1.IntroductionFluorescence molecular tomography (FMT) has become a promising optical imaging modality that allows studying disease noninvasively and quantitatively at the molecular level.1–3 It can be applied to drug discovery and preclinical oncological research.3–5 The reconstruction algorithm is crucial to recover the location of fluorophore, the fluorescence yield, and the images of the fluorescent lifetime. However, the solution to the inverse problem of FMT suffers from being ill-posed due to the diffusive behavior of light in tissue. In order to obtain meaningful and stable results, it is essential to incorporate adequate a priori information, such as the anatomical information provided by computed tomography (CT) or magnetic resonance imaging,6–8 and the distribution of optical properties obtained by diffuse optical tomography (DOT).7,9 Several multimodality FMT systems have been developed.6–9 In the reconstruction algorithms, the regularization technique is commonly used in the form of a penalty function to improve the solution. In the early stage of research, the Tikhonov regularization, namely -norm regularization, is routinely employed for the inverse problem.10,11 However, this method tends to create an oversmooth effect and generate spurious small elements, which produce noisy background when the inverse problem is severely ill-posed. Recently, the sparsity regularization, with -norm penalty function, has received considerable attention due to its good performance in FMT. In the application of early detection of a tumor, the observed biological mechanisms are generally confined to localized regions. In this case, the interested biological function is somewhat “sparse,” which can be utilized as a priori information in the form of -norm penalty.12,13 However, for the application of reconstructing the distribution of the fluorescent dye or drugs in organs, the spatial distribution of biological activity usually occurs in a large area. As a consequence, -norm regularization would be no longer suitable for this problem, while -norm regularization is potentially beneficial. In this paper, we intend to fill the gap in the existing studies to systematically show the performance of -norm and -norm regularization methods for the two imaging models of FMT using a fair and consistent experimental setting. The first imaging model is the reconstruction of fluorophore with small size, which is to mimic the early detection of tumor. The second model is the reconstruction of the distribution of fluorescent dye in organs, which is essential to drug pharmacokinetics (PK).14 In order to evaluate the performance of the two regularization methods, apart from numerical simulation experiments, in vivo experiments were also performed. A homemade dual-modality FMT/micro-computed tomography (micro-CT) imaging system is used as the experimental setup. The outline of this paper is as follows. In Sec. 2, the methods, models, and system setup are detailed. In Sec. 3, the simulation and in vivo experiments are shown, and the results are described. Finally, we discuss and make conclusions of this paper. 2.Methodologies2.1.Establishment of the System EquationSince practical FMT operates in the wavelength range of the near-infrared, scattering predominates in the light–matter interaction as light propagates through biological tissues.15 The diffusion equation is sufficiently accurate for the light propagation.16,17 Here, we only reconstructed the fluorophore distribution in the tissue and the optical properties are known a priori. The Robin boundary condition is performed on the boundary.18 When FMT is excited by a continuous wave point source, the coupled diffusion equations can be presented as follows: where , is the domain under consideration. The subscripts and denote excitation light and emission light, respectively. is the diffusion coefficient with as the scattering coefficient, is the anisotropy parameter, and is the absorption coefficient. denotes the photon density. The fluorescent yield is the unknown parameter to be reconstructed, which is denoted as hereafter. Using a finite element method to solve Eq. (1), we obtained the following matrix forms:19–21 where is the system matrix. is obtained by discretizing the unknown fluorescent yield distribution. For total excitation point sources, we have the final weighted matrixThis is a relationship between the measurements and the desired unknown fluorophore distribution . It should be noted that the determination of a unique solution to Eq. (4) suffers from being ill-posed. A proper regularization method is necessary for finding a unique and stable solution. In general, Eq. (4) can be solved by the following minimization problem with the regularization term: where is the regularization parameter and is the -norm of . When , Eq. (5) becomes the -norm regularization problem, which is commonly called Tikhonov regularization. When , it is the -norm regularization