5 February 2024 Sounding out the dynamics: a concise review of high-speed photoacoustic microscopy
Author Affiliations +
Abstract

Significance

Photoacoustic microscopy (PAM) offers advantages in high-resolution and high-contrast imaging of biomedical chromophores. The speed of imaging is critical for leveraging these benefits in both preclinical and clinical settings. Ongoing technological innovations have substantially boosted PAM’s imaging speed, enabling real-time monitoring of dynamic biological processes.

Aim

This concise review synthesizes historical context and current advancements in high-speed PAM, with an emphasis on developments enabled by ultrafast lasers, scanning mechanisms, and advanced imaging processing methods.

Approach

We examine cutting-edge innovations across multiple facets of PAM, including light sources, scanning and detection systems, and computational techniques and explore their representative applications in biomedical research.

Results:

This work delineates the challenges that persist in achieving optimal high-speed PAM performance and forecasts its prospective impact on biomedical imaging.

Conclusions

Recognizing the current limitations, breaking through the drawbacks, and adopting the optimal combination of each technology will lead to the realization of ultimate high-speed PAM for both fundamental research and clinical translation.

1.

Introduction

Photoacoustic (PA) imaging (PAI) has emerged as a promising biomedical technique based on the PA effect, wherein ultrasonic emission occurs upon the absorption of a short-pulsed laser by irradiating objects, such as biological tissues.13 The resultant ultrasonic waves, termed PA waves, are captured by a transducer to form PA images for diverse applications [Fig. 1(a)]. PA microscopy (PAM) is a prevalent form of PA imaging that achieves micron-level spatial resolution by focusing either the optical or acoustic beam or both. According to the focusing type, the PAM system is categorized into optical-resolution PAM (OR-PAM) and acoustic-resolution PAM (AR-PAM).5,6 Using wide-field optical excitation, AR-PAM can achieve a 3 to 5 times larger penetration depth than OR-PAM [Fig. 1(b)]. By contrast, because an optical focus is approximately 10 times tighter than an acoustic focus, OR-PAM excels in delivering higher resolution, which has the capability to image physiological and pathological changes in single capillaries.7 Furthermore, the remarkable optical absorption contrast makes PAM a promising imaging tool across various research domains including biology, dermatology, neurology, oncology, ophthalmology, and pathology.8

Fig. 1

Principle of PA imaging and penetration limits of PAM. (a) Illustration of PA imaging principle. (b) Penetration depth of OR-PAM and AR-PAM with different optical and acoustic focusing configurations. The images are reprinted with permission from Ref. 4.

JBO_29_S1_S11521_f001.png

Traditional PAM systems employ a confocal and coaxial configuration for the excitation light and acoustic emission beams, maximizing the signal-to-noise ratio (SNR) and spatial resolution. Because the acquisition of three-dimensional (3D) images is conducted by raster scanning of the PA probe over the sample, with the typical lasers and scanning systems, the combination of a low pulse repetition rate (PRR) and a fine scanning step size has led to a prolonged imaging time. Consequently, typical OR-PAM faces limitations in capturing dynamic tissue information, such as transient drug responses in brain microvasculature.9 Thus, many efforts have concentrated on improving the imaging speed of PAM to monitor the rapid dynamic changes in biomedical research.

Fundamentally, the imaging speed of PAM is limited by the PRR of the light source because a single laser pulse can generate a time-resolved PA signal. Assuming that there is no need for signal averaging for improving the SNR, the laser PRR generally corresponds to the A-line rate of PAM. Various types of pulsed lasers can be classified according to the wavelength, pulse duration, pulse energy, and PRR. Among them, Q-switched lasers, which usually have a nanosecond pulse duration, have been mainly used in PAI studies.10 Because the first-generation PAI needed the millijoule (mJ) level of pulse energy for the wide-field illumination, the initial research of PAM was conducted with slow and bulky lasers such as an optical parametric oscillator (OPO),11 a Ti:Sapphire laser,12 or a dye laser13 pumped by a flash-pumped solid-state laser (FPSS). However, as the need for speed improvement increases in the high-resolution scheme, greater attention was paid to the high-PRR lasers, which can replace conventional lasers.14 Therefore, in recent years, fiber lasers have been increasingly used to achieve ultra high-speed PAM, with a short pulse duration (<10  ns), a large pulse energy (>hundreds nJ), and an ultra-high PRR (hundreds kHzMHz).1524

From the perspective of a scanning system, a wide field-of-view (FOV) with a short imaging time is another need for high-speed PAM. Because diverse temporal and spatial scales exist in biological functions, a strategy to match the FOVs and scanning speed is necessary. Because typical PAM is sensitive to significant motion artifacts, such as those induced by respiratory movements, high-speed scanning methods with large FOVs have been reported to maximize the functional imaging capabilities of PAM. Initially, researchers have explored the use of faster scanning stages, such as voice coil scanners25 and slider-crank scanners,26,27 while maintaining the same setup as conventional PAM systems. These mechanical scanning approaches allowed the systems to achieve a high SNR; however, they consistently faced speed limitations due to the bulky scanners and associated vibrations. Moreover, scanning speeds were still insufficient to meet the desired requirements for dynamic imaging. On the other hand, the optical scanning approach employs stationary PA signal detection without any mechanical scanning of the sample/transducer; thus, the scanning speed is only limited by the laser’s pulse rate or the speed of PA signal generation.28 The use of a water-immersible mirror driven by a microelectromechanical systems (MEMS) scanner2931 introduced an innovative approach that led to a significant scanning speed improvement in PAM. Following this breakthrough, other optical scanning solutions such as galvanometer scanners32 and polygon scanners22 have emerged; each offers a different configuration to optimize the scanning speed. These advancements have pushed the boundaries of PAM speed, approaching the physical limit dictated by the speed of sound.

To further improve the imaging speed of PAM, in addition to the scanning mechanism, it is necessary to explore alternative approaches that can expedite the detection of PA signals. Optical sensors have emerged as a promising avenue, with a focus on achieving faster signal detection speeds. These optical sensors offer several advantages, including their compact size, wide reception angle, broad detection bandwidth, strong responsiveness to low-frequency signals, and seamless integration with the PA light path. Notably, the sensitivity of optical sensors tends to exhibit less dependence on sensor size, resulting in superior sensitivity compared with piezoelectric transducers of similar dimensions, particularly at higher frequencies (>2.5  MHz).33

Computational techniques to improve the imaging speed of PAM are inextricably linked to hardware limitations, such as the PRR of laser sources and scanning mechanisms. These software techniques serve to either circumvent or correct some of the hardware limitations of high-speed PAM systems. When scanning speed limitations are reached from the hardware perspective, due to the limited PRR of laser sources or scanning mechanisms, undersampling is often performed by covering the same FOV with fewer sampling events.34 When using traditional pulsed laser sources, this undersampling typically results in a lower sampling density and violates the Nyquist sampling theorem, causing overall poorer image quality and introducing aliasing (spatial undersampling).34,35 When performing localization microscopy, undersampling typically takes the form of using fewer frames, thus resulting in images with a poorer SNR or incomplete structures due to insufficient sampling (temporal framewise undersampling).36 Therefore, many computational techniques for high-speed PAM aim to overcome the tradeoff between imaging speed and its subsequent degradation of image quality through a process of image restoration. Other computational techniques speed up PAM image formation by attempting to correct imperfections in the scanning mechanism, such as by quickly restoring misaligned/distorted aspects of the scanning path or denoising low-energy images acquired within certain laser dosage limits.22,3739

In this concise review, we cover a brief overview of (i) light sources, (ii) scanning and detection, and (iii) computational techniques to improve the PAM imaging speed. Then we introduce the recent technical advances compared with conventional PAM system in the last 5 years or so. Finally, we conclude by briefly discussing the remaining challenges and future potential for high-speed PAM.

2.

Light Sources

The pulsed light source is one the most critical components of PAM, which often determines its maximal A-line rate, SNR, and lateral spatial sampling density. Accelerating the speed of PAM requires the engineering of novel light sources that can provide a high PRR, high pulse energy, and ideally, multi-wavelengths for functional studies, which are usually competing parameters for any given laser type.

2.1.

