CT images,” Med. Phys., 39 (10), 6397 –6406 (2012). https://doi.org/10.1118/1.4754654 MPHYA6 0094-2405 Google Scholar

143. 

S. Ramamurthy, C. J. D’Orsi and I. Sechopoulos, “X-ray scatter correction method for dedicated breast computed tomography: improvements and initial patient testing,” Phys. Med. Biol., 61 (3), 1116 –1135 (2016). https://doi.org/10.1088/0031-9155/61/3/1116 PHMBA7 0031-9155 Google Scholar

144. 

I. Sechopoulos, “X-ray scatter correction method for dedicated breast computed tomography,” Med. Phys., 39 (5), 2896 –2903 (2012). https://doi.org/10.1118/1.4711749 MPHYA6 0094-2405 Google Scholar

145. 

K. Yang, J. G. Burkett and J. M. Boone, “A breast-specific, negligible-dose scatter correction technique for dedicated cone-beam breast CT: a physics-based approach to improve Hounsfield Unit accuracy,” Phys. Med. Biol., 59 (21), 6487 –6505 (2014). https://doi.org/10.1088/0031-9155/59/21/6487 PHMBA7 0031-9155 Google Scholar

146. 

R. Ning, X. Tang and D. L. Conover, “X-ray scatter suppression algorithm for cone-beam volume CT,” Proc. SPIE, 4682 774 –781 (2002). https://doi.org/10.1117/12.465628 PSISDG 0277-786X Google Scholar

147. 

L. Shi et al., “Library based x-ray scatter correction for dedicated cone beam breast CT,” Med. Phys., 43 (8 Pt. 1), 4529 –4544 (2016). https://doi.org/10.1118/1.4955121 MPHYA6 0094-2405 Google Scholar

148. 

L. Shi et al., “X-ray scatter correction for dedicated cone beam breast CT using a forward- projection model,” Med. Phys., 44 (6), 2312 –2320 (2017). https://doi.org/10.1002/mp.12213 MPHYA6 0094-2405 Google Scholar

149. 

M. J. Berger et al., “XCOM: Photon Cross Section Database (version 1.3),” (2005). http://physics.nist.gov/xcom Google Scholar

150. 

N. D. Prionas et al., “Contrast-enhanced dedicated breast CT: initial clinical experience,” Radiology, 256 (3), 714 –723 (2010). https://doi.org/10.1148/radiol.10092311 Google Scholar

151. 

J. Uhlig et al., “Contrast-enhanced cone-beam breast-CT: analysis of optimal acquisition time for discrimination of breast lesion malignancy,” Eur. J. Radiol., 99 9 –16 (2018). https://doi.org/10.1016/j.ejrad.2017.12.003 Google Scholar

152. 

K. Michielsen et al., “Dose reduction in breast CT by spectrum switching,” Proc. SPIE, 10718 107180J (2018). https://doi.org/10.1117/12.2318425 PSISDG 0277-786X Google Scholar

153. 

J. Chung, J. G. Nagy and I. Sechopoulos, “Numerical algorithms for polyenergetic digital breast tomosynthesis reconstruction,” SIAM J. Imaging Sci., 3 (1), 133 –152 (2010). https://doi.org/10.1137/090749633 Google Scholar

154. 

V. M. Bustamante et al., “Iterative breast tomosynthesis image reconstruction,” SIAM J. Sci. Comput., 35 (5), S192 –S208 (2013). https://doi.org/10.1137/120881440 SJOCE3 1064-8275 Google Scholar

155. 

J. Hahn et al., “Motion compensation in the region of the coronary arteries based on partial angle reconstructions from short-scan CT data,” Med. Phys., 44 (11), 5795 –5813 (2017). https://doi.org/10.1002/mp.12514 MPHYA6 0094-2405 Google Scholar

156. 

Y. Arai et al., “Development of a compact computed tomographic apparatus for dental use,” Dentomaxillofac. Radiol., 28 (4), 245 –248 (1999). https://doi.org/10.1038/sj.dmfr.4600448 Google Scholar

157. 

K. Maki et al., “Application of cone-beam x-ray CT in dento-maxillofacial region,” CARS 2002 Computer Assisted Radiology and Surgery, 1003 –1008 Springer(2002). Google Scholar

158. 