problem. The -norm regularization problem can be solved efficiently using standard minimization methods, like the Newton method and the conjugate gradient method.22 In the past years, -norm regularization method has received considerable attention due to the development of a compressive sensing theory, and it can be solved by primal-dual interior-point, orthogonal matching pursuit, iterative shrinkage-thresholding, and so on.23–352.2.Imaging ModelsTwo imaging models of FMT were studied in this paper. The first imaging model is the reconstruction of fluorophore with small size, denoted as model A, which is to mimic the early detection of a tumor. Early detection of a tumor greatly increases the chances for successful treatment, and it represents one of the most promising approaches to reducing the growing tumor burden.26 The localization of the early tumor is absolutely crucial to diagnosis and treatment. Considerable attention has been paid on the early detection of various diseases.27–31 In the early stage of tumor, the biological mechanisms of interest are active in a very small region, which is “sparse” compared to the reconstruction region, as shown in Fig. 1(a). The second model is reconstructing the distribution of fluorescent dye in organs at a specific time point, denoted as model B, which is a part and parcel of the drug PK.14 It is helpful in the development of drug discovery and studying disease treatment. The reconstruction of concentration of fluorescent dye distributed in organs usually occurs with a large volume, as shown in Fig. 1(b), and does not concentrate in a localized region as model A. The different characteristics of these two models motivate our investigation of a suitable reconstruction algorithm. In the past years, Tikhonov regularization and sparsity regularization were developed for reconstruction based on the model of the early detection of tumor.10,12 To the best of our knowledge, they have not been performed on model B. This further motivates us to make a comparative study of -norm regularization and -norm regularization methods based on the two imaging models. 2.3.Imaging SystemWe developed our prototype FMT/micro-CT dual-modality imaging system for acquiring experimental data and high-resolution anatomical information. This hybrid system, as shown in Fig. 2, avoids the issue of anatomical transmogrification of coregistration between two separate FMT and micro-CT scanners. A continuous wave laser source (CrystaLaser, Reno, Nevada, and Model No. CL671-050-O) with a power of 3 mw provides an illumination at 670 nm. The laser is a stabilized compact red laser and the spot diameter of the laser beam is approximately 0.85 mm. The target is placed on a rotational stage, which is controlled by the computer. A highly sensitive charge-coupled device (CCD) camera (Princeton Instruments PIXIS 2048B, Roper Scientific, Trenton, NJ), which is cooled to , is employed to collect fluorescence signals. A Nikon Micro-NIKKOR 55 mm manual focus lens is mounted on the CCD camera. A custom-made 35 nm band-pass filter (HZXD, Beijing, China) centered at 720 nm is used to allow light transmission at the emission wavelength of 710 nm. The axial direction of camera lens is vertical to the x-ray central projection direction. The micro-CT imaging consists of an x-ray tube (Oxford Instruments series 5000 Apogee X-ray tube, X-ray Technology Inc., CA) with a focal spot size of 35 μm and a high-resolution flat panel x-ray detector (Hamamatsu C7921CA-02, Hamamatsu city, Japan) with a photo diode array with a 50 μm pixel pitch. It can provide three-dimensional (3-D) anatomic information.32 When rotating the stage to different angles, multiple view measurements can be obtained. The FMT experiments were performed in a totally dark environment. An anesthesia machine (Matrix Medical Inc., Minnesota) is employed to keep the mouse sedated during the experiment. 3.Experiments and ResultsIn this section, -norm regularization and -norm regularization methods were performed on the two aforementioned imaging models of FMT, respectively. Numerical and in vivo experiments have been designed for testing these two methods. Here, -norm regularization is solved by incomplete variables truncated conjugate gradient method, which has been demonstrated as an effective sparse regularization method in bioluminescence tomography.33 The -norm regularization problem is solved by conjugate gradient method.22 For each experiment, the algorithm was tested using a range of regularization parameters, and the best reconstruction was selected based on visual assessment. All of the reconstructions were conducted on a personal computer with a 2.66 GHz Intel(R) Xeon(R) CPU E5430 and 8.00 GB RAM. For all the numerical experiments, the fluorescent targets were excited by nine point sources at different positions in sequence, as shown in Fig. 3. For each point source, the surface data on the opposite side with a 120-deg field of view were measurable. The fluorescent yield was set to be . 