Conventional Excitation Lasers for PAM

In traditional PA imaging, FPSS lasers were used due to their extremely high energy (∼ several J), which enables obtaining the various scales of PA image and pumping the dye laser or OPO.1113 However, the FPSS lasers has a low PRR (10100  Hz) and a limited lamp lifetime of a few million pulses.40,41 Thus, diode-pumped solid-state (DPSS) lasers, utilizing a high-power semiconductor laser diode as a pumping source for Q-switching, have been widely used with a higher speed (kHz range).42 Laser diodes can optimize crystal efficiency for the lowest possible pump energy by tuning the wavelength controlled by temperature and driving current, which allows DPSS lasers to have high energy efficiency, good beam quality, high compactness, and long lifetime.43 PAM can also be implemented using Q-switched microchip lasers, which are alignment-free monolithic solid-state lasers (10  kHz100  kHz). The microchip cavity consists of a thick disk sandwiched between mirrors with extremely short cavity lengths, which allows the laser to oscillate on a single longitudinal mode generating a nanosecond Q-switched pulse.44 Microchip lasers can reach up to above 100 kHz with passive Q-switching, but in practice, PAM typically operates at tens of kHz PRRs with a few microjoules due to trade-offs between the PRR and pulse energy.45 When a gain crystal (e.g., Nd:YAG or Nd:YVO4) is overloaded by high pump power for a high PRR (hundreds of kHz), the thermal lensing effect and thermal damage occur, leading to serious performance degradation.46 Thus, solid-state lasers with crystals are not capable of providing high-energy nanosecond pulses at a high PRR.

2.2.

High-Speed Excitation Lasers

In terms of thermal resistance, fiber lasers can be a solution for high-speed PAM. The optical fiber has the merit of a large surface-to-volume aspect ratio, which prevents the thermal lensing effect by facilitating heat dissipation. This advantage enables the fiber lasers to withstand the frequent on/off intensity modulation and high peak power at a high PRR. Thus, current high-speed PAM systems have mostly adopted fiber lasers to provide a stable output with microjoule energy at the MHz-level PRR.1618,2022

2.2.1.

Single wavelength excitation

Shi et al. first discussed the potential of Q-switched fiber lasers as a high-speed OR-PAM light source in 2010.14 Because the optical absorption of hemoglobin at 1064 nm is low, a custom-built passively Q-switched ytterbium doped (Yb-doped) fiber laser at 1075 nm with a PRR of 100 kHz and a pulse energy of a 13  μJ was converted to 537 nm with a pulse energy of 0.06  μJ and a pulsewidth of 250 ns pulse, by coupling the 1075 nm light into a potassium titanium oxide phosphate (KTP) frequency-doubling crystal. The capability of fiber lasers with a high PRR (hundreds of kHz) was demonstrated for real-time OR-PAM, though the long pulse duration resulted in a poor SNR. The same group later demonstrated in vivo OR-PAM using a 532-nm commercial Yb-doped fiber laser with a PRR up to 600 kHz.24 The short pulse duration (1  ns) met the thermal and stress-confinement conditions, and the relatively high pulse energy (150  nJ) enabled high-contrast in vivo images of microvasculature in a mouse ear. In 2015, Yao et al. adopted a picosecond fiber laser with a PRR of 500 kHz for fast functional PAM of a mouse brain based on the single-wavelength pulse-width method.15 Despite using only single-wavelength excitation, the oxygen saturation of hemoglobin (sO2) in the mouse brain was mapped by acquiring the relative concentration of oxyhemoglobin and deoxyhemoglobin under nanosecond and picosecond excitations, respectively. This fiber laser-based OR-PAM with a submersible MEMS scanner achieved a C-scan rate of 1 Hz over an FOV of 3×2  mm2. In 2018, Allen et al. demonstrated high-speed laser-scanning OR-PAM using a custom-built fiber laser operating at a PRR of 2 MHz.16 The maximum PRR, fundamentally limited by the depth-range ambiguity condition, was set at 2 MHz assuming the absorbers are located at a depth range of 750  μm. In combination with an MHz fiber laser and a fiber-optic ultrasound detector scanned by a galvanometer, an in vivo OR-PAM image of the mouse ear with an FOV of 10×10  mm2 was obtained within 8 s.

2.2.2.

Multi-wavelength excitation

The fast advancement of commercial nanosecond fiber lasers has in turn promoted the development of multi-wavelength light sources for high-speed PAM capable of functional imaging. Supercontinuum (SC) sources have been increasingly applied for multi-wavelength excitation in PAM.4752 The use of highly nonlinear fibers, such as photonic crystal fibers, broadens the output light spectrum to hundreds of nanometers, which leads to a low energy density within each specific band. Recently, Chang et al. reported the use of a commercial SC source with a PRR of up to 1 MHz for multispectral PAM and OCT.17 Two wavelength bands around 528 nm and 558 nm light with a 10 nm bandwidth were used to image sO2 of the mouse ear. However, the two PA signals, generated at a PRR of 500 kHz, were respectively averaged 250 times to improve the SNR.

The stimulated Raman scattering (SRS) source is another technique for multi-wavelength PA excitation; it generates cascaded Stokes outputs based on the nonlinear SRS effect in optical fibers. Because SRS sources have highly condensed energy in a narrow band converted from the pump light [Fig. 2(a)], they have been adopted for implementing high-speed multi-wavelength PAM.1823 Hajireza et al. first reported an SRS source pumped by a Yb-doped fiber laser for multi-wavelength OR-PAM and conducted in vivo functional imaging of the mouse ear.23 Although the pump laser output had a PRR of up to 600 kHz, the carbon fibers were imaged at a PRR of 160 kHz, and the in vivo experiment was demonstrated at a PRR of 40 kHz, where the desired wavelengths at 545 and 558 nm had the maximum energy through a polarization-maintaining single-mode fiber (PM-SMF). Cho et al. reported the optimal generation of SRS sources with a PRR of 300 kHz for high-speed molecular PAM.19 To generate the various wavelengths with sufficient energy, 10 Stokes SRS wavelengths ranging from green to red were optimized with the pulse energy of 116756  nJ through a PM-SMF. By employing the optimized wavelengths, in vivo molecular PAM was used to distinguish the gold nanoparticles from the blood vessels at a PRR of 300 kHz. Wang’s group employed a 1-MHz dual-wavelength pulse laser system based on the SRS effect for high-speed PAM.18,20 The pump laser was a Q-switched fiber laser with a PRR of 1 MHz. Using a dual-path configuration with optical delay [Fig. 2(b)], a pump wavelength (532 nm) and a Raman Stokes wavelength (558 nm) were generated with a 150 ns interval. This study demonstrated functional OR-PAM at a 1 MHz A-line rate, which allowed for the observation of the rapid sO2 change in the mouse ear after intravenous epinephrine injection. He et al. proposed another 1-MHz dual-wavelength PAM using a crystal-based SRS source21 and combined a picosecond fiber laser and a potassium gadolinium tungstate [KGd(WO4)2] crystal with high Raman gain coefficient to enhance the SRS effect. As shown in Fig. 2(c), a dual-wavelength PAM was implemented using two 532-nm picosecond-pulsed lasers, one for the 532-nm path and the other for the first-Stoke 558-nm path. At a PRR of 1-MHz, PAM was used to observe the single-impulse-stimulated initial dip from cerebral micro-vessels without repeated stimulation. In a recent work, Zhu et al. developed high-speed PAM based on an SRS source pumped by a fiber laser with a PRR of up to 2 MHz.22 Dual-wavelength excitation at 532 nm and 558 nm with a PRR of 800 kHz enabled functional imaging of rapid whole-brain hemodynamics, with a volumetric frame rate of 2 Hz over an FOV of 11×7.5×1.5  mm3.

Fig. 2

High-speed PAM using an SRS source. (a) Schematic illustration of the SRS effect in an optical fiber and the discrete Raman spectrum.10 (b) Schematic of the high-speed PAM system using an SRS source with an optical fiber pumped by a 1-MHz nanosecond pulsed laser.20 (c) Schematic of the high-speed PAM using an SRS source with a Raman crystal pumped by a picosecond pulsed laser.21 Images (a), (b), and (c) are reprinted with permission from Refs. 10, 20, and 21, respectively.

JBO_29_S1_S11521_f002.png

3.

Scanning Systems

High-speed PAM systems have been focused on expanding the FOV while preserving high-sensitivity detection. Although the scanning time is technically determined by the PRR of the laser and scanning mechanism, it is fundamentally constrained by the speed of sound in biological tissues.16 Various high-speed scanning methods have been demonstrated for PAM, including mechanical scanning methods, such as voice-coil scanners and slider-crank scanners, and optical scanning methods, such as galvanometer scanners, MEMS scanners, and polygon-mirror scanners. Different scanning methods have their own merits and drawbacks.

3.1.

Mechanical Scanning

Mechanical scanning enables high-speed imaging with confocal and coaxial alignment of the light and sound over a large flat focal plane, and it consistently offers the best SNR. For example, the voice-coil scanner is an effective mechanical solution for high scanning speeds of up to 40 Hz B-scan rate.25 Nevertheless, as the scanning speed increases, the bulky voice-coil scanner head induces vibrations due to high driving forces, degrading its performance. Recently, a slider-crank scanner with a compact design and reduced vibration was reported.27 Using two sets of PA scanning probes, the slider-cranks scanner enabled a large scanning range of 24 mm26 [Figs. 3(a) and 3(b)]. However, it is still limited by the physical properties of the motor (e.g., speed and torque) and can provide a scanning speed of up to 32 Hz B-scan rate, which is much lower than most optical scanning systems.

