|
1.IntroductionOptical imaging in turbid media saw rapid development over the last few years. Applications in cancer detection,1, 2 the monitoring of brain activation,3, 4 and small animal fluorescence imaging5, 6, 7 were demonstrated. These developments also led to new commercial offerings. With this emergence of diffuse optical tomography (DOT), systems with an increasing number of sources and detectors were built to enhance image quality by exploiting a larger set of measures. In particular, it has been recently recognized that with dense spatial measurements, it is possible to image complex structures with high-resolution.8 In this work, we focus on noncontact transmission geometry with planar source and detector planes. Such geometry is commonly employed in optical mammography9 and small animal imaging associated with coupling liquid to mitigate high dynamical range in the detection signals.10 Previous diffuse optical tomography systems were implemented in this geometry by either positioning a series of punctual light sources and detection points at different positions on the sample surface or scanning the sources. This scan was done mechanically9 or optically with galvanometer mirrors8 providing high-density measurements. Detection in these configurations was mostly performed with either single detectors or a sensitive camera. While these acquisition strategies are experimentally straightforward to implement, they suffer from a major drawback. To acquire dense illumination and detection measurements, every light position has to be spatially scanned or time multiplexed, thereby limiting the frame rate that can be achieved. This is especially crippling in functional imaging applications where spectral information at multiple wavelengths is required.11 Furthermore, the large amount of data acquired leads to massive inverse problems and necessitates extensive computation resources. The incentive for this work is found in two simple observations: (1) even if high-density measurements were used to reconstruct complex structures in Ref. 8, only a smaller number of Fourier modes were employed for the reconstructions; (2) by using digital micromirror devices (DMDs) not only for illumination but also for detection, it is possible to generate specifically those relevant measures for reconstruction. With this approach, the number of measurements required is drastically reduced, leading to significant data compression and potential frame rate increase. The ability to illuminate specific spatial Fourier modes was exploited in recent work under the denomination of modulated imaging.12 By exploiting new technologies—more precisely, DMD arrays—it is possible to shine arbitrary patterns of light on tissues and use a model of light propagation to infer information about the media. For example, baseline tissue optical properties were estimated by illuminating distinct spatial Fourier modes and using diffusion theory.13 A recent extension included the study of time-domain spatial mode propagation14 and two-layer optical parameter estimation.12 All these initial works were not tomographic but “sectioning” methods, as the tomographic information was encoded in the frequency and not retrieved through an inverse problem. Only recently has tomography been explored.15 Furthermore, “pure tones” or spatial frequencies were used for illumination, diminishing significantly the acquisition rate. The technical reason for the reduced acquisition rates is that mirrors in DMDs can only be switched “on” or “off” and gray levels are produced by applying a pulse-width modulation (PWM) of the mirrors to create the intensity effect. While it is estimated that with current micromirror devices, the whole array can be controlled at rates exceeding , the frame rate in PWM mode is decreased to around . In these previous applications, DMD technology was used only for illumination in combination with a camera for detection. The wide-field acquisition capabilities of the camera allowed the computation of all detection Fourier modes with a single frame. However, acquisition frame rates are limited by the integration time of the camera—typically, —due to the low-light nature of diffuse optical applications. Conversely, it is possible to acquire sequentially all the detection Fourier modes directly with a detection channel comprised of a DMD and a single detector such as a photomultiplier tube (PMT) or an avalanche photodiode (APD). Thanks to the increased level of light, measures can be acquired at increased frame rates to obtain a complete set of measures for high-resolution DOT at unprecedented speeds. In this work, a dual DMD acquisition technique to perform diffuse optical imaging is experimentally demonstrated for the first time in phantoms. Previous work on diffusion theory is used to guide an approximation of the Fourier measures by simple “on-off” mirror patterns, thereby increasing the DMD switching speed (from ). We show that by selecting a subset of measurements, it is possible to get sufficient data for high-fidelity three-dimensional (3-D) image reconstruction at real-time frame rates. 