3.1.Model A: Small Fluorophore Reconstruction3.1.1.Numerical simulationIn this section, a 3-D digital mouse model was considered and the cryosection data were utilized to provide anatomical information.34 We supposed that there was only one small fluorophore to be detected and only the torso section of the mouse was selected as the region to be investigated, as shown in Fig. 1(a). The small sphere was to imitate the location of fluorophore. A coarse mesh (3905 nodes and 20,380 tetrahedral-elements) was used for the inverse problem. Three different locations of fluorophore were considered, as shown in Fig. 4(a), 4(e), and 4(i). The fluorophore was positioned 10, 8, and 6 mm off the bound of the left for case 1, case 2, and case 3, respectively. The optical properties for different organs are listed in Table 1.35,36 The second and third columns are the reconstructed results at plane by -norm and -norm regularization methods, respectively. We defined the location error (LE) to analyze the results quantitatively. where is the actual center of the fluorophore and is the reconstructed coordinate of node with maximum reconstructed value for that fluorophore. The relative error (RE) of fluorescent yield is also considered to analyze the results by the following formula: where is the reconstructed fluorescent yield and is the actual one. Table 2 gives the comparative results of these two regularization methods according to LE, RE of fluorescent yield, and the full width at half maximum (FWHM). For case 2, the recovered results at coronal and sagittal views are shown in Fig. 5. It is clear that the solution of -norm regularization is concentrated on a localized region, as depicted in Figs. 4(b), 4(f), 4(j), 5(a), and 5(d). However, the solution of -norm regularization has a large area than that of -norm regularization, as shown in Figs. 4(c), 4(g), 4(k), 5(b), and 5(e). In Table 2, the LEs of -norm regularization under the three cases are less than 1 mm, which are smaller than that of -norm regularization. In addition, the fluorescent yields of -norm regularization for three cases are 0.050, 0.055, and , respectively, with all REs less than 20%. In contrast, -norm provides results with larger REs of the fluorescent yield. The FWHMs by -norm are less than 2.5 mm, while the ones by -norm are larger than 3 mm, which are greater than the actual value of 2 mm. It means -norm provides a solution confined in a localized region. The last column of Table 2 displays the reconstruction time of -norm and -norm regularization methods. It is obvious that -norm regularization consumed less time than -norm regularization as it was solved by conjugate gradient method. It should be noted that the recovered fluorescent yield values are small than the actual ones. This may be caused by the absence of a priori knowledge. For quantitative fluorescence tomography, Lin et al. showed that the true fluorophore concentration could be recovered when both functional optical background and structural a priori information, including the location of the fluorophore, are utilized to guide the FMT reconstruction algorithm.37 In practical applications, the location of the fluorescent target is usually unknown. Our experiments were conducted on this assumption.Table 1Optical parameters of the mouse organs (units of μa and μs′: mm−1).
Table 2Quantitative results for the three cases by two regularization methods.
The influence of the optical parameters for the reconstruction has been investigated in this paper. We studied the robustness of these two regularization methods against inaccurate optical parameters by adding different perturbation levels (, , ) to absorption and reduced scattering coefficients at both excitation and emission wavelengths. The perturbation is added by the following formula: where is the optical parameters in Table 1 or Table 3, namely , , , or, is the perturbation level parameter, and it is set to 5%, 10%, and 15% in this paper. is a random number generated by a MATLAB function rand with . We carried out 100 independent reconstructions for each perturbation level. The reconstructed results in Table 4 are the average of 100 independent reconstructions. It is obvious that the LE becomes large with the increase of the perturbation levels of optical parameters for both the regularization methods, while the recovered fluorescent yield almost has no change. We have also found that the standard deviations of LE and fluorescent yield by both methods increase with the raise of the perturbation levels. It indicates that the reconstruction results will become more unstable with the increase of perturbation levels.Table 3Optical parameters of the mouse organs at 670 and 710 nm (units of μa and μs′: mm−1).
Table 4Quantitative analysis of the reconstruction results for simulation experiments of model A with different perturbation levels of optical parameters.