Fig. 3

High-speed PAM system with mechanical scanning and optical scanning approaches. (a) Schematic of slider-crank scanner using two-channel PA probes. (b) High-resolution nude mouse ear PA image acquired by the slider-crank-scanner PAM system.26 (c) The illustration of high-speed PAM using a water-immersible polygon scanner with 12 facets. (d) PA images of the mouse brain vasculature and oxygenation over the entire cortex obtained from the polygon-scanner PAM system.22 Images (b) and (d) are reprinted from Refs. 26 and 22, respectively.

JBO_29_S1_S11521_f003.png

3.2.

Optical Scanning

Galvanometer scanners are known for fast and stable scanning in conventional optical microscopy. Galvanometer scanners in PAM were previously employed for steering the light beam in air, leading to the non-confocal alignment of the light and sound.16,53,54 Consequently, the SNR was relatively low and the optical scanning range was limited within the transducer’s detection zone. To address this issue, water-immersible optical scanners were used in PAM to concurrently and confocally steer both the optical beams and emitted acoustic waves. Although this approach increased the B-scan rate to several hundred Hz with a high SNR, the scanning range remains limited. For example, a semi-water-immersible galvanometer scanner achieved a B-scan rate of 500 Hz within a small FOV of 2.4  mm.32 On the other hand, employing cylindrically-focused or unfocused acoustic detection can expand the FOV significantly to 40  mm.55 However, this method has the cost of a reduced SNR.

Water-immersible MEMS scanners facilitated high-speed B-scan scanning at 400 Hz29,30 with a scanning range of 3 mm. However, these MEMS scanners had a highly reduced FOV unless operated at their resonant frequency. Furthermore, MEMS scanners were prone to thermal damage during prolonged use, resulting in distorted scanning patterns and warped PA images. Recently, Chen et al.31 introduced a torsion-bending-based MEMS scanner, achieving a B-scan rate of 400 Hz and a volumetric imaging speed of 1 Hz, covering a FOV measuring 1.5×2.5  mm2. This system capitalized on the benefits of independent scanning axes, resulting in enhanced stability, particularly in aqueous environments. However, it was still limited by the relatively small FOV, compared with other mechanical or galvanometer scanners.

To achieve a large FOV, Zhu et al. used a 12-facet polygon-mirror scanner, which provided a B-scan rate of 2 kHz across an 11-mm scanning range22 [Figs. 3(c) and 3(d)]. Nonetheless, uneven step sizes across the scanning range degraded the PA image quality, necessitating data interpolation for final image reconstruction. Furthermore, steering the laser and ultrasound beams after the objective lens induced a curved scanning plane, further limiting the image quality.56 Moreover, the long travel distance for the ultrasound waves from the sample to the transducer increased PA signal attenuation.

4.

Detecting systems

Piezoelectric ultrasound transducers continue to dominate the PAM technologies due to their broad availability, high detection sensitivity, low cost, and ease of use. Nonetheless, piezoelectric transducers are often not transparent to light and thus complicate the PAM system design for light and sound alignment. Recently, optical ultrasound sensors have presented distinct advantages for PAM and gained increasing interest. Notably, the Fabry–Perot (FP) interferometer, functioning as an optical sensor, offers nearly isotropic spatial resolutions that are primarily determined by the size of the optical probing beam.57 The high-resolution imaging can be consistently maintained through 2D dense spatial sampling across the entire FOV. Traditionally, the imaging speed of the FP interferometer-based PAM system has been limited by the slow point-by-point raster scanning of the probing beam and the relatively low PRR of the PA excitation laser, typically operating at 50 Hz. However, a recent work conducted by the UCL group showcased a remarkable 32-fold increase in imaging speed.58 This improvement was achieved by simultaneously employing a total of eight parallel probing beams, all scanning over the sensor concurrently. For high-speed PA applications, a planoconcave fiber-optic sensor, which is a deposited polymer FP interferometer at the tip of an optical fiber, offers high detection sensitivity and an extensive bandwidth. Its wide acceptance angle (±90  deg) allowed for the fast detection of PA signals generated across a substantial FOV without requiring any sensor translation. In a recent work, Allen et al. adopted a custom-built fiber laser operating at a PRR of 2 MHz and demonstrated, for the first time, high-speed PAM using a planoconcave fiber-optic sensor [Figs. 4(a) and 4(b)].16 This high-speed PAM system employed a rapid galvanometer scanner to perform a raster scan of the laser beam across the sample surface while employing a stationary fiber-optic ultrasound detector for PA signal recording. Nonetheless, one limitation of the planoconcave fiber-optic sensor is its uneven sensitivity across a wide FOV, resulting in non-uniform PA image quality. Another approach used an unfocused side-looking fiber optic-sensor that was based on a dual-polarized fiber laser [Figs. 4(c) and 4(d)].59 This method combined stationary acoustic detection using the side-looking optical sensor with the fast laser beam scanning by a 2D galvanometer mirror. As a result, hemodynamic imaging was successfully demonstrated within a 2×2  mm2 region at a volumetric frame rate of 2 Hz on the mouse ear. With a B-scan rate of 400 Hz, this technique allowed for imaging physiological dynamics in both trunk vessels and capillaries. Other optical ultrasound sensors, such as polymer micro-rings61 and Bragg-grating fibers,62 have recently been demonstrated as point-like detectors in PA imaging systems. These optical sensors often have sizes that are orders of magnitude smaller than their piezoelectric counterparts. However, a significant challenge faced by these optical sensors is the complexity of scaling up their fabrication while maintaining consistent optical properties such as the optical resonant wavelength, Q factor, and transmission efficiency. Therefore, they have not yet been applied on high-speed PAM systems. A comprehensive comparison of optical sensors and piezoelectric transducers can be found in the review article authored by Wissmeyer et al.63

Fig. 4

Various PA detection methods and their representative results by applying the optical scanning scheme. (a) Schematic of the planoconcave fiber-optic sensor.16 (b) in vivo PAM image of mouse ear obtained at 2 million A-lines per second. The scan area was 10  mm×10  mm, and the total acquisition time was 8 seconds. (c) Schematic of the fiber-laser based optic sensor method.59 (d) In vivo PAM image of the mouse ear at a C-scan frame rate of 0.2 Hz. The scanning FOV was 2.2×2.2  mm2. (e) Schematic of the transparent transducer method.60 (f) In vivo PAM image of whole cortical vasculature of the mouse brain. The total acquisition time was 5 s to scan an FOV of 6×6  mm2. Images (a), (b), and (c) are reproduced with permission from Refs. 16, 59, and 60, respectively.

JBO_29_S1_S11521_f004.png

Different from the conventional point-by-point scanning methods, parallel excitation and/or detection over a wide FOV allows for high-speed volumetric PA imaging. A recent study by Li et al. introduced a high-speed PAM implementation known as PA Topography through an Ergodic Relay (PATER).64 This method was based on the simultaneous encoding of PA signals from a wide FOV based on their distinct time-delay characteristics. In PATER, each individual excitation laser pulse enabled the parallel detection of encoded PA signals using a single-element ultrasound transducer. These signals were subsequently mathematically decoded to reconstruct a 2D projection image. By incorporating a point-by-point scanning calibration process, PATER showcased a topographic frame rate of 2 kHz over a FOV of 8×6  mm2. This approach was applied to visualize the blood pulse wave velocity and monitor the circulation of melanoma cells within the mouse brain. PATER’s imaging speed was primarily limited by the acoustic transit time within the ergodic relay as it eliminated the need for optical or acoustic beam scanning. However, the existing calibration method lacks depth information; thus, it can only provide topographic images. Xia et al. introduced multifocal PA-computed microscopy, utilizing a 2D microlens array and a 512-element full-ring ultrasonic transducer array.65 They transmitted 1800 optical foci within the transducer’s focal plane and performed raster scanning with a 25  μm step size. As a result, they obtained a cross-sectional PAM image across a FOV of 10×10  mm2 in just 36 s. However, the resolution (36  μm) was quite poor in comparison with other PAM systems. This approach was also applied in ultraviolet PAM by Imai et al., who developed a high-throughput multifocal Ultraviolet PAM (MF-UV-PAM) that employed a linear microlens array, which requires only 1D scanning for volumetric imaging by utilizing a 256-channel data acquisition system.66 They accomplished a lateral resolution of 1.6  μm and achieved a remarkable 40-fold increase in imaging speed compared with conventional raster scanning. The transparent ultrasound transducer is another novel approach aimed at simplifying and accelerating PAM systems.67,68 As shown in Figs. 4(e) and 4(f), with this design, the laser beam can pass through the optical transparent transducer, without the need for an optical acoustic combiner or/and steering mirror. In the context of high-speed PAM, a cylindrically-focused transparent high-frequency ultrasound transducer (CFT-UT) combined with a galvo scanner was used.60 This setup managed to achieve a B-scan frame rate of 500 Hz over a scanning range of 9 mm. However, it is important to note that the current prototype of the CFT-UT exhibited non-uniform optical transparency and detection sensitivity, resulting in non-uniform PA images (Table 1).