2.ModelIn the diffusion approximation to the radiative transport equation, the propagation of light in tissues (valid if ) is modeled by a diffusion equation for the fluence :16, 17, 18 where is the diffusion coefficient that will be assumed constant in the following, the absorption coefficient, the speed of light in the medium, and a source located on the boundary. In the experiments described in the following, all measures were taken in the continuous wave (CW) mode so that the source is considered to have no time dependence. To completely define the problem, boundary conditions are specified bywhere is an outward normal to the boundary, and is an extrapolation distance.19 In the slab geometry, the source and detectors lie in the planes, and , with describing two-dimensional (2-D) vectors on the surface on which measures are made. Define the Green function:Then by using the Rytov approximation and denoting , the following equation relates the measures to variations in absorption:16 where and are, respectively, the measured intensities of the perturbed and unperturbed sample. For example, with time-varying changes, the baseline could be defined to be the data taken at the beginning of the temporal changes. This expression can also be extended beyond the diffusion approximation, e.g., for small spatial volumes where the diffusion approximation does not hold, using other techniques to compute the Green function such as Monte Carlo simulations or numerical solutions of the radiative transport equation. Discretizing the integral in Eq. 4, the approximate linear forward model relating measures to changes in absorption can be inverted by a regularized pseudo-inverse or by other techniques.By assuming translational symmetry in the directions, define the Fourier transform in the spatial coordinates that are “in-plane,” i.e., , as Equation 4 can be written asTaking into account the boundary conditions,14 we can writewhere and . The Fourier transform provides a block-diagonalized form and can be numerically estimated for each . In Ref. 8, this formulation was used to image a complex phantom by scanning a source with a galvanometer mirror and taking a large set of measurements with a sensitive camera. While a dense set of measures were used , many fewer combinations of these measures were actually needed in the reconstruction formula 7 since both and were discretized. Thus, if one could generate directly the functions for a series of ’s, only the data used in the inversion formula would be measured.As mentioned in Sec. 1, a few issues arise with this strategy when focusing on this equation. First, the pure Fourier measures are slower to generate with DMDs due to the necessity of generating gray levels. Until now in modulated imaging, sinusoidal measures were used, but these carry a temporal cost due to the PWM of some mirrors and integration time of the camera.15 Second, the optimal spatial patterns described by Eq. 7 rely on translational invariance, which in some cases will not be valid. Last, in situations where the diffusion theory does not hold, this approach is not adequate since the forward model used is inaccurate. However, these issues can be mitigated by the following observations: While the diffusion theory and spatial invariance are not always true, it is possible to build a forward model that takes into account those two issues. For example, a Monte Carlo simulation of the fluence induced by a pattern is simple to compute by convolution, taking into account both the issue of spatial invariance (by modeling an arbitrary medium) and the diffusion theory approximation. While the analytical inversion formula requires pure Fourier modes that are longer to measure, the DMDs can be used to get measures that are approximates of the Fourier modes. Formula 7 is no longer available in that case, but it provides guidance as to which measures