3.1.2.In vivo implanted fluorophore experimentAll animal studies were performed in accordance with the Fourth Military Medical University Guide for the Care and Use of Laboratory Animals formulated by the National Society for Medical Research. An in vivo implantation experiment was performed on an adult BALB/C. A glass tube with 0.6 mm radius and 2.8 mm height, which contains Cy5.5 solution (with the extinction coefficient of about and quantum efficiency of 0.23 at the peak excitation wavelength of 671 nm),38 was implanted into the abdomen of the mouse that was anesthetized beforehand. The glass tube, with Cy5.5 (Fanbo, Beijing, China) solution at a concentration of 4000 nM, is the small fluorophore to be recovered. The actual fluorescent yield of Cy5.5 is according to published literature.39 The experimental data were acquired by the dual-modality FMT/micro-CT system (see Fig. 2). After the optical image acquisition, the anesthetized mouse was scanned by micro-CT. The CT slices were segmented into major anatomical components, including heart, lungs, liver, kidneys, and muscle. The optical properties at both the excitation and emission wavelengths for the main organs were calculated based on the literature as listed in Table 3.35 A landmarks-based rigid-body registration method was adopted to make the CT data registered with the fluorescence imaging.33 After the registration, the absolute irradiance distribution in two-dimensional (2-D) fluorescence images was mapped on the 3-D mouse surface.40 The segmented mouse torso was discretized into 18,504 tetrahedral-elements and 3823 nodes for reconstruction. Figure 6 shows the results by -norm and -norm regularization methods, which are compared with the corresponding CT slices. The recovered center deviation is 1.73 mm using -norm regularization, while using -norm regularization method, the recovered center deviation is 2.05 mm. Figure 7 provides the fluorescent yield profiles of in vivo experiment and Table 5 provides the quantitative results. From Figs. 6 and 7, it is clear that the solution of -norm regularization is still confined in a small region with clean background, while the one of -norm regularization has a large area, which is consistent with the numerical experiment results. Furthermore, -norm regularization generated spurious small elements in Fig. 6(b), which also appear in Fig. 4(c) and 4(k). The green solid lines in the longitudinal of Fig. 6(a) are closer to the red dashed lines than that of Fig. 6(b), which indicates that -norm regularization provides a better location of fluorophore than -norm regularization method does. Table 5Comparative results for in vivo experiment of model A.
The optical parameters for the main organs are calculated according to literature,35 not experimentally measured. However, the tissue optical parameters of the mouse in our experiments may be significantly different from the literature data. So we have also studied the influence by changing the optical parameters for reconstruction. Table 6 provides the quantitative analysis for each perturbation level. From Table 6, we have found that the LE by two methods increases with the raise of the perturbation levels of optical parameters, while the recovered fluorescent yield increases slightly. It is clear that the standard deviation increases with the raise of the perturbation levels, which indicates that the reconstruction results are very unstable when the perturbation level of optical parameters is large. This conclusion is consistent with the numerical experiment. Table 6Reconstruction results with optical properties perturbation for in vivo experiment of model A.
3.2.Model B: Reconstruction of Fluorescent Dye Distribution in Organs3.2.1.Numerical simulationIn the numerical experiment, the mouse chest region is considered instead of the torso section, as shown in Fig. 1(b). The heart was considered as an entirety, and the atriums, ventricles, and vessels were not considered. Our aim was to recover the fluorophore distribution in the heart. The fluorescence yield was set to be . The optical parameters for major organs are shown in Table 1. A coarse mesh (2292 nodes and 11,737 tetrahedral-elements) was used for the inverse problem. Figure 8 presents the recovered results by -norm and -norm regularization methods. For clarification, we display the results at slices , 6.5, and 7.5 mm, respectively. It seems that the solution of -norm regularization is confined in small regions. On the contrary, the solution of -norm is smooth and has a large area. This indicates that -norm regularization still provides a solution with a localized region, while -norm regularization gives a reasonable solution. This is because -norm regularization tends to find sparse solutions, while -norm regularization tends to penalize large elements. The profiles along the yellow dotted lines in Fig. 8 are shown in Fig. 9. Figure 9(a) is a profile of horizontal lines and Fig. 9(b) is one of vertical lines. Table 7 provides the FWHM for simulation experiments of model B. From Fig. 9 and Table 7, the FWHM of either horizontal or vertical lines, it is clear that the FWHM of -norm is much closer to the actual FWHM. It means the solution provided by -norm has large volume and is more approximated to the actual one. Table 7Full width at half maximum of simulation experiment for model B.