Table 1

Specifications of PAM with high imaging speed. PRR, pulse repetition rate; fc, center frequency of ultrasound sensor; Δf, bandwidth of ultrasound sensor; FOV, field-of-view; δx, lateral resolution; δz, axial resolution; PT, piezoelectric transducer; and FP, Fabry–Pérot.

JBO_29_S1_S11521_f007.png

Often in PAM, a volumetric image can be generated by 2D raster scanning with depth-resolved PA signals, and thus depth scanning is not necessary. With that said, due to the limited depth of focus of the light beam in OR-PAM, depth scanning is sometimes needed to extend the imaging depth on thick samples. Several recent works focus on extending the depth of focus of the light beam and thus increasing the volumetric imaging speed. Techniques such as using dual non-diffracting Bessel beams,69 synthetic Bessel light needles,70 and deep learning methods71 have been employed to extend the depth of field in OR-PAM. These advancements enable an extended focus and faster volume scans. Bessel beams, in particular, have shown promise in maintaining a narrow focal region over a longer range,70 which is beneficial for high-resolution imaging over extended depths.

5.

Computational Techniques

As the imaging speed of PAM continuously increases, the inherent drawbacks with a high imaging speed may not be addressed by improving the system configurations alone. Such drawbacks include the spatial undersampling, scanning-line misalignment, reduced SNR, and deteriorated spatial resolution. Advanced computational techniques can be used to enhance the high-speed PAM image quality, mostly by focusing on the post-imaging processing.

5.1.

Upsampling Methods

5.1.1.

Spatial undersampling

One of the primary ways computational techniques have been used to speed up PAM imaging is to reduce the number of spatial sampling events required for a complete PAM scan, particularly when the laser’s PRR is limited. Although often less effective than modern deep learning techniques, non-deep learning approaches can be more interpretable and, without the need for extensive training datasets or complex data augmentation regimes, can still be more generally robust. Similar to compressed sensing, these non-deep learning approaches can recover sparsely sampled data by exploiting known information about the data’s underlying sparsity to overcome typical Nyquist–Shannon limitations. One method utilized to transform sparsely sampled PAM data into an approximation of the fully sampled images frames the recovery task as a matrix completion problem. In Lui et al.’s work,72 the alternating direction method of multipliers (ADMM) based low-rank and sparse matrix recovery method is used on sparse PAM data to enable fast vascular imaging. By enforcing a sparsity constraint (total variation norm) and a low-rank constraint (nuclear norm), this optimization problem can be solved with ADMM, resulting in superior signal recovery compared with other contemporary algorithms such as GoDec.73 Dictionary learning is another popular non-deep learning approach for recovering quickly acquired, sparsely sampled PAM images. In dictionary learning, a sparsely coded dictionary is used to represent the data, whereby the dictionary is iteratively updated to minimize the representation error while maintaining a certain sparsity constraint. The desired result is a dictionary representation that best represents the acquired data given the specified level of sparsity. In Sathyanarayana et al.’s work,74 a dictionary learning technique known as K-SVD is implemented with a dictionary size of 1024, sparsity of 20, 5 inner loop iterations of weighted rank 1 SVD approximation, and 5 outer loop iterations of K-SVD alternating minimization. This method was able to recover randomly downsampled PAM images of 50% and 75% sparsity with a minimal impact on the peak-signal-to-noise ratio (PSNR) compared with the fully sampled images. This work was later expanded upon in another work by Sathyanarayana et al.,75 which could recover hemodynamic parameters from undersampled multi-parametric PAM data with up to eight times undersampling, with less error than bicubic interpolation. More recently in Sulistyawan et al.’s work,76 a curvelet transform was used to reconstruct the boundary and separability of cells in images randomly undersampled up to eight-fold with better noise rejection and edge recovery compared with nearest neighbor interpolation with a smoothing filter.

DiSpirito et al. published the first work to use deep learning to accelerate PAM through reconstructing undersampled images.34 In this work, several deep learning model architectures were compared, and a modified fully dense U-Net (FD U-Net) was found to perform the best at the various downsampling ratios. At each of these downsampling ratios, FD U-Net was able to reconstruct vessel structures with smoother edges and less jagged aliasing artifacts than bicubic interpolation as the model learned vasculature features during its supervised training [Fig. 5(a)]. This sense of vesselness was reinforced using a multifactorial loss function that included pixelwise loss in the spatial and Fourier domains, as well as perceptual loss in the form of structural similarity index measure and PSNR.

Fig. 5

Deep learning methods for upsampling PAM data. (a) PAM MAPs upsampled using FD U-Net with supervised training at various downsampling ratios (i.e., ratios [Dx, Dy]) compared with interpolation.34 (b) PAM MAPs upsampled using the unsupervised DIP iterative method compared with interpolation methods and FD U-Net.35 (c) PAM data upsampled in 3D using a combination of axial z-axis upsampling (SA-SRResNet) and lateral x-y-axis upsampling (DA-SRResNet).77 Images (a), (b), and (c) are reprinted with permission from Refs. 34, 35, and 77, respectively.

JBO_29_S1_S11521_f005.png

Soon after, Zhou et al.78 reported that a convolutional neural network architecture with 16 residual blocks and 8 squeeze-and-excitation (SE) blocks was able to reconstruct undersampled PAM images. Unlike DiSpirito et al.,34 the proposed model was trained on experimentally downsampled PAM images of 25% and 6.25% effective pixels respectively, rather than synthetically downsampled images. The trained model was shown to outperform bicubic interpolation and several other popular single image super-resolution deep learning architectures. The authors emphasized that the high quality of the images reconstructed using their technique was in part due to training on experimentally downsampled data with a VGG19 perceptual loss function.

Rather than using a traditional supervised deep learning approach, whereby one needs a large dataset of ground truth high-resolution PAM images, Vu et al.35 used an unsupervised method to upsample PAM images [Fig. 5(b)]. In this work, the deep image prior (DIP) technique utilized a modified FD U-Net to iteratively convert Gaussian noise into a given undersampled PAM image using the known downsampling binary mask. Though this unsupervised method had a much longer inference time due to the need to iteratively retrain for every new image, it also benefited from being able to create more natural vessel structures, without the need for ground truth or the presence of transposed convolution checkboard artifacts.

Most spatial undersampling works in PAM have been restricted to working with 2D maximum amplitude projection (MAP) images and only with lateral (or x-y) direction undersampling. However, PAM images are originally acquired as 3D volumes, with a depth direction z in addition to the usual x-y directions shown in MAP images. To utilize a lower sampling frequency and reduce PAM’s volumetric data size, Seong et al.77 used a modified super-resolution ResNet (SRResNet) to improve PAM data undersampled in 3D volume (along the z and x-y directions) [Fig. 5(c)]. This work used two deep learning models, a modified single axis SRResNet and a dual axis (DA) SRResnet, to perform upsampling along the z-direction and the x-y directions respectively. Similar to DiSpirito et al.’s work, Seong et al. used fully sampled data during model training as ground truth and then simulated undersampled data by downsampling along the x-, y-, and z-directions respectively, with the ability to vary the downsampling ratio used for each spatial dimension. The model also performed well on experimentally downsampled test data.

5.1.2.

Temporal undersampling

In addition to reducing the required spatial sampling density of PAM, deep learning has also been used to reduce the number of temporal frames required to form PA localization microscopy images. In PA localization microscopy, N number of frames capturing localized targets are combined to attain a high-quality dense localization-based image. This image formation technique can acquire high resolution images of vasculature at the expense of poor temporal resolution as a sufficient number of frames must be acquired and combined to form a high-quality image. To circumvent the speed limitation, Kim et al.36 used a deep learning model based on the Pix2Pix generative adversarial network (GAN) framework to convert low quality (low N number) images into their higher quality (high N number) equivalents. A 3D U-Net was used as the generator, with seven 3D convolutional layers and roughly 43 million trainable parameters, and the discriminator consisted of five convolutional layers connected in series and contained 5 million trainable parameters. Model performance was exhibited after being trained on various input N frame total values, with potential imaging speed enhancements of 12-fold.

5.2.

Computational Correction Methods

5.2.1.

Alignment of the scanning system

Computational techniques have also been used alongside fast hardware to help speed up PAM through performing imaging system corrections. For example, the manufacturing of polygon scanning mirrors can introduce minor imperfections that present as distortions and misalignments among the various polygon facet faces. In Zhu et al.’s work,22 a combination of affine transformations between facet images and deep learning upsampling was used to correct for this facet misalignment [Fig. 6(a)]. The calibrated affine transformations removed most of the global distortions that were caused by polygon facet misalignment, and the remaining minor local distortions were mitigated further by the deep learning upsampling procedure.