are important in the inversion. The decrease of the number of measures (by using approximate Fourier modes) can then be exploited by using a pseudo-inverse (that can be precomputed) to provide a real-time inversion. These issues are investigated in the following, both in simulations and with experimental work on phantoms. We first use Monte Carlo simulations to generate measures by convolution in a simple homogeneous volume and use perturbation theory to compare different illumination/detection schemes for tomography. Once identified, we study the effect of noise on the reconstructions in a given scheme. Last, we provide experimental evidence of the feasibility of the approach proposed in this paper by using liquid phantoms. 3.SimulationsThe availability of a large number of illumination and detection patterns ( combinations given that each array has mirrors) makes it impossible to explore the entire measurement space. Since our goal was to provide a proof of principle for this dual DMD technique, we considered and compared two types of patterns. Many others can be explored, but doing so requires an extensive study. To model light propagation, Monte Carlo simulations were performed for a single source in a homogeneous medium.20 Assuming perpendicular translation invariance, the resulting fluence was then convolved with each illumination pattern to create the fluence distribution for each case. A sensitivity matrix was computed by doing the same for the detector configuration. Using this forward model, a numerical phantom was created and simulated measures generated. These data were used for the ensuing inversion, which was done with a regularized pseudo-inverse. Forward and inverse problems were performed on distinct grids to minimize the “inverse crime.” Figure 1 shows the two types of patterns used in simulations. In each case, the same pattern sets were used for both illumination and detection ( patterns). The total number was chosen to achieve a frame-rate of , which was the initial target but can be chosen arbitrarily. In simulations, a numerical phantom of rectangular geometry of was used. Optical parameters were set to and to mimic the experimental configuration described in the following. Two numerical inclusions were added to the phantom consisting of two bars, each of width and situated at . The contrast of the inclusions was set to 100%, i.e., . Figure 1 compares reconstructions between the two illumination/detection patterns: checkerboard and moving bars. Taking a line profile across the axis leads to a FWHM of for both bars, but the contrast for pattern 1 was 50% lower than that of pattern 2 (moving bars). For inclusions having 100% contrasts, the reconstructed value was 40%, coherent with the observed spatial extent of the solution. Expectations were that the checkerboard pattern would perform better based on the analytic block diagonalization of Eq. 7 and a singular value analysis (Fig. 2 ). However, it was the moving bars that enabled the best reconstructions for the chosen phantom. The numerical noise introduced by the discretization and the associated truncation performed during the computation of the pseudo-inverse may explain the slightly better performance by the moving bars pattern. To evaluate sensitivity to noise, simulations were performed with 0, 5, and 10% noise added to the simulated measures with respect to the mean. In each case, the regularization parameter was optimized based on the mean-squared distance between the reconstruction and the gold-standard. Results are presented in Fig. 3 for the moving bars (pattern 2). While spatial aliasing was present in all reconstructions, degradation of the reconstruction appeared only at high noise levels (10%). These simulations were used to find the optimal regularization parameter for noise levels similar to that measured experimentally (1 to 2%; see the following). These latter values were used for the following experimental reconstructions. In all of the preceding simulations, the reconstructions were performed both on an eight-core computer ( RAM, ) and a graphical processor unit (GPU; GForce 280GTX ) system to evaluate the feasibility of real-time 3-D imaging. On both computational platforms, reconstructions based on precomputed pseudo-inverse were executed in a few milliseconds. For the inverse problem investigated herein, 1296 pattern pairs and 131,072 voxels, the multicore and the GPU platform led to and inversion time, respectively. 