5 August 2019 Propagating-path uniformly scanned light sheet excitation microscopy for isotropic volumetric imaging of large specimens
Author Affiliations +
Abstract

We demonstrate a propagating-path uniformly scanned light sheet excitation (PULSE) microscopy based on the oscillation of voice coil motor that can rapidly drive a thin light sheet along its propagation direction. By synchronizing the rolling shutter of a camera with the motion of laser sheet, we can obtain a uniform plane-illuminated image far beyond the confocal range of Gaussian beam. A stable 1.7-μm optical sectioning under a 3.3  mm  ×  3.3  mm wide field of view (FOV) has been achieved for up to 20 Hz volumetric imaging of large biological specimens. PULSE method transforms the extent of plane illumination from one intrinsically limited by the short confocal range (μm scale) to one defined by the motor oscillation range (mm scale). Compared to the conventional Gaussian light sheet imaging, our method greatly mitigates the compromise of axial resolution and successfully extends the FOV over 100 times. We demonstrate the applications of PULSE method by rapidly imaging cleared mouse spinal cord and live zebrafish larva at isotropic subcellular resolution.

In the past decade, light sheet fluorescence microscopy has become an emerging technique for three-dimensional (3-D) imaging of thick samples at high speed and low photo toxicity.19 For a conventional focused Gaussian sheet, a conflict always exists between the beam confocal range and its axial extent, thereby causing a trade-off between high axial resolution and large field of view (FOV). One possible approach is to iteratively illuminate the sample and stitch these tiles together.10,11 But the mechanical stitching greatly limits the throughput. Recently, light sheet combined with nondiffracting beams, such as Airy beam6,12 or Bessel beam,3,13,14 can generate a thin light sheet over a long distance, which mitigates the conflict of axial resolution and FOV. However, compared to Gaussian beam that focuses most of photons into the beam waist, these nondiffracting beams have more dispersed energy at the side lobes. Though a few techniques such as confocal virtual slit15,16 have been developed to eliminate the fluorescence emission by the side lobes’ excitation, the specimen still receives the photon burden from the side lodes. More complicated modalities, such as two-photon Bessel light sheet3,17,18 and lattice light sheet,14 are reported to physically suppress the side lodes, but these systems are in a more complex format and at a higher cost.

An alternative approach is to sweep a thin Gaussian light sheet along the propagation direction to dynamically extend the confocal range of illumination. Some special elements, such as a mirror based on a piezo stage or voice coil motor (VCM),19,20 electrical tunable lens (ETL),2124 or tunable acoustic gradient lens25,26 are combined with confocal virtual slit to achieve this goal. For example, with using an ASLM the extended confocal depth can reach 162  μm with an axial resolution of 390 nm.19 The use of ETL, which has a larger adjustable range than a piezo scanner, can further extend the confocal range to 1  mm with 4 to 6  μm resolution.23 However, for these methods, the beam scanning is performed before the illumination objective. The convergence angle of incident light entering the rear pupil of illuminating objective varies during the light-sheet sweeping. Thus, it is difficult for these methods to maintain a uniform light-sheet thickness at very large scanning range of several millimeters.24

Here, we propose a propagating-path uniformly scanned light sheet excitation (PULSE) method via directly oscillating the illumination objective by a VCM. By synchronization of the VCM-driven light-sheet motion and the confocal slit of an sCMOS camera, we can achieve a uniform light sheet sectioning of 1.7  μm over a large FOV of 3.3  mm×3.3  mm in single acquired frame. At the same time, the VCM can push a speed limit of 20 frames per second. We successfully apply this technique to the volumetric imaging of mouse spinal cord and live zebrafish larva, proving isotropic subcellular resolution across entire samples.

In conventional LSFM, the incident Gaussian beam is focused in one dimension and forms a thin laser sheet. The axial extent of laser sheet, which usually determines the system axial resolution, can be defined as the diameter of beam waist d (FWHM value). According to the Gaussian beam propagation theory,27 d can be calculated as

d=2ln2λfπω0,
where f is the focal length, λ is the excitation wavelength, and ω0 is the radius of the incident beam. Meanwhile, the confocal range, which is appropriate for imaging, can be expressed as
l=πd2ln2λ=4λf2ln2πω02.