We have also investigated the influence by changing optical parameters on the recovered results for model B. Figure 10 shows the recovered results for perturbation levels at , , and . The third column is the profiles along the yellow dotted lines in the first and second columns. The blue solid lines are the real distribution along the yellow dotted lines. The red dashed lines and green solid lines represent the reconstructed distribution by -norm and -norm regularization. There is a slight variation in the distribution, as shown in Fig. 10(c), 10(f), and 10(i). In order to show the reconstruction results clearly, we have introduced the average fluorescent yield, which is the ratio of total fluorescent yield to the total volume of the organ. The average fluorescent yields of the heart by two methods for each perturbation level are shown in the third column of Table 8. There is little variation in the average fluorescent yield by the two methods, but the standard deviations of average fluorescent yield increase with the increase of perturbation levels. Table 8Average fluorescent yield in the organ for simulation experiment and in vivo experiment of model B.
3.2.2.Reconstruction of Cy5.5 solution distribution in the bladder of mouseThe Cy5.5 solution of 0.12 ml, with a concentration of 4000 nM, was injected to an adult BALB/C mouse via tail vein. We considered the lower body of the mouse as the reconstruction region, and the problem was to recover the distribution of Cy5.5 in the bladder. After 6 h, we acquired the data using our prototype FMT/micro-CT dual-modality imaging system for reconstruction. The optical properties of main organs are listed in Table 3. The segmented mouse torso was discretized into 14,500 tetrahedral-elements and 3085 nodes for reconstruction. The actual concentration of Cy5.5 in the bladder is unknown, and the recovered results are normalized by the maximum value of -norm for comparison. Figure 11 shows the results, which are compared to the corresponding CT slices. It is clear that the solution of -norm regularization concentrates on some small regions, while the solution of -norm regularization has a large area and is smooth, which are consistent with the results in the numerical experiments. Figure 12 is profiles along the yellow dotted lines in Fig. 11, and Table 9 is the FWHM of in vivo experiment. Table 9Full width at half maximum of in vivo experiment for model B.
Figure 13 shows the reconstructed results under the optical parameters perturbation at 0, , , and . All of the results are normalized by the maximum value recovered by -norm regularization with no optical properties perturbation. The third column is the profiles along the yellow dotted lines in the first and second columns. The blue solid lines are the real distributions along the yellow dotted lines. The red dash-dot lines and green solid lines represent the reconstructed distribution by -norm and -norm regularization. The first row is results with no optical parameters perturbation, and the second row is for perturbation at , the third row is for , and the fourth row is for . The average fluorescent yield of the bladder is shown in the fourth column of Table 8. We have found that there is almost no change in the average value, but the standard deviations of average fluorescent yield by two methods increase with the raise of perturbation levels. 