Fig. 6

Improving PAM images by (a) aligning polygon facets in a fast polygon scanning system combined with deep learning FD U-Net upsampling,22 (b) recovering misaligned signal from bidirectional scan using deep learning,37 (c) correcting MEMS scanning distortion using a deep learning spatial weight matrix (SWM) with a dimensionality reduction,38 and (d) improving the image quality of low energy PAM using MT-RDN.39 Images (a) and (b)–(d) are reprinted with permission from Refs. 22 and 3738.39, respectively.

JBO_29_S1_S11521_f006.png

When using high-speed water-immersible MEMS scanners, the consecutively acquired bidirectional PAM images are typically misaligned because of unstable mirror performance. This usually results in one of the bidirectional scans being discarded, thereby doubling the scanning time necessary to meet Nyquist requirements. In Kim et al.’s work,37 the misalignment in these bidirectional scans was corrected using a deep learning network based on a FD U-Net, known as multiscale (MS) FD U-Net, in a GAN configuration [Fig. 6(b)]. When compared with traditional enhancement filtering (such as bicubic, bilateral, and median filters) and previously reported deep learning networks (Dense GAN and FD U-Net), the proposed MS FD-U-Net was shown to be superior in terms of perceptual similarity to the ground truth.

MEMS scanning mirrors are often used to accelerate OR-PAM. However, the nonlinear tilt angular-voltage characteristic of a MEMS mirror introduces distortions into the MAP image. In addition, other factors such as the size of the airy disk, ultrasonic sensor properties, and thermal effects can decrease the resolution. Ma et al.38 proposed a spatial weight matrix (SWM) with a dimensionality reduction for image reconstruction that aimed to improve these downsides of traditional MEMS scanners [Fig. 6(c)]. The three-layer SWM was calibrated to the constant distortion factors of the system, allowing for both spatial dependent distortion correction and 3D deconvolution. An ordinal-valued Markov random field and Harris Stephen algorithm were used in tandem with a modified delay-and-sum method during reconstruction. Using this multistep adaptive calibration approach, Ma et al. showcased that many of the distortions caused by MEMS system characteristics could be effectively mitigated.

5.2.2.

Denoising of low laser pulse energy

Clinical translation of high-speed OR-PAM requires safe laser energy levels as regulated by the American National Standards Institute (ANSI) safety standards. This necessitates a tradeoff between laser energy, imaging speed, and image quality. Conventionally, one may sacrifice laser pulse energy and thus image quality to maximize imaging speed. The work of Zhao et al.39 shows that deep learning can circumvent this tradeoff by improving the image quality caused by low laser energy under the ANSI limitations [Fig. 6(d)]. By integrating multi-supervised learning, dual-channel sample collection, and a reasonable weight distribution, the proposed deep learning model, known as multitask residual dense network (MT-RDN), was shown to have superior image quality, even with a 16 to 32-fold reduction in laser energy. With three subnetworks, the proposed model learned the multi-supervised task of simultaneous image denoising, super-resolution, and vascular enhancement. Zhao et al. also demonstrated the robustness of their technique with its ability to extrapolate to ear vasculature after only being trained on brain vasculature images.

6.

Conclusion and Outlook

Imaging speed is an essential priority of PAM when monitoring biological functions and capturing dynamic changes in biological processes. Due to the fast development of key technologies, including (1) light sources, (2) scanning systems, (3) detecting systems, and (4) computational techniques, PAM has experienced remarkable advances in terms of imaging speed in the last decade.

For the light source, the laser’s PRR plays a key role in determining the A-scan rate. Therefore, increasing the maximum PRR is an ideal approach for high-speed imaging. In PAM, even an MHz-level PRR is allowable because the target is located at a shallow depth. However, conventional PAM systems used slow bulky solid-state lasers utilizing the crystal as the gain medium, which was vulnerable to thermal lensing effect and thermal damage, thus limiting the input threshold power for a high PRR. Therefore, the emergence of fiber lasers has become a powerful driving force for ultrafast PAM. Because optical fibers have strong thermal resistance and are less affected by the thermal lensing effect, even commercial fiber lasers can produce the pulse energy of microjoules at an MHz-level PRR. Harnessing the development of commercial single-wavelength fiber lasers, techniques using nonlinear optical effects (e.g., SRS) have been utilized for multi-wavelength PA excitation, especially targeting oxy- and deoxy-hemoglobin. Consequentially, PAM using an MHz-level SRS source has enabled the monitoring of rapid sO2 changes, observing the initial dip from cerebral micro-vessels without repeated simulation, and achieving rapid whole-brain hemodynamics imaging. However, current light sources still have challenges for ultrafast PAM, such as the limited wavelength tunability and the low pulse energy. For example, when coupled at an MHz-level PRR, the optical fiber is prone to thermal damage, which prevents a stable output at tunable wavelengths. Further optimizing high-speed tunable lasers may focus on improving the fiber characteristics such as the effective area, nonlinear efficiency, and thermal resistance.

The development of scanning systems has proven to be an effective approach to optimizing the scanning speed over a large FOV. Mechanical scanning systems excel in high SNRs but often have low speed. Traditional optical scanning systems have high scanning speed while maintaining a good SNR, although they often have limitations in the FOV. Recent advancement of optical scanning systems has pushed the PAM speed limit to approximately the physical limit determined by ultrasound propagation. Various methods to detect the PA signals have been suggested as an alternative approach. One solution might be remote sensing technology that detects the local optical reflection change induced by an acoustic pressure rise.7981 The PA remote sensing does not rely on acoustic wave propagation and thus can achieve a higher signal detection speed. Optical sensors have emerged as a promising solution for achieving faster signal detection speeds, mostly coupled with fast optical scanning of the PA excitation beam. They offer several advantages, such as a compact size, wide reception angle, broad detection bandwidth, strong responsiveness in the low-frequency range, and seamless integration with the PA light path. However, optical sensors face challenges when it comes to multiplexing as even minor fabrication inaccuracies can significantly affect their performance. Another drawback is their susceptibility to environmental instability in biological settings. For example, micro-ring resonators can be sensitive to surface contamination, causing scattering and absorption losses, and FP interferometers can be affected by temperature variations, altering the properties of the polymer spacer. This instability in biological environments can lead to a rapid decline in sensor sensitivity and hinder their use in longitudinal in vivo studies.

With the introduction of ever more efficient computing resources over the past few years, there has been an explosion of new computational techniques to address the issues of spatial and temporal undersampling, as well as misalignment, distortion, and low laser energy correction. Hardware limitations, such as the limited laser PRR, have been transcended by upsampling the spatially/temporally undersampled PAM data, and already-fast hardware can overcome traditional sampling-imposed limitations by decreasing the sampling density required to achieve high image quality. Additionally, computational techniques have facilitated faster scanning mechanisms (e.g., polygon or MEMS scanners) by correcting the misalignment or distortion induced by rapid scanning speeds. Similarly, these techniques have enabled high-speed PAM while using laser pulse energy under the ANSI limit. Despite these many advances in the computational enhancement of PAM, there still exist significant hurdles regarding the need for large ground truth datasets. These datasets are often time consuming to gather, and ground truth can sometimes be difficult to access, especially in in vivo settings. Simulation or other synthetic data generation approaches have been used to address this issue. However, without appropriate augmentation and regularization techniques, there can exist a domain gap when applying deep learning models trained on synthetic datasets to improve experimentally acquired, complex in vivo images. Some steps have been taken to address this limitation by creating a large high-resolution open-source dataset,82 using experimentally-acquired ground-truth images,37,78 or adopting the realm of less data-hungry self-supervised or unsupervised deep learning methods.35 However, future work is still needed, likely by incorporating multiple approaches to address system specific-requirements with a fusion of hardware innovations and tailored computational interventions, to speed up PAM systems in a less data-hungry and more interpretable/experimentally-grounded way. It should be noted that the new computational, data-driven techniques also pose their own inevitable risks and challenges. One set of difficulties when using deep learning emerges from the nature of these “black-box” models as being fundamentally challenging to interpret.8385 This issue of interpretability is then exacerbated by the occasional pattern of deep learning models to add features not based in reality, so called “hallucinations,” when confronted with uncertain inputs.86 For example, when faced with corrupted input, such as aliased or noisy PAM data, it is possible for deep learning models to add vessels that do not exist in areas of model uncertainty. Finally, without the use of standardized, open-source, deep learning datasets and the release of code as open source, it can be difficult to build upon the work of other researchers and quantify method improvements.87 Many of these issues are still being addressed by deep learning practitioners, such as with more realistic simulation techniques, data augmentation, regularization, and model uncertainty quantification; thus this remain an area of growth for the field.

Overall, we expect that more efforts for implementing high-speed PAM will continue for monitoring the rapid dynamic changes in biomedical processes, enabled by further development of key technologies. Recognizing the current limitations, breaking through the drawbacks mentioned above, and adopting the optimal combination of each technology will lead to the realization of ultimate high-speed PAM for both fundamental research and clinical translation.

Disclosures

The authors have no relevant financial interests and no other potential conflicts of interest to disclose.