It is obvious that as the thickness of laser sheet goes smaller (higher axial resolution), the illumination range (FOV) decreases drastically. As a reference point, a 2-μm-thick light-sheet only yields 38-μm confocal range, which is only suited for imaging single cell. Therefore, imaging millimeter-size large samples at single-cell axial resolution becomes a choice dilemma of either tens of microns low axial resolution or slow stepwise mosaic stitching.

In our PULSE setup, the incident beam was line-focused first by a cylindrical lens (Thorlabs, LJ1695RM) and then formed the illuminating laser sheet through a long working-distance objective (Mitutoyo, 10×/0.28  NA). The objective was mounted steadily on a VCM (TMEC0050-025-000) for high-speed oscillation under a millimeter-scale large amplitude, inducing an oscillating light-sheet covering the entire FOV [Fig. 1(a)]. In the static mode, the thin light sheet measured 1.7  μm only covers a small imaging area 28  μm. Under 4× objective detection, the quality of acquired image degrades seriously beyond this range [Fig. 1(h)]. Through controlling the rolling shutter of an sCMOS camera (Hamamatsu. ORCA-Flash 4.0 V2), we formed a linear array that contains 15 to 20 rows of active pixels sweeping at the sensor plane and tightly synchronized it with the VCM-driven movement of the Gaussian sheet waist [Figs. 1(b), 1(e), and 1(f)] using a customized Labview program. This dynamic synchronization mode transformed the extent of plane illumination from one limited by the short confocal range to one defined by the motor oscillation range. The formed confocal electronic slit further blocked the fluorescence resides excited by the diverged laser out of the confocal range. Thus, we can obtain sharp plane-excited fluorescence signals in single wide-FOV frame recorded by 4× objective [Fig. 1(g)]. Then through a z-scan controlled by another independent actuator (Thorlabs, Z825B), a 3-D stack can be obtained at a high speed of tens of PULSE frames per second. Video 1 [Fig. 1(d)] shows the process of light-sheet moving and synchronizing with the camera. The photograph of VCM mounted with objective is shown in Fig. 1(c).

Fig. 1

(a) Schematic of the propagational light sheet scan using VCM. The confocal range of the laser sheet rapidly moving along the x direction. (b) Control timing diagram of the VCM motion and rolling shutter of the camera. (c) Device photo of VCM mounted with objective. (d) Through the synchronization, a virtual confocal slit is formed to exclusively collect the fluorescence excited by the Gaussian sheet waist (Video 1), as shown in (e) and compared to conventional static light-sheet geometry in (f). (g), (h) The plane images acquired under PULSE mode and conventional mode, respectively. (Video 1, MPEG, 5.47 MB [URL: https://doi.org/10.1117/1.JBO.24.8.086501.1]).

JBO_24_8_086501_f001.png

Compared to propagational scanning using ETL,22 scanning the laser sheet by directly driving the illuminating lens can generate a more uniform optical sectioning when the sample is across a large range of several millimeters. We used ZEMAX software to simulate the profiles of the focused Gaussian sheet generated by PULSE and ETL methods (EL-10-30-VIS-LD, Optotune, Switzerland). PULSE method maintained a very uniform laser sheet thickness of 1.7  μm after the illuminating objective traveling a 3.3-mm distance along propagation direction [Fig. 2(a), bottom row and Fig. 2(b), orange line]. In contrast, the thickness of light sheet would varies over 35% (from 1.7 to 2.3  μm) via using the ETL to shift the focal point of Gaussian sheet with the same 3.3 mm distance [Fig. 2(a), top row and blue line]. This nonuniform plane illumination would result suboptimal contrast and resolution while imaging large samples.

Fig. 2

(a) The beam profile at different propagation position adjusted by ETL (top row) and VCM (bottom row). (b) The thickness of light sheet at different focusing position by ETL (blue line) and VCM (orange line).

JBO_24_8_086501_f002.png

We used fluorescent beads (500  nm, Lumisphere) as point source to experimentally characterize our system. We first evaluated the excitation efficiency of PULSE via comparing it with conventional static plane illumination set-ups. A 10X/0.4 detection objective was used for collecting the fluorescence signals excited by the laser sheets of different thickness. The narrow confocal range of 1.7  μm sheet was measured around 28  μm in static mode [Fig. 3(a)]. The confocal virtual slit of the camera was accordingly set to 40-pixel lines (26  μm) under PULSE mode, with rolling speed over 10 full frames per seconds synchronized with the VCM objective scanning. The acquired plane image of the beads is shown in Fig. 3(b). Due to the more intensive energy condensed at the beam waist, the thinner light sheet enables stronger fluorescence emission and higher signal contrast as compared to the thicker light sheets [3 and 8  μm in Figs. 3(c) and 3(d)]. Meanwhile, comparing the PULSE and static excitation, the application of the confocal virtual slit does not affect the emission intensity, generating uniform and strong signals across the entire frame [Figs. 3(e) and 3(f)].