Y. Nakagawa et al., “Preoperative application of limited cone beam computerized tomography as an assessment tool before minor oral surgery,” Int. J. Oral Maxillofac. Surg., 31 (3), 322 –326 (2002). https://doi.org/10.1054/ijom.2001.0222 Google Scholar

159. 

K. Yamamoto et al., “Development of dento-maxillofacial cone beam x-ray computed tomography system,” Orthod. Craniofacial Res., 6 160 –162 (2003). Google Scholar

160. 

P. Sukovic et al., “DentoCAT™—a novel design of a cone-beam CT scanner for dentomaxillofacial imaging: introduction and preliminary results,” in Int. Congress Series, 700 –705 (2001). Google Scholar

161. 

R. Baba, K. Ueda and M. Okabe, “Using a flat-panel detector in high resolution cone beam CT for dental imaging,” Dentomaxillofac. Radiol., 33 (5), 285 –290 (2004). https://doi.org/10.1259/dmfr/87440549 Google Scholar

162. 

A. G. Farman, W. C. Scarfe, “Historical perspectives on CBCT,” Maxillofacial Cone Beam Computed Tomography, 3 –11 Springer, Cham (2018). Google Scholar

163. 

R. Molteni, “The way we were (and how we got here): fifty years of technology changes in dental and maxillofacial radiology,” Dentomaxillofac. Radiol., 50 (1), 20200133 (2020). https://doi.org/10.1259/dmfr.20200133 Google Scholar

164. 

J. W. Casselman et al., “Cone beam CT: non-dental applications,” J. Belgian Soc. Radiol., 96 (6), 333 –353 (2013). https://doi.org/10.5334/jbr-btr.453 Google Scholar

165. 

W. C. Scarfe, A. G. Farman and P. Sukovic, “Clinical applications of cone-beam computed tomography in dental practice,” J. Can. Dent. Assoc., 72 (1), 75 –80 (2006). JCDAAS 0008-3372 Google Scholar

166. 

R. Pauwels et al., “Technical aspects of dental CBCT: state of the art,” Dentomaxillofacial Radiol., 44 20140224 (2015). Google Scholar

167. 

M. Loubele et al., “Comparison between effective radiation dose of CBCT and MSCT scanners for dentomaxillofacial applications,” Eur. J. Radiol., 71 (3), 461 –468 (2009). https://doi.org/10.1016/j.ejrad.2008.06.002 EJRADR 0720-048X Google Scholar

168. 

K. Tsiklakis et al., “Dose reduction in maxillofacial imaging using low dose Cone Beam CT,” Eur. J. Radiol., 56 (3), 413 –417 (2005). https://doi.org/10.1016/j.ejrad.2005.05.011 EJRADR 0720-048X Google Scholar

169. 

J. B. Ludlow and M. Ivanovic, “Comparative dosimetry of dental CBCT devices and 64-slice CT for oral and maxillofacial radiology,” Oral Surg. Oral Med. Oral Pathol. Oral Radiol. Endodontol., 106 (1), 106 –114 (2008). https://doi.org/10.1016/j.tripleo.2008.03.018 Google Scholar

170. 

W. De Vos, J. Casselman and G. R. J. Swennen, “Cone-beam computerized tomography (CBCT) imaging of the oral and maxillofacial region: a systematic review of the literature,” Int. J. Oral Maxillofac. Surg., 38 (6), 609 –625 (2009). https://doi.org/10.1016/j.ijom.2009.02.028 Google Scholar

171. 

P. Sukovic, “Cone beam computed tomography in craniofacial imaging,” Orthod. Craniofac. Res., 6 (Suppl. 1), 31 –36 (2003). https://doi.org/10.1034/j.1600-0544.2003.259.x Google Scholar

172. 

H.-L. Chan, K. Misch and H.-L. Wang, “Dental imaging in implant treatment planning,” Implant Dent., 19 (4), 288 –298 (2010). https://doi.org/10.1097/ID.0b013e3181e59ebd Google Scholar

173. 

S. Sato et al., “Clinical application of a new cone-beam computerized tomography system to assess multiple two-dimensional images for the preoperative treatment planning of maxillary implants: case reports,” Quintessence Int., 35 (7), 525 –528 (2004). Google Scholar

174. 

J. F. Yepes and M. Al-Sabbagh, “Use of cone-beam computed tomography in early detection of implant failure,” Dent. Clin. North Am., 59 (1), 41 –56 (2015). https://doi.org/10.1016/j.cden.2014.09.003 DCNAAC 0011-8532 Google Scholar

175. 