Code and Data Availability

Data sharing is not applicable to this article as no new data were created or analyzed.

Acknowledgments

This work was supported by the National Institutes of Health (Grant Nos. R21EB027981, R21 EB027304, RF1 NS115581, R01 NS111039, R01 EB028143, and R21 EB030072); National Science Foundation CAREER (Grant No. 2144788); Chan Zuckerberg Initiative (Grant No. 2020-226178); and National Research Foundation of Korea (Grant Nos. 2021R1A5A1032937 and 2022R1C1C2010976).

References

1. 

J. L. Su et al., “Advances in clinical and biomedical applications of photoacoustic imaging,” Expert Opin. Med. Diagn., 4 (6), 497 –510 https://doi.org/10.1517/17530059.2010.529127 (2010). Google Scholar

2. 

J. Xia, J. Yao and L. V. Wang, “Photoacoustic tomography: principles and advances,” Electromagn. Waves, 147 1 https://doi.org/10.2528/PIER14032303 (2014). Google Scholar

3. 

D. Das et al., “Another decade of photoacoustic imaging,” Phys. Med. Biol., 66 (5), 05TR01 https://doi.org/10.1088/1361-6560/abd669 (2021). Google Scholar

4. 

L. V. Wang and J. Yao, “A practical guide to photoacoustic tomography in the life sciences,” Nat. Methods, 13 (8), 627 –638 https://doi.org/10.1038/nmeth.3925 1548-7091 (2016). Google Scholar

5. 

J. Yao and L. V. Wang, “Photoacoustic microscopy,” Laser Photonics Rev., 7 (5), 758 –778 https://doi.org/10.1002/lpor.201200060 (2013). Google Scholar

6. 

S. Jeon et al., “Review on practical photoacoustic microscopy,” Photoacoustics, 15 100141 https://doi.org/10.1016/j.pacs.2019.100141 (2019). Google Scholar

7. 

K. Maslov et al., “Optical-resolution photoacoustic microscopy for in vivo imaging of single capillaries,” Opt. Lett., 33 (9), 929 –931 https://doi.org/10.1364/OL.33.000929 OPLEDP 0146-9592 (2008). Google Scholar

8. 

Z. Hosseinaee, J. A. Tummon Simmons and P. H. Reza, “Dual-modal photoacoustic imaging and optical coherence tomography,” Front. Phys., 8 616618 https://doi.org/10.3389/fphy.2020.616618 (2021). Google Scholar

9. 

S. Mirg et al., “Photoacoustic imaging for microcirculation,” Microcirculation, 29 (6–7), e12776 https://doi.org/10.1111/micc.12776 MCCRD8 0275-4177 (2022). Google Scholar

10. 

S.-W. Cho et al., “High-speed photoacoustic microscopy: a review dedicated on light sources,” Photoacoustics, 24 100291 https://doi.org/10.1016/j.pacs.2021.100291 (2021). Google Scholar

11. 

A. Agarwal et al., “Targeted gold nanorod contrast agent for prostate cancer detection by photoacoustic imaging,” J. Appl. Phys., 102 (6), 064701 https://doi.org/10.1063/1.2777127 JAPIAU 0021-8979 (2007). Google Scholar

12. 

M.-L. Li et al., “In-vivo photoacoustic microscopy of nanoshell extravasation from solid tumor vasculature,” J. Biomed. Opt., 14 (1), 010507 https://doi.org/10.1117/1.3081556 JBOPFO 1083-3668 (2009). Google Scholar

13. 

H. F. Zhang et al., “Functional photoacoustic microscopy for high-resolution and noninvasive in vivo imaging,” Nat. Biotechnol., 24 (7), 848 –851 https://doi.org/10.1038/nbt1220 NABIF9 1087-0156 (2006). Google Scholar

14. 

W. Shi et al., “Optical resolution photoacoustic microscopy using novel high-repetition-rate passively Q-switched microchip and fiber lasers,” J. Biomed. Opt., 15 (5), 056017 https://doi.org/10.1117/1.3502661 JBOPFO 1083-3668 (2010). Google Scholar

15. 

J. Yao et al., “High-speed label-free functional photoacoustic microscopy of mouse brain in action,” Nat. Methods, 12 (5), 407 –410 https://doi.org/10.1038/nmeth.3336 1548-7091 (2015). Google Scholar

16. 

T. J. Allen et al., “Ultrafast laser-scanning optical resolution photoacoustic microscopy at up to 2 million A-lines per second,” J. Biomed. Opt., 23 (12), 126502 https://doi.org/10.1117/1.JBO.23.12.126502 JBOPFO 1083-3668 (2018). Google Scholar

17. 

Y. Chang et al., “Co-impulse multispectral photoacoustic microscopy and optical coherence tomography system using a single supercontinuum laser,” Opt. Lett., 44 (18), 4459 –4462 https://doi.org/10.1364/OL.44.004459 OPLEDP 0146-9592 (2019). Google Scholar

18. 

Y. Liang et al., “2 MHz multi-wavelength pulsed laser for functional photoacoustic microscopy,” Opt. Lett., 42 (7), 1452 –1455 https://doi.org/10.1364/OL.42.001452 OPLEDP 0146-9592 (2017). Google Scholar

19. 

S.-W. Cho et al., “Optimal generation of ten individual green-to-red Raman source for wavelength-dependent real-time OR-PAM images,” IEEE J. Sel. Top. Quantum Electron., 25 (1), 1 –9 https://doi.org/10.1109/JSTQE.2018.2869646 IJSQEN 1077-260X (2018). Google Scholar

20. 

J. Chen et al., “Wide-field polygon-scanning photoacoustic microscopy of oxygen saturation at 1-MHz A-line rate,” Photoacoustics, 20 100195 https://doi.org/10.1016/j.pacs.2020.100195 (2020). Google Scholar

21. 

Y. He et al., “Wave of single-impulse-stimulated fast initial dip in single vessels of mouse brains imaged by high-speed functional photoacoustic microscopy,” J. Biomed. Opt., 25 (6), 066501 https://doi.org/10.1117/1.JBO.25.6.066501 JBOPFO 1083-3668 (2020). Google Scholar

22. 

X. Y. Zhu et al., “Real-time whole-brain imaging of hemodynamics and oxygenation at micro-vessel resolution with ultrafast wide-field photoacoustic microscopy,” Light-Sci. Appl., 11 (1), 138 https://doi.org/10.1038/s41377-022-00836-2 (2022). Google Scholar

23. 

P. Hajireza, A. Forbrich and R. Zemp, “In-vivo functional optical-resolution photoacoustic microscopy with stimulated Raman scattering fiber-laser source,” Biomed. Opt. Express, 5 (2), 539 –546 https://doi.org/10.1364/BOE.5.000539 BOEICL 2156-7085 (2014). Google Scholar

24. 

W. Shi et al., “In vivo near-realtime volumetric optical-resolution photoacoustic microscopy using a high-repetition-rate nanosecond fiber-laser,” Opt. Express, 19 (18), 17143 –17150 https://doi.org/10.1364/OE.19.017143 OPEXFF 1094-4087 (2011). Google Scholar

25. 

L. Wang et al., “Fast voice-coil scanning optical-resolution photoacoustic microscopy,” Opt. Lett., 36 (2), 139 –141 https://doi.org/10.1364/OL.36.000139 OPLEDP 0146-9592 (2011). Google Scholar

26. 

N. T. P. Truong et al., “Ultra-widefield photoacoustic microscopy with a dual-channel slider-crank laser-scanning apparatus for in vivo biomedical study,” Photoacoustics, 23 100274 https://doi.org/10.1016/j.pacs.2021.100274 (2021). Google Scholar

27. 

C. D. Ly et al., “Development of fast photoacoustic and ultrasound imaging system based on slider-crank scanner for small animals and humans study,” Expert Syst. Appl., 206 117939 https://doi.org/10.1016/j.eswa.2022.117939 ESAPEH 0957-4174 (2022). Google Scholar

28. 

K. Wang et al., “Recent advances in high-speed photoacoustic microscopy,” Photoacoustics, 24 100294 https://doi.org/10.1016/j.pacs.2021.100294 (2021). Google Scholar

29. 

J. Yao et al., “Wide-field fast-scanning photoacoustic microscopy based on a water-immersible MEMS scanning mirror,” J. Biomed. Opt., 17 (8), –080505 https://doi.org/10.1117/1.JBO.17.8.080505 JBOPFO 1083-3668 (2012). Google Scholar

30. 

J. Y. Kim et al., “Fast optical-resolution photoacoustic microscopy using a 2-axis water-proofing MEMS scanner,” Sci. Rep., 5 (1), 7932 https://doi.org/10.1038/srep07932 (2015). Google Scholar

31. 

M. Chen et al., “High-speed functional photoacoustic microscopy using a water-immersible two-axis torsion-bending scanner,” Photoacoustics, 24 100309 https://doi.org/10.1016/j.pacs.2021.100309 (2021). Google Scholar

32. 