Fig. 3

(a)–(d) The signals distribution of fluorescent beads acquired using 1.7-μm static sheet, 1.7-μm PULSE, 3-μm static sheet, and 8-μm static sheet, respectively. (e), (f) Comparison of peak and averaged intensities of six selected signals shown in (a)–(d). Scale bars: 50  μm in (a)–(d) and 3  μm in insets.

JBO_24_8_086501_f003.png

Then, a lower magnification of 4× detection was used to compare the effective imaging FOV under PULSE and conventional static models. With obtaining the z-stacks using both methods, seven reconstructed beads at different x positions were accordingly extracted from both setups to analyze their axial performance [Figs. 4(c) and 4(d)]. The measured axial extents of the beads from conventional light sheet shows over 10-times divergence (1.7 to 20  μm) within a 500-μm short propagation. This is in significant contrast with very sharp-and-uniform optical sectioning of PULSE (1.7 to 1.8  μm) across entire 3.3-mm FOV. Our method obviously mitigates the compromise of axial resolution while greatly extends the effective FOV over 100 times.

Fig. 4

(a), (b) The reconstructed xz planes of fluorescent beads imaged by a 1.7-μm thin light sheet under conventional and PULSE modes. (c), (d) Vignette high-resolution views of seven selected beads at different x positions in (a) and (b). (e) The variations of axial intensity profiles of the selected beads in (c) and (d). Scale bars: 200  μm in (a), (b) and 5  μm in (c), (d).

JBO_24_8_086501_f004.png

We demonstrated PULSE’s ability via three-dimensionally imaging fluorescence-tagged mouse spinal cord (thy1-GFP) and zebrafish larva (Tg flk1: mCherry). The PULSE setup was identical with that described above, and a 10×/0.4 apochromatic objective lens was used to collect fluorescent signals at adequate lateral resolution. Figure 5(a) showed the 3-D reconstruction of spinal cord by PULSE microscopy. The center of confocal range is around x=700  μm. Three yz planes with 200-μm interval in propagation direction (x) were extracted and shown in Fig. 5(b). The same yz sample planes from the 3-D reconstructions by 1.7 and 8  μm static laser sheet were shown in Figs. 5(c) and 5(d) as comparison. Due to the tradeoff between the axial resolution and the FOV, 8-μm laser sheet has a better over performance across the large scale [Figs. 5(d2), 5(d3)] while the 1.7-μm sheet shows thinner optical sectioning ability at the narrow beam waist [Fig. 5(c1)]. In consistence with the PSF test, static excitation mode caused either signal deterioration out of small confocal range [Figs. 5(c2), 5(c3)] or insufficient overall resolution [Fig. 5(d)] while PULSE showed high-and-uniform axial resolution capable of discerning neural fibers in both planes. In Fig. 5(e), the 3-D visualization of the vascular system in live zebrafish larva also verified an isotropic resolution has been achieved in all transverse (xy), coronal (xz), and sagittal (yz) views.

Fig. 5

(a) The 3-D reconstruction of cleared mouse spinal cord with neurons labeled by GFP (thy1-GFP). (b) Three reconstructed yz planes with 200  μm interval in the propagation direction (x), showing the uniform isotropic resolution spanning a wide view. These results are further compared with the images by 1.7 and 8  μm static sheets, as shown in (c), (d), respectively. Insets in (b)–(d), line intensity profiles of the neuron fibers, indicating the different resolving power of three methods. (e) PULSE imaging of a blood vessel-tagged zebrafish larval (Tg flk1: mCherry). (e1)–(e3), The xy, xz, and yz views of the anterior part of the zebrafish. (e4) The 3-D volume rendering of the zebrafish. Scale bars: 50  μm in (b)–(d) and 100  μm in (e).