E. Benavides et al., “Use of cone beam computed tomography in implant dentistry,” Implant Dent., 21 (2), 78 –86 (2012). https://doi.org/10.1097/ID.0b013e31824885b5 Google Scholar

176. 

D.-G. Kim, “Can dental cone beam computed tomography assess bone mineral density?,” J. Bone Metab., 21 (2), 117 (2014). https://doi.org/10.11005/jbm.2014.21.2.117 Google Scholar

177. 

K. Tsiklakis, K. Syriopoulos and H. C. Stamatakis, “Radiographic examination of the temporomandibular joint using cone beam computed tomography,” Dentomaxillofac. Radiol., 33 (3), 196 –201 (2004). https://doi.org/10.1259/dmfr/27403192 Google Scholar

178. 

M. L. P. dos Anjos et al., “Evaluation of bone changes in the temporomandibular joint using cone beam CT,” Dentomaxillofac. Radiol., 41 (1), 24 –29 (2012). https://doi.org/10.1259/dmfr/17815139 Google Scholar

179. 

C. Hannig et al., “Three-dimensional, non-destructive visualization of vertical root fractures using flat panel volume detector computer tomography: an ex vivo in vitro case report,” Int. Endod. J., 38 (12), 904 –913 (2005). https://doi.org/10.1111/j.1365-2591.2005.01033.x Google Scholar

180. 

M. Dahmani-Causse et al., “Morphologic examination of the temporal bone by cone beam computed tomography: comparison with multislice helical computed tomography,” Eur. Ann. Otorhinolaryngol. Head Neck Dis., 128 (5), 230 –235 (2011). https://doi.org/10.1016/j.anorl.2011.02.016 Google Scholar

181. 

K. Eibenberger et al., “A novel method of 3D image analysis of high-resolution cone beam CT and multi slice CT for the detection of semicircular canal dehiscence,” Otol. Neurotol., 35 (2), 329 –337 (2014). https://doi.org/10.1097/MAO.0000000000000199 Google Scholar

182. 

P. Révész et al., “Comparative analysis of preoperative diagnostic values of HRCT and CBCT in patients with histologically diagnosed otosclerotic stapes footplates,” Eur. Arch. Otorhinolaryngol., 273 (1), 63 –72 (2016). https://doi.org/10.1007/s00405-015-3490-3 Google Scholar

183. 

T. A. Kennedy et al., “Flat-panel CT for cochlear implant electrode imaging,” Otol. Neurotol., 37 (10), 1646 –1653 (2016). https://doi.org/10.1097/MAO.0000000000001216 Google Scholar

184. 

J. Zou et al., “Imaging cochlear implantation with round window insertion in human temporal bones and cochlear morphological variation using high-resolution cone beam CT,” Acta Otolaryngol., 135 (5), 466 –472 (2015). https://doi.org/10.3109/00016489.2014.993090 Google Scholar

185. 

S. R. Saeed et al., “The use of cone-beam computed tomography to determine cochlear implant electrode position in human temporal bones,” Otol. Neurotol., 35 (8), 1338 –1344 (2014). https://doi.org/10.1097/MAO.0000000000000295 Google Scholar

186. 

R. A. Zoumalan et al., “Flat panel cone beam computed tomography of the sinuses,” Otolaryngol. Head Neck Surg., 140 (6), 841 –844 (2009). https://doi.org/10.1016/j.otohns.2009.01.010 Google Scholar

187. 

M. Maillet et al., “Cone-beam computed tomography evaluation of maxillary sinusitis,” J. Endod., 37 (6), 753 –757 (2011). https://doi.org/10.1016/j.joen.2011.02.032 Google Scholar

188. 

L. F. P. Nicolielo et al., “Presurgical CBCT assessment of maxillary neurovascularization in relation to maxillary sinus augmentation procedures and posterior implant placement,” Surg. Radiol. Anat., 36 (9), 915 –924 (2014). https://doi.org/10.1007/s00276-014-1309-3 Google Scholar

189. 

M. Ahmad, J. Jenny and M. Downie, “Application of cone beam computed tomography in oral and maxillofacial surgery,” Aust. Dent. J., 57 (1), 82 –94 (2012). https://doi.org/10.1111/j.1834-7819.2011.01661.x ADEJA2 0045-0421 Google Scholar

190. 