4.ExperimentsThe experimental setup is depicted in Fig. 4 . The illumination consisted of a single LED ( , ) collimated with an biconvex lens. The collimated beam was aligned with the first micromirror matrix (DMD, TI Discovery 3000 board), and the pattern generated by the DMD was then imaged on the surface of the object through a focal length magnifying lens. The average surface illuminated was , leading to an estimated fluence, well below ANSI standards. The detection side was symmetrical: the light coming out of the turbid medium was imaged on a second DMD with another biconvex lens. The detection patterns generated by the second mirrors then redirected the light in the direction of a collimating lens adjusted to image the width of the DMD onto the surface of a photomultiplier tube. Data from the PMT, as well as the trigger signals controlling the DMD boards, went through a homemade adaptive electronic circuit to a DAQ card (National Instruments PCI6009; transistor-transistor logic (TTL) levels differ between the DAQ and the DMD boards). For the experiments, moving bars were chosen based on the simulations results. In the following pilot study, a complete acquisition frame consisted of 1296 measurements (36 emission patterns by 36 detection patterns), as in the preceding simulations. This number is not limited, however; it is possible to include larger numbers of source detector pairs by reprogramming the DMD boards. DMDs took approximately to generate a pattern. Once the pattern was generated, 10 measurements from the voltage outputs of the PMT were taken by the acquisition board, sampling at , adding a time lapse of . Last, software averages and communication between computer and DMD boards took , for a total of approximately by pattern. Thus, a full frame of 1296 measures was acquired in leading to a frame rate of including overhead. Since all patterns were known ahead of time, a precomputed pseudo-inverse with Tikhonov regularization was used to reconstruct the images in real time. The data was sent by TCP/IP from the acquisition computer to a second reconstruction computer with little transmit time given the intrinsic data compression afforded by this illumination-detection approach. Data was reconstructed in and displayed in real-time on this second computer. The experimental acquisitions were done on a controlled phantom with optical properties set to and a (dilution of Intralipid 10% with India ink). An imaging chamber with Plexiglas windows was used to hold the liquid. Since the reconstruction algorithm required a baseline measure, a set of frames was taken on the homogeneous liquid, phantom. Following this, the inclusion was inserted in the liquid, and further frames were taken. In all cases, data were measured at the frame rate and stability of the system was first assessed. The mean noise level of the measurements was observed to be between 1 to 2% for a frame rate. Then, the system was run continuously following an initial frame, and no drift was observed while maintaining the rate over periods exceeding . Figure 5 shows the baseline and phantom frames with their noise. Using a second DMD for detection seems to provide faster acquisitions than previously obtained in the literature. In previous work using a CCD,20 was necessary to integrate light information in similar conditions as the ones presented here. Given that only 36 illumination patterns would be necessary in this case, since detection patterns would be computed in software, the estimated acquisition time for an equivalent data set is . This is approximately 7 times slower than the method presented in this study but is also dependent on the camera and optical configuration. Figure 6 shows the reconstructions for distinct inclusion positions and a comparison of these reconstructions with data simulated for the same phantom positions. In this case, the phantom consisted of a -diam graphite rod (completely absorbing). For the three positions, the FWHM of the reconstructed inclusion was estimated to be for the ones positioned near the boundaries and for the inclusion located at the center of the liquid. These results are consistent with a simulation done numerically for the same inclusion size and the associated reconstruction (similar noise levels were used in the simulation). Contrast is observed to be similar in both simulations and experimental data, but it is difficult to infer any meanings to the values since the inclusion was completely absorbing and a Born approximation was used for reconstruction. A mild aliasing phenomenon is observed and is due to the regularity of the chosen measurement patterns for emission and detection. This aliasing is expected to disappear if a set of incoherent measures are used akin to work performed in compressed sensing in recent years.21 Resolution in both simulations and experiments are similar, as are the simulated and experimental measurement frames (data not shown). 