JBO_24_8_086501_f005.png

In summary, we have reported the PULSE microscopy method, which is based on the synchronization of VCM-driven light-sheet oscillation and the rolling virtual slit of the sCMOS camera. PULSE method greatly extends the FOV of conventional Gaussian light-sheet imaging by two orders while maintaining a thin-and-uniform optical sectioning. As a demonstration, PULSE can readily provide a gigavoxel image of millimeter-thick samples, such as mouse spine cord and zebrafish larval, with isotropic subcellular resolution achieved at a speed up to 20 fps. The highest frame rate of PULSE is currently limited by the load of VCM and its large travel range. We believe this speed limitation could be overcome via further optimizing the design. By choosing VCM with a better actuator, improving the mechanical design, and integration of vibration isolator, it is possible to obtain a higher scanning frequency well matched to the frame rate of the camera. The advancements of simplified setup and thin-and-uniform illumination across ultralarge range render the PULSE method a promising tool for a variety of biomedical applications, such as embryo development, tissue histology, pathology, and neuron science research, in which high-resolution cellular profiling at whole-tissue scale is highly desired.

Disclosures

The authors have no relevant financial interests in the manuscript.

Acknowledgments

The authors acknowledge Yusha Li and Jiayi Tu for providing the mouse and zebrafish samples and funding support from National Key R&D Program of China (P.F., 2017YFA0700500), the National Natural Science Foundation of China (21874052), Research Program of Shenzhen (P.F., JCYJ20160429182424047), and 1000 Youth Talents Plan of China (P.F.).

References

1. 

J. Huisken et al., “Optical sectioning deep inside live embryos by selective plane illumination microscopy,” Science, 305 1007 –1009 (2004). https://doi.org/10.1126/science.1100035 SCIEAS 0036-8075 Google Scholar

2. 

P. J. Keller et al., “Reconstruction of zebrafish early embryonic development by scanned light sheet microscopy,” Science, 322 1065 –1069 (2008). https://doi.org/10.1126/science.1162493 SCIEAS 0036-8075 Google Scholar

3. 

T. A. Planchon et al., “Rapid three-dimensional isotropic imaging of living cells using Bessel beam plane illumination,” Nat. Methods, 8 417 –423 (2011). https://doi.org/10.1038/nmeth.1586 1548-7091 Google Scholar

4. 

U. Krzic et al., “Multiview light-sheet microscope for rapid in toto imaging,” Nat. Methods, 9 730 –733 (2012). https://doi.org/10.1038/nmeth.2064 1548-7091 Google Scholar

5. 

M. B. Ahrens et al., “Whole-brain functional imaging at cellular resolution using light-sheet microscopy,” Nat. Methods, 10 413 –420 (2013). https://doi.org/10.1038/nmeth.2434 1548-7091 Google Scholar

6. 

T. Vettenburg et al., “Light-sheet microscopy using an Airy beam,” Nat. Methods, 11 541 –544 (2014). https://doi.org/10.1038/nmeth.2922 1548-7091 Google Scholar

7. 

S. Wolf et al., “Whole-brain functional imaging with two-photon light-sheet microscopy,” Nat. Methods, 12 379 –380 (2015). https://doi.org/10.1038/nmeth.3371 1548-7091 Google Scholar

8. 

J. Lee et al., “4-Dimensional light-sheet microscopy to elucidate shear stress modulation of cardiac trabeculation,” J. Clin Invest., 126 1679 –1690 (2016). https://doi.org/10.1172/JCI83496 Google Scholar

9. 

R. M. Power and J. Huisken, “A guide to light-sheet fluorescence microscopy for multiscale imaging,” Nat. Methods, 14 360 –373 (2017). https://doi.org/10.1038/nmeth.4224 1548-7091 Google Scholar

10. 

L. Gao, “Extend the field of view of selective plan illumination microscopy by tiling the excitation light sheet,” Opt. Express, 23 6102 –6111 (2015). https://doi.org/10.1364/OE.23.006102 OPEXFF 1094-4087 Google Scholar

11. 

Q. Fu et al., “Imaging multicellular specimens with real-time optimized tiling light-sheet selective plane illumination microscopy,” Nat. Commun., 7 11088 (2016). https://doi.org/10.1038/ncomms11088 NCAOBW 2041-1723 Google Scholar

12. 

G. Siviloglou et al., “Observation of accelerating Airy beams,” Phys. Rev. Lett., 99 213901 (2007). https://doi.org/10.1103/PhysRevLett.99.213901 PRLTAO 0031-9007 Google Scholar

13. 

L. Gao et al., “Noninvasive imaging beyond the diffraction limit of 3D dynamics in thickly fluorescent specimens,” Cell, 151 1370 –1385 (2012). https://doi.org/10.1016/j.cell.2012.10.008 CELLB5 0092-8674 Google Scholar

14. 