A. J. Huang et al., “Using cone-beam CT as a low-dose 3D imaging technique for the extremities: initial experience in 50 subjects,” Skeletal Radiol., 44 797 –809 (2015). https://doi.org/10.1007/s00256-015-2105-9 Google Scholar

191. 

J. B. Ludlow, B. K. Johnson and M. Ivanovic, “Estimation of effective doses from MDCT and CBCT imaging of extremities,” J. Radiol. Prot., 38 1371 –1383 (2015). https://doi.org/10.1088/1361-6498/aae44b Google Scholar

192. 

E. K. Tuominen et al., “Weight-bearing CT imaging of the lower extremity,” Am. J. Roentgenol., 200 (1), 146 –148 (2013). https://doi.org/10.2214/AJR.12.8481 Google Scholar

193. 

J. A. Carrino et al., “Dedicated cone-beam CT system for extremity imaging,” Radiology, 270 (3), 816 –824 (2014). https://doi.org/10.1148/radiol.13130225 RADLAX 0033-8419 Google Scholar

194. 

A. Hirschmann et al., “Upright Cone CT of the hindfoot: comparison of the non-weight- bearing with the upright weight-bearing position,” Eur. Radiol., 24 (3), 553 –558 (2014). https://doi.org/10.1007/s00330-013-3028-2 Google Scholar

195. 

S. K. Koskinen et al., “CT arthrography of the wrist using a novel, mobile, dedicated extremity cone-beam CT (CBCT),” Skeletal Radiol., 42 (5), 649 –657 (2013). https://doi.org/10.1007/s00256-012-1516-0 Google Scholar

196. 

H. Lang et al., “A retrospective, semi-quantitative image quality analysis of cone beam computed tomography (CBCT) and MSCT in the diagnosis of distal radius fractures,” Eur. Radiol., 26 (12), 4551 –4561 (2016). https://doi.org/10.1007/s00330-016-4321-7 Google Scholar

197. 

P. M. Ricci et al., “Cone-beam computed tomography compared to x-ray in diagnosis of extremities bone fractures: a study of 198 cases,” Eur. J. Radiol. Open, 6 119 –121 (2019). Google Scholar

198. 

P. Kröpil et al., “Cone beam CT in assessment of tibial bone defect healing: an animal study,” Acad. Radiol., 19 (3), 320 –325 (2012). https://doi.org/10.1016/j.acra.2011.10.022 Google Scholar

199. 

G. M. Osgood et al., “Image quality of cone beam computed tomography for evaluation of extremity fractures in the presence of metal hardware: visual grading characteristics analysis,” Br. J. Radiol., 90 (1073), 20160539 (2017). https://doi.org/10.1259/bjr.20160539 BJRAAP 0007-1285 Google Scholar

200. 

Q. Cao et al., “Quantitative cone-beam CT of bone mineral density using model-based reconstruction,” Proc. SPIE, 10948 109480Y (2019). https://doi.org/10.1117/12.2513216 PSISDG 0277-786X Google Scholar

201. 

S. Subramanian et al., “Quantitative evaluation of bone microstructure using high-resolution extremity cone-beam CT with a CMOS detector,” Proc. SPIE, 10953 1095317 (2019). https://doi.org/10.1117/12.2515504 PSISDG 0277-786X Google Scholar

202. 

Y. Aurell, M. L. E. Andersson and K. Forslind, “Cone-beam computed tomography, a new low-dose three-dimensional imaging technique for assessment of bone erosions in rheumatoid arthritis: reliability assessment and comparison with conventional radiography: a BARFOT study,” Scand. J. Rheumatol., 47 (3), 173 –177 (2018). https://doi.org/10.1080/03009742.2017.1381988 Google Scholar

203. 

R. Schouten et al., “Intra-operative cone-beam CT (O-arm) and stereotactic navigation in acute spinal trauma surgery,” J. Clin. Neurosci., 19 (8), 1137 –1143 (2012). https://doi.org/10.1016/j.jocn.2012.01.020 Google Scholar

204. 

A. Sisniega et al., “High-fidelity artifact correction for cone-beam CT imaging of the brain,” Phys. Med. Biol., 60 (41), 415 –439 (2015). https://doi.org/10.1088/0031-9155/60/4/1415 Google Scholar

205. 

P. Wu et al., “Cone-beam CT for imaging of the head/brain: development and assessment of scanner prototype and reconstruction algorithms,” Med. Phys., 47 (6), 2392 –2407 (2020). https://doi.org/10.1002/mp.14124 MPHYA6 0094-2405 Google Scholar

206. 