5.ConclusionIn this work, simulations and experimental results were presented to validate the feasibility of a fast tomographic imager using structured illumination and detection. Simulations demonstrated the possibility to perform reconstructions with high resolution and accurate quantification without using a large number of measures. The reconstructions’ sensitivity to noise was also evaluated, and the acquisition scheme was shown to be stable in the presence of noise equivalent to that found in experiments. The use of a small number of measures enabled us to perform a full pseudo-inverse on a desktop computer. With precomputation of the pseudo-inverse, a frame rate of , including 3-D image reconstruction, was achieved on a GPU. These results were further exploited to provide ultra-fast reconstruction display to assess our experiments in real time. While not as accurate as the reconstructions found in, Ref. 8 the reconstructions were comparable to the literature in resolution. It is expected that by increasing the number of patterns used, the resolution of the reconstructions will increase and may be able to achieve that found in Ref. 8. However, the frame rate will diminish, and further work is required to assess the best patterns to optimize the resolution to frame rate trade-off. We believe that the ability of performing real-time 3-D images is the main benefit of this system, and applications to dynamical imaging of optically thick tissue such as pharmacokinetics will be the subject of further studies.22 A more extensive characterization of the reconstructions could have been performed, but since only simple illumination-detection patterns were explored, it is not clear that it provides a quantitative evaluation of the system. The goal of this paper was to provide a proof of principle for the dual DMD approach and estimate achievable frame rates and noise levels. A separate study to identify the optimal illumination-detection strategy is needed, but it is beyond the scope of this work. With these optimal measures, the trade-off between frame rate and image quality can be evaluated more accurately. Another drawback of this work is the use of precomputed pseudo-inverses to perform image reconstructions. It is not clear that this is the optimal reconstruction method, but it does provide a means to visualize data rapidly. Besides illuminating only the relevant modes, there are significant advantages to this new approach when compared to other techniques. First, by using wide-field illumination, it is possible to increase the light intensity while preserving the ANSI limit for tissue exposition since the limit is per area. Here data were acquired with a simple LED, but using a more powerful source, one should be able to perform fluorescence tomography. Second, although higher total light intensity is used, it is spatially distributed, and at any point in the volume imaged, the light intensity is lower than in local illumination schemes. This has the advantage of decreasing bleaching effects for potential fluorescence applications while still increasing signal-to-noise ratio (SNR). Last, compared to previous reports, when using a DMD for both illumination and detection, it is possible to limit the measurements to only those necessary and recollimate the light detected to spread it in distinct spectral components, providing spectral measurements without time penalty. We hope to revisit these issues in future work. AcknowledgmentsThis work was supported by NSERC and CIHR grants to C.C. and an NSERC grant to F.L. FRSQ provided most of C.C.’s salary (Chercheur National Program). ReferencesS. Nioka and
B. Chance,
“NIR spectroscopic detection of breast cancer,”
Technol. Cancer Res. Treat., 4 497
–512
(2005). 1533-0346 Google Scholar
B. J. Tromberg,
N. Shah,
R. Lanning,
A. Cerussi,
J. Espinoza,
T. Pham,
L. Svaasand, and
J. Butler,
“Noninvasive in vivo characterization of breast tumors using photon migration spectroscopy,”
Neoplasia, 2 26
–40
(2000). https://doi.org/10.1038/sj.neo.7900082 1522-8002 Google Scholar
J. P. Culver,
B. L. Schlaggar,