B.-C. Chen et al., “Lattice light-sheet microscopy: imaging molecules to embryos at high spatiotemporal resolution,” Science, 346 1257998 (2014). https://doi.org/10.1126/science.1257998 SCIEAS 0036-8075 Google Scholar

15. 

E. Baumgart and U. Kubitscheck, “Scanned light sheet microscopy with confocal slit detection,” Opt. Express, 20 21805 –21814 (2012). https://doi.org/10.1364/OE.20.021805 OPEXFF 1094-4087 Google Scholar

16. 

F. O. Fahrbach and A. Rohrbach, “Propagation stability of self-reconstructing Bessel beams enables contrast-enhanced imaging in thick media,” Nat. Commun., 3 632 (2012). https://doi.org/10.1038/ncomms1646 NCAOBW 2041-1723 Google Scholar

17. 

F. O. Fahrbach et al., “Light-sheet microscopy in thick media using scanned Bessel beams and two-photon fluorescence excitation,” Opt. Express, 21 13824 –13839 (2013). https://doi.org/10.1364/OE.21.013824 OPEXFF 1094-4087 Google Scholar

18. 

M. Zhao et al., “Cellular imaging of deep organ using two-photon Bessel light-sheet nonlinear structured illumination microscopy,” Biomed. Opt. Express, 5 1296 –1308 (2014). https://doi.org/10.1364/BOE.5.001296 BOEICL 2156-7085 Google Scholar

19. 

K. M. Dean et al., “Deconvolution-free subcellular imaging with axially swept light sheet microscopy,” Biophys. J., 108 2807 –2815 (2015). https://doi.org/10.1016/j.bpj.2015.05.013 BIOJAU 0006-3495 Google Scholar

20. 

T. Chakraborty et al., “Light-sheet microscopy with isotropic, submicron resolution and solvent-independent large-scale imaging,” (2019). Google Scholar

21. 

A. K. Chmielewski et al., “Fast imaging of live organisms with sculpted light sheets,” Sci. Rep., 5 9385 (2015). https://doi.org/10.1038/srep09385 SRCEC3 2045-2322 Google Scholar

22. 

P. N. Hedde and E. Gratton, “Selective plane illumination microscopy with a light sheet of uniform thickness formed by an electrically tunable lens,” Microsc. Res. Tech., 81 924 –928 (2018). https://doi.org/10.1002/jemt.v81.9 MRTEEO 1059-910X Google Scholar

23. 

B. Migliori et al., “Light sheet theta microscopy for rapid high-resolution imaging of large biological samples,” BMC Biol., 16 57 (2018). https://doi.org/10.1186/s12915-018-0521-8 1741-7007 Google Scholar

24. 

F. F. Voigt et al., “The mesoSPIM initiative: open-source light-sheet mesoscopes for imaging in cleared tissue,” (2019). Google Scholar

25. 

K. M. Dean and R. Fiolka, “Uniform and scalable light-sheets generated by extended focusing,” Opt. Express, 22 26141 –26152 (2014). https://doi.org/10.1364/OE.22.026141 OPEXFF 1094-4087 Google Scholar

26. 

W. Zong et al., “Large-field high-resolution two-photon digital scanned light-sheet microscopy,” Cell Res., 25 254 –257 (2015). https://doi.org/10.1038/cr.2014.124 1001-0602 Google Scholar

27. 

M. Born and E. Wolf, Principles of Optics, Cambridge University Press, Cambridge (1999). Google Scholar

Biographies of the authors are not available.

CC BY: © The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License. Distribution or reproduction of this work in whole or in part requires full attribution of the original publication, including its DOI.
Junyu Ping, Fang Zhao, Jun Nie, Tingting Yu, Dan Zhu, Mugen Liu, and Peng Fei "Propagating-path uniformly scanned light sheet excitation microscopy for isotropic volumetric imaging of large specimens," Journal of Biomedical Optics 24(8), 086501 (5 August 2019). https://doi.org/10.1117/1.JBO.24.8.086501
Received: 4 June 2019; Accepted: 15 July 2019; Published: 5 August 2019
Lens.org Logo
CITATIONS
Cited by 8 scholarly publications and 2 patents.
Advertisement
Advertisement
KEYWORDS
Confocal microscopy

Microscopy

Image resolution

Cameras

Biomedical optics

Camera shutters

Luminescence

Back to Top