J. Kim et al., “Super-resolution localization photoacoustic microscopy using intrinsic red blood cells as contrast absorbers,” Light Sci. Appl., 8 (1), 103 https://doi.org/10.1038/s41377-019-0220-4 (2019). Google Scholar

33. 

A. M. Winkler, K. Maslov and L. V. Wang, “Noise-equivalent sensitivity of photoacoustics,” J. Biomed. Opt., 18 (9), 097003 https://doi.org/10.1117/1.JBO.18.9.097003 JBOPFO 1083-3668 (2013). Google Scholar

34. 

A. DiSpirito et al., “Reconstructing Undersampled Photoacoustic Microscopy Images Using Deep Learning,” IEEE Trans. Med. Imaging, 40 (2), 562 –570 https://doi.org/10.1109/TMI.2020.3031541 ITMID4 0278-0062 (2021). Google Scholar

35. 

T. Vu et al., “Deep image prior for undersampling high-speed photoacoustic microscopy,” Photoacoustics, 22 100266 https://doi.org/10.1016/j.pacs.2021.100266 (2021). Google Scholar

36. 

J. Kim et al., “Deep learning acceleration of multiscale superresolution localization photoacoustic imaging,” Light: Sci. Appl., 11 (1), 131 https://doi.org/10.1038/s41377-022-00820-w (2022). Google Scholar

37. 

J. Kim et al., “Deep learning alignment of bidirectional raster scanning in high speed photoacoustic microscopy,” Sci. Rep., 12 (1), 16238 https://doi.org/10.1038/s41598-022-20378-2 (2022). Google Scholar

38. 

Y. Ma et al., “Spatial weight matrix in dimensionality reduction reconstruction for micro-electromechanical system-based photoacoustic microscopy,” Vis. Comput. Ind. Biomed. Art., 3 (1), 22 https://doi.org/10.1186/s42492-020-00058-6 (2020). Google Scholar

39. 

H. Zhao et al., “Deep learning enables superior photoacoustic imaging at ultralow laser dosages,” Adv. Sci., 8 (3), 2003097 https://doi.org/10.1002/advs.202003097 1936-6612 (2021). Google Scholar

40. 

V. Baryshnikov, N. Sarukura and Y. Segawa, Solid State Tunable Lasers with Nanosecond Flashlamp Pumping, SPIE PressBellingham, Washington, (1996). Google Scholar

41. 

H. Jelínková, Lasers for Medical Applications: Diagnostics, Therapy and Surgery, Elsevier( (2013). Google Scholar

42. 

Y. Wang and H. Kan, “Optimization algorithm for the pump structure of diode side-pumped solid-state lasers,” Opt. Lasers Eng., 45 (1), 93 –105 https://doi.org/10.1016/j.optlaseng.2006.06.005 OLENDN 0143-8166 (2007). Google Scholar

43. 

W. Koechner, Solid-State Laser Engineering, Springer( (2013). Google Scholar

44. 

G. Spühler et al., “Experimentally confirmed design guidelines for passively Q-switched microchip lasers using semiconductor saturable absorbers,” J. Opt. Soc. Am. B, 18 (6), 886 –886 https://doi.org/10.1364/JOSAB.18.000886 JOBPDE 0740-3224 (2001). Google Scholar

45. 

J. J. Zayhowski, “Passively Q-switched Nd: YAG microchip lasers and applications,” J. Alloys Compd., 303 393 –400 https://doi.org/10.1016/S0925-8388(00)00647-2 JALCEU 0925-8388 (2000). Google Scholar

46. 

T. Waritanant and A. Major, “Thermal lensing in Nd: YVO 4 laser with in-band pumping at 914 nm,” Appl. Phys. B, 122 1 –4 https://doi.org/10.1007/s00340-016-6417-9 (2016). Google Scholar

47. 

Y. N. Billeh, M. Liu and T. Buma, “Spectroscopic photoacoustic microscopy using a photonic crystal fiber supercontinuum source,” Opt. Express, 18 (18), 18519 –18524 https://doi.org/10.1364/OE.18.018519 OPEXFF 1094-4087 (2010). Google Scholar

48. 

C. Lee et al., “Combined photoacoustic and optical coherence tomography using a single near-infrared supercontinuum laser source,” Appl. Opt., 52 (9), 1824 –1828 https://doi.org/10.1364/AO.52.001824 APOPAI 0003-6935 (2013). Google Scholar

49. 

C. Lee et al., “In vitro photoacoustic measurement of hemoglobin oxygen saturation using a single pulsed broadband supercontinuum laser source,” Appl. Opt., 53 (18), 3884 –3889 https://doi.org/10.1364/AO.53.003884 APOPAI 0003-6935 (2014). Google Scholar

50. 

M. Bondu et al., “High energy supercontinuum sources using tapered photonic crystal fibers for multispectral photoacoustic microscopy,” J. Biomed. Opt., 21 (6), 061005 https://doi.org/10.1117/1.JBO.21.6.061005 JBOPFO 1083-3668 (2016). Google Scholar

51. 

M. K. Dasa et al., “All-fibre supercontinuum laser for in vivo multispectral photoacoustic microscopy of lipids in the extended near-infrared region,” Photoacoustics, 18 100163 https://doi.org/10.1016/j.pacs.2020.100163 (2020). Google Scholar

52. 

S.-W. Cho et al., “Efficient label-free in vivo photoacoustic imaging of melanoma cells using a condensed NIR-I spectral window,” Photoacoustics, 29 100456 https://doi.org/10.1016/j.pacs.2023.100456 (2023). Google Scholar

53. 

R. H. Silverman et al., “High-resolution photoacoustic imaging of ocular tissues,” Ultrasound Med. Biol., 36 (5), 733 –742 https://doi.org/10.1016/j.ultrasmedbio.2010.02.006 USMBA3 0301-5629 (2010). Google Scholar

54. 

C. Tian et al., “Novel photoacoustic microscopy and optical coherence tomography dual-modality chorioretinal imaging in living rabbit eyes,” J. Vis. Exp., 132 e57135 https://doi.org/10.3791/57135 (2018). Google Scholar

55. 

W. Qin et al., “Large-field-of-view optical resolution photoacoustic microscopy,” Opt. Express, 26 (4), 4271 –4278 https://doi.org/10.1364/OE.26.004271 OPEXFF 1094-4087 (2018). Google Scholar

56. 

C. Taboada et al., “Glassfrogs conceal blood in their liver to maintain transparency,” Science, 378 (6626), 1315 –1320 https://doi.org/10.1126/science.abl6620 SCIEAS 0036-8075 (2022). Google Scholar

57. 

E. Zhang, J. Laufer and P. Beard, “Backward-mode multiwavelength photoacoustic scanner using a planar Fabry-Perot polymer film ultrasound sensor for high-resolution three-dimensional imaging of biological tissues,” Appl. Opt., 47 (4), 561 –577 https://doi.org/10.1364/AO.47.000561 APOPAI 0003-6935 (2008). Google Scholar

58. 

N. Huynh et al., “Photoacoustic imaging using an 8-beam Fabry-Perot scanner,” Proc. SPIE, 9708 97082L https://doi.org/10.1117/12.2214334 PSISDG 0277-786X (2016). Google Scholar

59. 

Y. Liang et al., “Fast-scanning photoacoustic microscopy with a side-looking fiber optic ultrasound sensor,” Biomed. Opt. Express, 9 (11), 5809 –5816 https://doi.org/10.1364/BOE.9.005809 BOEICL 2156-7085 (2018). Google Scholar

60. 

M. Chen et al., “High-speed wide-field photoacoustic microscopy using a cylindrically focused transparent high-frequency ultrasound transducer,” Photoacoustics, 28 100417 https://doi.org/10.1016/j.pacs.2022.100417 (2022). Google Scholar

61. 

W. J. Westerveld et al., “Sensitive, small, broadband and scalable optomechanical ultrasound sensor in silicon photonics,” Nat. Photonics, 15 (5), 341 –345 https://doi.org/10.1038/s41566-021-00776-0 NPAHBY 1749-4885 (2021). Google Scholar

62. 

R. Shnaiderman et al., “A submicrometre silicon-on-insulator resonator for ultrasound detection,” Nature, 585 (7825), 372 –378 https://doi.org/10.1038/s41586-020-2685-y (2020). Google Scholar

63. 

G. Wissmeyer et al., “Looking at sound: optoacoustics with all-optical ultrasound detection,” Light: Sci. Appl., 7 (1), 53 https://doi.org/10.1038/s41377-018-0036-7 (2018). Google Scholar

64. 

Y. Li et al., “Snapshot photoacoustic topography through an ergodic relay for high-throughput imaging of optical absorption,” Nat. Photonics, 14 (3), 164 –170 https://doi.org/10.1038/s41566-019-0576-2 NPAHBY 1749-4885 (2020). Google Scholar

65. 