4.Discussion and ConclusionIn this paper, we make a comprehensive comparison between -norm and -norm regularization methods based on two imaging models of FMT. For the early detection of tumor model, numerical experiments and in vivo experiments were conducted to test the two methods. The experiment results indicate that -norm regularization produced the solution confined in a localized region with clean background, while -norm regularization created a smooth solution with a large volume and generated spurious small elements. It should be noted that the recovered LEs of in vivo experiment were larger than that of the numerical experiments. This may be caused by many factors: (1) the photon propagation was described by diffusion approximation, not the radiative transfer equation,41 (2) the optical properties for major organs were using the published values, not using DOT,7,9 (3) experimental operations may also affect the reconstruction results. In the model of reconstructing the distribution of fluorescent dye in organs, -norm regularization shrunk the support of the fluorescent and still provided the solution with small volume, while -norm regularization produced a smooth solution with a large volume. The FWHM of fluorescent yield profiles by -norm is more close to the actual one than that of -norm. We infer that for these two imaging models of FMT, the -norm regularization is a more suitable choice for the reconstruction of small fluorophore as it fully takes advantage of the sparse characteristic of the fluorophore; -norm regularization is more accurate for reconstructing the distribution of fluorescent dye in organs since it tends to penalize large elements. It should be noted that the size and shape of fluorescent target cannot be accurately recovered by either -norm or -norm regularization. It is desirable to design more appropriate regularization methods for this problem. Gao et al. have developed total variation and data fidelity for bioluminescence tomography based on radiative transfer equation. 42 This algorithm showed good performance for 2-D numerical simulation experiments, especially on preserving the boundary of the object. The extensive clinical applications could encounter difficulties for the complex 3-D reconstruction. This is our future research. In conclusion, we have made a detailed comparison between -norm and -norm regularization methods based on two imaging models of FMT. Some interesting conclusions were obtained in this paper, which would provide useful information for the researcher in selecting and designing algorithms for FMT reconstruction. AcknowledgmentsThis work was supported by the Program of the National Basic Research and Development Program of China (973) under Grant No. 2011CB707702, the National Natural Science Foundation of China under Grant Nos. 81090270, 81090272, 81227901, 81000632, 81101084, the National Key Technology Support Program under Grant No. 2012BAI23B06, and the Fundamental Research Funds for the Central Universities. ReferencesV. Ntziachristoset al.,
“Looking and listening to light: the evolution of whole body photonic imaging,”
Nat. Biotechnol., 23
(3), 313
–320
(2005). http://dx.doi.org/10.1038/nbt1074 NABIF9 1087-0156 Google Scholar
E. E. GravesR. WeisslederV. Ntziachristos,
“Fluorescence molecular imaging of small animal tumor models,”
Curr. Mol. Med., 4
(4), 419
–430
(2004). http://dx.doi.org/10.2174/1566524043360555 CMMUBP 1566-5240 Google Scholar
A. Corluet al.,
“Three-dimensional in vivo fluorescence diffuse optical tomography of breast cancer in humans,”
Opt. Express, 15
(11), 6696
–6716
(2007). http://dx.doi.org/10.1364/OE.15.006696 OPEXFF 1094-4087 Google Scholar
M. J. Niedreet al.,
“Early photon tomography allows fluorescence detection of lung carcinomas and disease progression in mice in vivo,”
Proc. Nat. Acad. Sci. U.S.A., 105
(49), 19126
–19131
(2008). http://dx.doi.org/10.1073/pnas.0804798105 PMASAX 0096-9206 Google Scholar
J. K. Willmannet al.,
“Molecular imaging in drug development,”
Nat. Rev. Drug Discov., 7
(7), 591
–607
(2008). http://dx.doi.org/10.1038/nrd2290 NRDDAG 1474-1776 Google Scholar
R. B. Schulzet al.,
“Hybrid system for simultaneous fluorescence and x-ray computed tomography,”
IEEE Trans. Med. Imag., 29
(2), 465
–473
(2010). http://dx.doi.org/10.1109/TMI.2009.2035310 ITMID4 0278-0062 Google Scholar
Y. Linet al.,
“Quantitative fluorescence tomography using a combined tri-modality FT/DOT/XCT system,”
Opt. Express, 18
(8), 7835
–7850
(2010). http://dx.doi.org/10.1364/OE.18.007835 OPEXFF 1094-4087 Google Scholar
