24 March 2023 Imaging effects due to pixel distortions in CdZnTe detectors—results from the HREXI Calibration Facility
Author Affiliations +
Abstract

ProtoEXIST2 (P2) was a prototype imaging x-ray detector plane developed for wide-field time-domain astrophysics (TDA) in the 5 to 200 keV energy band. It was composed of an 8 × 8 array of 5 mm thick, 2 cm × 2 cm pixelated (32 × 32) CdZnTe (CZT) detectors with a 0.6-mm pitch that utilize the NuSTAR ASIC (NuASIC) for readout. During the initial detector development process leading up to postflight examination of the entire detector plane, distortions in expected pixel positions and shapes were observed in a significant fraction of the detectors. The High Resolution Energetic x-ray Imager (HREXI) calibration facility (HCF) was designed and commissioned to improve upon these early experiments and to rapidly map out and characterize pixel nonuniformities and defects within CZT detector planes at resolutions down to 50 μm. Using this facility, the subpixel level detector response of P2 was measured at 100 μm5 resolution and analyzed to extract and evaluate the area and profile of individual pixels, their morphology across the entire P2 detector plane for comparison with previous measurements and to provide additional characterization. In this article, we evaluate the imaging performance of a coded-aperture telescope using the observed pixel morphology for P2 detectors. This investigation will serve as an initial guide for detector selection in the development of HREXI detector planes, for the future implementation of the 4pi X-Ray Imaging Observatory (4piXIO)6 mission, which aims to provide simultaneous and continuous imaging of the full sky (4π sr) in the 3 to 200 keV energy band with ≃2 arcmin angular resolution and ≃10 arcsec source localization, as well as other, future coded-aperture instruments.

© 2023 Society of Photo-Optical Instrumentation Engineers (SPIE)
Arkadip Basak, Branden E. Allen, Jae Sub E. Hong, Daniel P. Violette, and Jonathan E. Grindlay "Imaging effects due to pixel distortions in CdZnTe detectors—results from the HREXI Calibration Facility," Journal of Astronomical Telescopes, Instruments, and Systems 9(1), 018005 (24 March 2023). https://doi.org/10.1117/1.JATIS.9.1.018005
Received: 12 December 2022; Accepted: 3 March 2023; Published: 24 March 2023
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Image restoration

Telescopes

Signal to noise ratio

Calibration

Detector arrays

X-rays

Spatial resolution

RELATED CONTENT

Status of ART-XC/SRG instrument
Proceedings of SPIE (July 24 2014)
ART XC SRG status of the x ray focal...
Proceedings of SPIE (July 25 2014)
Ground Calibration of the IXPE Optics and Telescope
Proceedings of SPIE (January 01 1900)
ART-XC / SRG overview
Proceedings of SPIE (July 06 2018)
Rapid detection of surface defects by x-ray scanning
Proceedings of SPIE (September 26 2000)
Wide-field coded-mask x-ray cameras for SAX
Proceedings of SPIE (October 16 1995)
Development of the photon-counting 256ch CdTe line detector
Proceedings of SPIE (October 06 1999)

Back to Top