J. Xia et al., “Wide-field two-dimensional multifocal optical-resolution photoacoustic-computed microscopy,” Opt. Lett., 38 (24), 5236 –5239 https://doi.org/10.1364/OL.38.005236 OPLEDP 0146-9592 (2013). Google Scholar

66. 

T. Imai et al., “High-throughput ultraviolet photoacoustic microscopy with multifocal excitation,” J. Biomed. Opt., 23 (3), 036007 https://doi.org/10.1117/1.JBO.23.3.036007 JBOPFO 1083-3668 (2018). Google Scholar

67. 

C. Fang, H. Hu and J. Zou, “A focused optically transparent PVDF transducer for photoacoustic microscopy,” IEEE Sens. J., 20 (5), 2313 –2319 https://doi.org/10.1109/JSEN.2019.2952971 ISJEAZ 1530-437X (2020). Google Scholar

68. 

J. Park et al., “Quadruple ultrasound, photoacoustic, optical coherence, and fluorescence fusion imaging with a transparent ultrasound transducer,” Proc. Natl. Acad. Sci. U. S. A., 118 (11), e1920879118 https://doi.org/10.1073/pnas.1920879118 (2021). Google Scholar

69. 

Y. Hu et al., “Extended depth-of-field all-optical photoacoustic microscopy with a dual non-diffracting Bessel beam,” Opt. Lett., 44 (7), 1634 –1637 https://doi.org/10.1364/OL.44.001634 OPLEDP 0146-9592 (2019). Google Scholar

70. 

R. Cao et al., “Optical-resolution photoacoustic microscopy with a needle-shaped beam,” Nat. Photonics, 17 (1), 89 –95 https://doi.org/10.1038/s41566-022-01112-w NPAHBY 1749-4885 (2023). Google Scholar

71. 

Y. Zhou, N. Sun and S. Hu, “Deep learning-powered bessel-beam multiparametric photoacoustic microscopy,” IEEE Trans. Med. Imaging, 41 (12), 3544 –3551 https://doi.org/10.1109/TMI.2022.3188739 ITMID4 0278-0062 (2022). Google Scholar

72. 

T. Liu et al., “ADMM based low-rank and sparse matrix recovery method for sparse photoacoustic microscopy,” Biomed. Signal Process. Control, 52 14 –22 https://doi.org/10.1016/j.bspc.2019.03.007 (2019). Google Scholar

73. 

T. Zhou and D. Tao, “Godec: randomized low-rank & sparse matrix decomposition in noisy case,” in Proc. 28th Int. Conf. Mach. Learn., ICML, (2011). Google Scholar

74. 

S. G. Sathyanarayana et al., “Simultaneous dictionary learning and reconstruction from subsampled data in photoacoustic microscopy,” in IEEE Int. Ultrasonics Symp. (IUS), 483 –486 (2019). https://doi.org/10.1109/ULTSYM.2019.8925747 Google Scholar

75. 

S. G. Sathyanarayana et al., “Recovery of blood flow from undersampled photoacoustic microscopy data using sparse modeling,” IEEE Trans. Med. Imaging, 41 (1), 103 –120 https://doi.org/10.1109/TMI.2021.3104521 ITMID4 0278-0062 (2022). Google Scholar

76. 

I. G. E. Sulistyawan et al., “Utilization of curvelet transform in reconstructing cellular images for undersampled optical-resolution photoacoustic microscopy,” IEEE Trans. Ultrasonics Ferroelectr. Freq. Control, 70 1631 –1641 https://doi.org/10.1109/TUFFC.2023.3272917 (2023). Google Scholar

77. 

D. Seong et al., “Three-dimensional reconstructing undersampled photoacoustic microscopy images using deep learning,” Photoacoustics, 29 100429 https://doi.org/10.1016/j.pacs.2022.100429 (2023). Google Scholar

78. 

J. Zhou et al., “Photoacoustic microscopy with sparse data by convolutional neural networks,” Photoacoustics, 22 100242 https://doi.org/10.1016/j.pacs.2021.100242 (2021). Google Scholar

79. 

P. Hajireza et al., “Non-interferometric photoacoustic remote sensing microscopy,” Light: Sci. Appl., 6 (6), e16278 https://doi.org/10.1038/lsa.2016.278 (2017). Google Scholar

80. 

K. Bell et al., “Reflection-mode virtual histology using photoacoustic remote sensing microscopy,” Sci. Rep., 10 (1), 19121 https://doi.org/10.1038/s41598-020-76155-6 (2020). Google Scholar

81. 

S. Abbasi, K. Bell and P. Haji Reza, “Rapid high-resolution mosaic acquisition for photoacoustic remote sensing,” Sensors, 20 (4), 1027 https://doi.org/10.3390/s20041027 SNSRES 0746-9462 (2020). Google Scholar

82. 

, “Duke PAM Dataset,” https://doi.org/10.5281/zenodo.4042171 (). Google Scholar

83. 

A. Hauptmann and B. Cox, “Deep learning in photoacoustic tomography: current approaches and future directions,” J. Biomed. Opt., 25 (11), 112903 https://doi.org/10.1117/1.JBO.25.11.112903 JBOPFO 1083-3668 (2020). Google Scholar

84. 

J. Gröhl et al., “Deep learning for biomedical photoacoustic imaging: a review,” Photoacoustics, 22 100241 https://doi.org/10.1016/j.pacs.2021.100241 (2021). Google Scholar

85. 

C. Yang et al., “Review of deep learning for photoacoustic imaging,” Photoacoustics, 21 100215 https://doi.org/10.1016/j.pacs.2020.100215 (2021). Google Scholar

86. 

H. Deng et al., “Deep learning in photoacoustic imaging: a review,” J Biomed Opt, 26 (4), 040901 https://doi.org/10.1117/1.JBO.26.4.040901 JBOPFO 1083-3668 (2021). Google Scholar

87. 

P. Rajendran, A. Sharma and M. Pramanik, “Photoacoustic imaging aided with deep learning: a review,” Biomed. Eng. Lett., 12 155 –173 https://doi.org/10.1007/s13534-021-00210-y (2022). Google Scholar

Biography

Soon-Woo Cho completed her PhD in the Department of Cogno-Mechatronics Engineering at Pusan National University, South Korea, in 2022. She is affiliated at Engineering Research Center for Color-modulated Extra-sensory Perception Technology, Pusan National University, and currently doing research at Duke University as a postdoctoral researcher. Her current research interest is the development of novel light sources for optical imaging systems, especially multimodal biomedical imaging systems.

Van Tu Nguyen obtained his BS degree in biomedical engineering from Hanoi University of Science and Technology, Vietnam in 2015. He received his MS and PhD degrees in biomedical engineering from Pukyong National University, Busan, South Korea, in 2018 and 2022, respectively. He currently joins as a post-doc fellow at Biomedical Engineering Department, Duke University, Durham, North Carolina, United States. His research focuses on biomedical imaging methods, including photoacoustic imaging and fluorescence imaging.

Anthony DiSpirito is a biomedical engineering PhD student at Duke University. He received his BSE degree in biomedical engineering from Case Western Reserve University with a focus in computational modeling, computer science, and biomedical imaging. He received his MS degree in biomedical engineering from Duke University, with a focus in artificial intelligence and biomedical imaging. As a PhD student at Duke’s Photoacoustic Imaging Lab, his research interests focus on the application of machine learning in biomedical imaging.

Joseph Yang is a biomedical engineering PhD student at Duke University. He completed both his MEng and BEng degrees at Stevens Institute of Technology in biomedical engineering. At Duke’s Photoacoustic Imaging Lab, his current research interests are in deep-tissue imaging, especially regarding the use of CEUS and PACT.

Chang-Seok Kim received his PhD from The Johns Hopkins University, Baltimore, Maryland, United States, in 2004. He is a professor at the Department of Optics and Mechatronics Engineering and the Department of Cogno-Mechatronics Engineering, Pusan National University, Busan, South Korea. His current research interests include novel fiber laser systems and their application in biomedical, telecommunication, and sensor areas.

Junjie Yao is associate professor of biomedical engineering at Duke University. He completed his PhD in biomedical engineering at Washington University in St. Louis in 2013. His research involves the integration of light and sound to enable high-speed functional brain imaging, deep-tissue molecular imaging, and early-stage cancer detection.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Soon-Woo Cho, Van Tu Nguyen, Anthony DiSpirito, Joseph Yang, Chang-Seok Kim, and Junjie Yao "Sounding out the dynamics: a concise review of high-speed photoacoustic microscopy," Journal of Biomedical Optics 29(S1), S11521 (5 February 2024). https://doi.org/10.1117/1.JBO.29.S1.S11521
Received: 13 October 2023; Accepted: 11 January 2024; Published: 5 February 2024
Advertisement
Advertisement
KEYWORDS
Fiber lasers

Biomedical optics

Deep learning

Imaging systems

Microelectromechanical systems

Signal detection

Pulsed laser operation

Back to Top