L. Shi et al., “Characterization and potential applications of a dual-layer flat-panel detector,” Med. Phys., 47 (8), 3332 –3343 (2020). https://doi.org/10.1002/mp.14211 MPHYA6 0094-2405 Google Scholar

207. 

M. Lu et al., “Dual energy imaging with a dual-layer flat panel detector,” Proc. SPIE, 10948 1094815 (2019). https://doi.org/10.1117/12.2513499 PSISDG 0277-786X Google Scholar

208. 

NEXIS, “Next generation x-ray imaging system, 2018-2021,” (2020). Google Scholar

209. 

G. Simone et al., “Organic photodetectors and their application in large area and flexible image sensors: the role of dark current,” Adv. Funct. Mater., 30 (20), 1904205 (2020). https://doi.org/10.1002/adfm.201904205 AFMDC6 1616-301X Google Scholar

210. 

K. Mueller et al., “Towards material decomposition on large field-of-view flat panel photon-counting detectors first in-vivo results,” in Proc. Fourth Int. Conf. Image Formation in X-Ray Computed Tomography, 479 –482 (2016). Google Scholar

211. 

M. Ahmad et al., “Assessment of a photon-counting detector for a dual-energy C-arm angiographic system,” Med. Phys., 44 (11), 5938 –5948 (2017). https://doi.org/10.1002/mp.12517 MPHYA6 0094-2405 Google Scholar

212. 

A. Shankar et al., “Spectroscopy with a CdTe-based photon-counting imaging detector (PCD) having charge sharing correction capability,” Proc. SPIE, 10573 1057352 (2018). https://doi.org/10.1117/12.2294632 PSISDG 0277-786X Google Scholar

213. 

K. Treb et al., “A hybrid photon counting and flat panel detector system for periprocedural hemorrhage monitoring in the angio suite,” Proc. SPIE, 11595 115950U (2021). https://doi.org/10.1117/12.2581242 PSISDG 0277-786X Google Scholar

Biography

Rebecca Fahrig is head of Innovation for Advanced Therapies at Siemens Healthcare GmbH, and a professor in the Department of Computer Science, at the Pattern Recognition Lab at Friedrich-Alexander Universität, Erlangen, Germany. She received her BSc degree in physics and her MSc degree in medical biophysics from the University of Toronto, and her PhD also in medical biophysics from the University of Western Ontario in London, Canada. Her research and predevelopment work focus on multimodality image guidance technologies for use during planning, guidance, and assessment of minimally invasive therapies. She is a fellow of the AAPM and AIMBE.

David A. Jaffray is a professor in the Departments of Radiation Physics and Imaging Physics at the University of Texas MD Anderson Cancer Center and their chief technology and digital officer. He served UHN/PMH/University of Toronto as head of medical physics, director of Techna, and professor in the Departments of Radiation Oncology, Medical Biophysics, and IBBME, as well as EVP, Technology, and Innovation. He is world-renowned in the field of medical physics and technology development in cancer, including his pioneering development of cone-beam CT for image-guided radiotherapy. He holds 47 patents and has >300 peer-reviewed publications in cancer science and technology.

Ioannis Sechopoulos leads the Advanced X-ray Tomographic Imaging Lab of the Department of Medical Imaging at Radboud University Medical Center. He has a secondary appointment at the Multi-Modality Medical Imaging Group at the University of Twente and is a scientific advisor of the Dutch Expert Center for Screening (LRCB). He received his PhD from Georgia Institute of Technology performing research in x-ray breast imaging at the Department of Radiology and Imaging Sciences of Emory University. Since then, his main area of research is the development of advanced x-ray-based imaging techniques, including for breast imaging and several body CT clinical applications.

J. Webster Stayman is an associate professor in the Biomedical Engineering Department at Johns Hopkins University. He has led efforts on the development of hardware and software for various imaging systems, including dedicated and interventional cone-beam-computed tomography devices. His research interests include system analysis and optimization, task-driven performance assessments, advanced algorithm development and machine learning, device design and prototyping, and clinical translation.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Rebecca Fahrig, David A. Jaffray, Ioannis Sechopoulos, and Joseph W. Stayman "Flat-panel conebeam CT in the clinic: history and current state," Journal of Medical Imaging 8(5), 052115 (28 October 2021). https://doi.org/10.1117/1.JMI.8.5.052115