F. Stukeret al.,
“Hybrid small animal imaging system combining magnetic resonance imaging with fluorescence tomography using single photon avalanche diode detectors,”
IEEE Trans. Med. Imag., 30
(6), 1265
–1273
(2011). http://dx.doi.org/10.1109/TMI.2011.2112669 ITMID4 0278-0062 Google Scholar
Y. TanH. Jiang,
“DOT guided fluorescence molecular tomography of arbitrarily shaped objects,”
Med. Phys., 35
(12), 5703
–5707
(2008). http://dx.doi.org/10.1118/1.3020594 MPHYA6 0094-2405 Google Scholar
X. ZhangC. T. BadeaG. A. Johnson,
“Three-dimensional reconstruction in free-space whole-body fluorescence tomography of mice using optically reconstructed surface and atlas anatomy,”
J. Biomed. Opt., 14
(6), 064010
(2009). http://dx.doi.org/10.1117/1.3258836 JBOPFO 1083-3668 Google Scholar
A. D. Zacharopouloset al.,
“A matrix-free algorithm for multiple wavelength fluorescence tomography,”
Opt. Express, 17
(5), 3025
–3035
(2009). http://dx.doi.org/10.1364/OE.17.003042 OPEXFF 1094-4087 Google Scholar
D. Hanet al.,
“Sparsity promoting tomographic fluorescence imaging with simplified spherical harmonics approximation,”
IEEE Trans. Biomed. Eng., 57
(10), 2564
–2567
(2010). http://dx.doi.org/10.1109/TBME.2010.2053538 IEBEAX 0018-9294 Google Scholar
P. Mohajeraniet al.,
“Optimal sparse solution for fluorescent diffuse optical tomography: theory and phantom experimental results,”
Appl. Opt., 46
(10), 1679
–1685
(2007). http://dx.doi.org/10.1364/AO.46.001679 APOPAI 0003-6935 Google Scholar
B. AlacamB. Yazici,
“Direct reconstruction of pharmacokinetic-rate images of optical fluorophores from NIR measurements,”
IEEE Trans. Med. Imag., 28
(9), 1337
–1353
(2009). http://dx.doi.org/10.1109/TMI.2009.2015294 ITMID4 0278-0062 Google Scholar
V. V. Tuchin,
“Light scattering studies of tissues,”
Phys. Usp., 40
(5), 495
–515
(1997). http://dx.doi.org/10.1070/PU1997v040n05ABEH000236 PHUSEY 1063-7869 Google Scholar
S. R. Arridge,
“Optical tomography in medical imaging,”
Inverse Probl., 15
(2), R41
–R93
(1999). http://dx.doi.org/10.1088/0266-5611/15/2/022 INPEEY 0266-5611 Google Scholar
A. KienleM. S. Patterson,
“Improved solutions of the steady-state and the time-resolved diffusion equations for reflectance from a semi-infinite turbid medium,”
J. Opt. Soc. Am. A., 14
(1), 246
–254
(1997). http://dx.doi.org/10.1364/JOSAA.14.000246 JOAOD6 1084-7529 Google Scholar
M. Schweigeret al.,
“The finite element method for the propagation of light in scattering media: boundary and source conditions,”
Med. Phys., 22
(11), 1779
–1792
(1995). http://dx.doi.org/10.1118/1.597634 MPHYA6 0094-2405 Google Scholar
S. R. Arridgeet al.,
“A finite element approach for modeling photon transport in tissue,”
Med. Phys., 20
(2), 299
–309
(1993). http://dx.doi.org/10.1118/1.597069 MPHYA6 0094-2405 Google Scholar
M. SchweigerS. R. ArridgeD. T. Delpy,
“Application of the finite-element method for the forward and inverse models in optical tomography,”
J. Math. Imag. Vis., 3
(3), 263
–283
(1993). http://dx.doi.org/10.1007/BF01248356 JIMVEC 0924-9907 Google Scholar
D. Wanget al.,
“A novel finite-element-based algorithm for fluorescence molecular tomography of heterogeneous media,”
IEEE Trans. Inf. Technol. Biomed., 13
(5), 766
–773
(2009). http://dx.doi.org/10.1109/TITB.2009.2015144 ITIBFX 1089-7771 Google Scholar
W. SunY. Yuan, Optimization Theory and Method, Springer, New York
(2006). Google Scholar
M. KojimaN. MegiddoS. Mizuno,
“Theoretical convergence of large-step primal-dual interior point algorithms for linear programming,”
Math. Program., 59
(1–3), 1
–21
(1993). http://dx.doi.org/10.1007/BF01581234 MHPGA4 1436-4646 Google Scholar
G. DavisS. MallatM. Avellaneda,
“Adaptive greedy approximations,”
Constr. Approx., 13
(1), 57
–98
(1997). http://dx.doi.org/10.1007/BF02678430 0176-4276 Google Scholar
I. DaubechiesM. DefriseC. Mol,
“An iterative thresholding algorithm for linear inverse problems with a sparsity constraint,”
Commun. Pure Appl. Math., 57
(11), 1413
–1457