H. Dehghani, and
B. W. Zeff,
“Diffuse optical tomography for mapping human brain function,”
Life Science Systems and Applications Workshop, IEEE/NLM, 1
–2
(2006) Google Scholar
M. A. Franceschini,
D. K. Joseph,
T. J. Huppert,
S. G. Diamond, and
D. A. Boas,
“Diffuse optical imaging of the whole head,”
J. Biomed. Opt., 11 054007
(2006). https://doi.org/10.1117/1.2363365 1083-3668 Google Scholar
V. Ntziachristos,
J. Ripoll,
L. V. Wang, and
R. Weissleder,
“Looking and listening to light: the evolution of whole-body photonic imaging,”
Nat. Biotechnol., 23 313
–320
(2005). https://doi.org/10.1038/nbt1074 1087-0156 Google Scholar
W. Akers,
F. Lesage,
D. Holten, and
S. Achilefu,
“In vivo resolution of multiexponential decays of multiple near-infrared molecular probes by fluorescence lifetime-gated whole-body time-resolved diffuse optical imaging,”
Molecular Imag., 6 237
–246
(2007). Google Scholar
S. Bloch,
F. Lesage,
L. McIntosh,
A. Gandjbakhche,
K. Liang, and
S. Achilefu,
“Whole-body fluorescence lifetime imaging of a tumor-targeted near-infrared molecular probe in mice,”
J. Biomed. Opt., 10 054003
(2005). https://doi.org/10.1117/1.2070148 1083-3668 Google Scholar
S. D. Konecky,
G. Y. Panasyuk,
K. Lee,
V. Markel,
A. G. Yodh, and
J. C. Schotland,
“Imaging complex structures with diffuse light,”
Opt. Express, 16 5048
–5060
(2008). https://doi.org/10.1364/OE.16.005048 1094-4087 Google Scholar
X. Intes,
“Time-domain optical mammography SoftScan: initial results,”
Acad. Radiol., 12 934
–947
(2005). https://doi.org/10.1016/j.acra.2005.05.006 1076-6332 Google Scholar
A. Koenig,
L. Hervé,
V. Josserand,
M. Berger,
J. Boutet,
A. Da Silva,
J. Dinten,
P. Peltié,
J. Coll, and
P. Rizo,
“In vivo mice lung tumor follow-up with fluorescence diffuse optical tomography,”
J. Biomed. Opt., 13 011008
(2008). https://doi.org/10.1117/1.2884505 1083-3668 Google Scholar
X. Intes and
B. Chance,
“Non-PET functional imaging techniques: optical,”
Radiol. Clin. North Am., 43 221
–234
(2005). https://doi.org/10.1016/j.rcl.2004.07.002 0033-8389 Google Scholar
J. R. Weber,
D. J. Cuccia, and
B. J. Tromberg,
“Modulated imaging in layered media,”
6674
–6676
(2006). Google Scholar
D. J. Cuccia,
F. Bevilacqua,
A. J. Durkin, and
B. J. Tromberg,
“Modulated imaging: quantitative analysis and tomography of turbid media in the spatial-frequency domain,”
Opt. Lett., 30 1354
–1356
(2005). https://doi.org/10.1364/OL.30.001354 0146-9592 Google Scholar
A. Bassi,
C. D’Andrea,
G. Valentini,
R. Cubeddu, and
S. Arridge,
“Temporal propagation of spatial information in turbid media,”
Opt. Lett., 33 2836
–2838
(2008). https://doi.org/10.1364/OL.33.002836 0146-9592 Google Scholar
S. D. Konecky,
A. Mazhar,
D. Cuccia,
A. J. Durkin,
J. C. Schotland, and
B. J. Tromberg,
“Quantitative optical tomography of sub-surface heterogeneities using spatially modulated structured light,”
Opt. Express, 17 14780
–14790
(2009). https://doi.org/10.1364/OE.17.014780 1094-4087 Google Scholar
S. R. Arridge,
“Optical tomography in medical imaging,”
Inverse Probl., 15 R41
–R93
(1999). https://doi.org/10.1088/0266-5611/15/2/022 0266-5611 Google Scholar
V. A. Markel,
V. Mital, and
J. C. Schotland,
“Inverse problem in optical diffusion tomography. III. Inversion formulas and singular-value decomposition,”
J. Opt. Soc. Am. A, 20 890
–902
(2003). https://doi.org/10.1364/JOSAA.20.000890 0740-3232 Google Scholar
V. A. Markel and
J. C. Schotland,
“Symmetries, inversion formulas, and image reconstruction for optical tomography,”
Phys. Rev. E, 70 056616
(2004). https://doi.org/10.1103/PhysRevE.70.056616 1063-651X Google Scholar
M. S. Patterson,
B. Chance, and
B. C. Wilson,
“Time-resolved reflectance and transmittance for the noninvasive measurement of tissue optical properties,”
Appl. Opt., 28 2331
–2336
(1989). https://doi.org/10.1364/AO.28.002331 0003-6935 Google Scholar
D. Boas,
J. Culver,
J. Stott, and
A. Dunn,
“Three-dimensional Monte Carlo code for photon migration through complex heterogeneous media including the adult human head,”
Opt. Express, 10 159
–170
(2002). 1094-4087 Google Scholar
J. Provost and
F. Lesage,
“The application of compressed sensing for photo-acoustic tomography,”
IEEE Trans. Med. Imaging, 28 585
–594
(2009). https://doi.org/10.1109/TMI.2008.2007825 0278-0062 Google Scholar
X. Intes,
J. Ripoll,
Y. Chen,
S. Nioka,
A. G. Yodh, and
B. Chance,
“In vivo continuous-wave optical breast imaging enhanced with Indocyanine Green,”
Med. Phys., 30 1039
–1047
(2003). https://doi.org/10.1118/1.1573791 0094-2405 Google Scholar
|