Although multilayer analytical models have been proposed to enhance brain sensitivity of diffuse correlation spectroscopy (DCS) measurements of cerebral blood flow, the traditional homogeneous model remains dominant in clinical applications. Rigorous in vivo comparison of these analytical models is lacking.

Aim

We compare the performance of different analytical models to estimate a cerebral blood flow index (CBFi) with DCS in adults.

Approach

Resting-state data were obtained on a cohort of 20 adult patients with subarachnoid hemorrhage. Data at 1 and 2.5 cm source-detector separations were analyzed with the homogenous, two-layer, and three-layer models to estimate scalp blood flow index and CBFi. The performance of each model was quantified via fitting convergence, fit stability, brain-to-scalp flow ratio (BSR), and correlation with transcranial Doppler ultrasound (TCD) measurements of cerebral blood flow velocity in the middle cerebral artery (MCA).

Results

The homogeneous model has the highest pass rate (100%), lowest coefficient of variation (CV) at rest (median [IQR] at 1 Hz of 0.18 [0.13, 0.22]), and most significant correlation with MCA blood flow velocities (Rs = 0.59, p = 0.010) compared with both the two- and three-layer models. The multilayer model pass rate was significantly correlated with extracerebral layer thicknesses. Discarding datasets with non-physiological BSRs increased the correlation between DCS measured CBFi and TCD measured MCA velocities for all models.

Conclusions

We found that the homogeneous model has the highest pass rate, lowest CV at rest, and most significant correlation with MCA blood flow velocities. Results from the multilayer models should be taken with caution because they suffer from lower pass rates and higher coefficients of variation at rest and can converge to non-physiological values for CBFi. Future work is needed to validate these models in vivo, and novel approaches are merited to improve the performance of the multimodel models.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 International License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Hongting Zhao, Eashani Sathialingam, Kyle R. Cowdrick, Tara Urner, Seung Yup Lee, Shasha Bai, Feras Akbik, Owen B. Samuels, Prem Kandiah, Ofer Sadan, and Erin M. Buckley "Comparison of diffuse correlation spectroscopy analytical models for measuring cerebral blood flow in adults," Journal of Biomedical Optics 28(12), 126005 (15 December 2023). https://doi.org/10.1117/1.JBO.28.12.126005
Received: 19 August 2023; Accepted: 21 November 2023; Published: 15 December 2023
Advertisement
Advertisement
KEYWORDS
Analytic models

Data modeling

Blood circulation

Performance modeling

Skull

Spectroscopy

Brain

Back to Top