(2004). http://dx.doi.org/10.1002/(ISSN)1097-0312 CPMAMV 0010-3640 Google Scholar
R. Etzioniet al.,
“The case for early detection,”
Nat. Rev. Cancer, 3
(4), 243
–252
(2003). http://dx.doi.org/10.1038/nrc1041 NRCAC4 1474-175X Google Scholar
R. Ellmanet al.,
“Breast self-examination programmes in the trial of early detection of breast cancer: ten years findings,”
Br. J. Cancer, 68
(1), 208
–212
(1993). http://dx.doi.org/10.1038/bjc.1993.315 BJCAAI 0007-0920 Google Scholar
P. Cairns,
“Gene methylation and early detection of genitourinary cancer: the road ahead,”
Nat. Rev. Cancer, 7
(7), 531
–543
(2007). http://dx.doi.org/10.1038/nrc2170 NRCAC4 1474-175X Google Scholar
A. McWilliamset al.,
“Innovative molecular and imaging approaches for the detection of lung cancer and its precursor lesions,”
Oncogene, 21
(45), 6949
–6959
(2002). http://dx.doi.org/10.1038/sj.onc.1205831 ONCNES 0950-9232 Google Scholar
A. LevinP. E. Stevens,
“Early detection of CKD: the benefits limitations and effects on prognosis,”
Nat. Rev. Nephrol., 7
(8), 446
–457
(2011). http://dx.doi.org/10.1038/nrneph.2011.86 NRNABO 1759-5061 Google Scholar
A. Herbstet al.,
“Methylation of NEUROG1 in serum is a sensitive marker of the detection of early colorectal cancer,”
Am. J. Gastroenterol., 106
(6), 1110
–1118
(2011). http://dx.doi.org/10.1038/ajg.2011.6 AJGAAR 0002-9270 Google Scholar
G. Yanet al.,
“Fast cone-beam CT image reconstruction using GPU hardware,”
J. X-Ray Sci. Technol., 16
(4), 225
–234
(2008). JXSTE5 0895-3996 Google Scholar
X. Heet al.,
“Sparse reconstruction for quantitative bioluminescence tomography based on the incomplete variables truncated conjugate gradient method,”
Opt. Express, 18
(24), 24825
–24841
(2010). http://dx.doi.org/10.1364/OE.18.024825 OPEXFF 1094-4087 Google Scholar
B. Dogdaset al.,
“Digimouse: a 3D whole body mouse atlas from CT and cryosection data,”
Phys. Med. Biol., 52
(3), 577
–587
(2007). http://dx.doi.org/10.1088/0031-9155/52/3/003 PHMBA7 0031-9155 Google Scholar
G. AlexandrakisF. R. RannouA. F. Chatziioannou,
“Tomographic bioluminescence imaging by use of a combined optical-PET (OPET) system: a computer simulation feasibility study,”
Phys. Med. Biol., 50
(17), 4225
–4241
(2005). http://dx.doi.org/10.1088/0031-9155/50/17/021 PHMBA7 0031-9155 Google Scholar
A. CongG. Wang,
“A finite-element-based reconstruction method for 3D fluorescence tomography,”
Opt. Express, 13
(24), 9847
–9857
(2005). http://dx.doi.org/10.1364/OPEX.13.009847 OPEXFF 1094-4087 Google Scholar
Y. Linet al.,
“Quantitative fluorescence tomography using a trimodality system: in vivo validation,”
J. Biomed. Opt., 15
(4), 040503
(2010). http://dx.doi.org/10.1117/1.3467495 JBOPFO 1083-3668 Google Scholar
F. Gaoet al.,
“A linear, featured-data scheme for image reconstruction in time-domain fluorescence molecular tomography,”
Opt. Express, 14
(16), 7109
–7124
(2006). http://dx.doi.org/10.1364/OE.14.007109 OPEXFF 1094-4087 Google Scholar
M. A. NaserM. S. Patterson,
“Improved bioluminescence and fluorescence reconstruction algorithms using diffuse optical tomography, normalized data, and optimized selection of the permissible source region,”
Biomed. Opt. Express, 2
(1), 169
–184
(2011). http://dx.doi.org/10.1364/BOE.2.000169 BOEICL 2156-7085 Google Scholar
X. Chenet al.,
“3D reconstruction of light flux distribution on arbitrary surfaces from 2D multi-photographic images,”
Opt. Express, 18
(19), 19876
–19893
(2010). http://dx.doi.org/10.1364/OE.18.019876 OPEXFF 1094-4087 Google Scholar
Y. Luet al.,
“Experimental bioluminescence tomography with fully parallel radiative-transfer-based reconstruction framework,”
Opt. Express, 17
(19), 16681
–16695
(2009). http://dx.doi.org/10.1364/OE.17.016681 OPEXFF 1094-4087 Google Scholar
H. GaoH. Zhao,
“Multilevel bioluminescence tomography based on radiative transfer equation. Part 2: total variation and l1 data fidelity,”
Opt. Express, 18
(3), 2894
–2912
(2010). http://dx.doi.org/10.1364/OE.18.002894 OPEXFF 1094-4087 Google Scholar
|