|
1.IntroductionAstro-H is a joint Japan Aerospace Exploration Agency (JAXA)/NASA x-ray satellite launched on February 17, 2016.1–4 The Astro-H has several instruments to cover a wide energy range from a few hundred eV to 600 keV. It is equipped with two soft x-ray telescopes (SXTs) covering up to about 15 keV.5,6 One of the two SXTs is for the soft x-ray spectroscopy (SXS: x-ray microcalorimeter detector) for high-resolution spectroscopy. The detector-limited field-of-view (FoV) of the SXS is , comprising . The other SXT is for the soft x-ray imaging (SXI: x-ray CCD detector) for general x-ray spectroscopy. The detector-limited FoV of the SXI is very wide and , comprising . The Astro-H SXTs are designed on the same basis as those built for the telescopes (XRT-Is) onboard Suzaku, launched in 2007.7 In order to achieve high throughput within a limited weight allowance, as many aluminum foils as possible, each of which is 101.6 mm long and is 0.15 to 0.31 mm thick, are nested within geometrical constraints. Each telescope consists of 203 nested shells of segmented thin-foil reflectors with a 200-nm gold, epoxy-replicated mirror surfaces, installed in primary and secondary mirror housings. The weight is , which is very light for the achieved effective area (EA). On the other hand, it is difficult to retain all the reflectors in the ideal shapes required by the original Wolter-I optics. We approximated both the primary and secondary reflectors by cones. The design parameters of the SXT-I and SXT-S are identical and summarized in Table 1. The system requirements are listed in Table 2. Each SXT is composed of the identically designed four “quadrants.” Each quadrant comprises primary and secondary housings, which contain reflectors. The reflectors in each housing are supported by seven alignment bars on the top and bottom sides. The precollimator is installed for stray-light protection, which consists of coaxially nested cylindrical aluminum blades placed above each reflector.8 A thermal shield is attached in front of the precollimator to stabilize the thermal environment of the SXTs.9 The shield is made of Al-coated thick polyimide film (PI) to ensure a large EA in the soft energy band. Table 1Design parameters of the Astro-H SXTs.
Table 2System requirement of the Astro-H SXTs.
Figure 1 shows the photos of the Astro-H SXT flight-models. The SXTs were developed at NASA/GSFC. The full details on the design concept, expected performance, fabrication, and interface to the satellite are described in Refs. 5 and 10. On ground, two measurement methods have been applied for the calibration of the telescope. One is the diverging beam illuminating the full aperture of the telescope (e.g., Ref. 11), where as the other is the raster scan with the pencil beam (e.g., Ref. 12). The full-aperture-illumination with the diverging beam with a long distance source such as the MSFC’s XRCF facility and the Panter facility well works to calibrate the image of a high-angular-resolution “arc sec” telescope, such as the Chandra HRMA (e.g., Ref. 13) and the XMM-Newton telescope (e.g., Ref. 11). The pencil-beam raster scan gives a precise solution for the EA calibration since the diverging angle of the beam is quite small, typically tens of arc secs are much smaller than the telescope vignetting. The EA taken with the raster scan method is known as well consistent with that obtained in orbit (e.g., Refs. 7 and 14). Also, the image with an “arc min” angular resolution can be made with the scan (e.g., Ref. 15). The initial ground calibration was performed at the 100-m beamline at NASA/GSFC, using a diverging beam, and the angular resolutions at 8.0 keV were determined to be and in a half-power diameter (HPD) for SXT-I and SXT-S, respectively.16 We later conducted the comprehensive ground calibration at the 30-m x-ray beam facility at the Institute of Space and Astronautical Science (ISAS)/JAXA, Japan. We adopted a raster scan method with a narrow x-ray pencil beam with a divergence of . In this paper, we report the summary of the ground calibration of the Astro-H SXTs at ISAS/JAXA. In Sec. 2, we describe the measurement system at ISAS. The results of the EA, image quality, and stray light are presented in Secs. 3, 4, and 5, respectively. We also compare the results with those of the Suzaku XRT-Is. 2.Measurement SetupFigure 2 shows a schematic view of the x-ray beam facility at ISAS/JAXA. The measurements are carried out by scanning the SXT aperture with a parallel narrow x-ray beam as follows. To archive high parallelism, we place an x-ray generator at one end of the system and collimate the beam with the slit placed 27 m away from the x-ray source. The resultant beam is highly parallel with the size of , and the beam divergence is square, on each side. To illuminate the entire aperture of the SXT by the fixed x-ray pencil beam, SXTs and detectors are moved synchronously. In addition, thin metal filters and double-crystal monochromator are installed in the beam path in order to obtain monochromatic x-rays. The same ISAS facility has been also used for the calibration of the x-ray telescopes of Suzaku. The measurement system for Astro-H is almost identical except for two minor changes; the chamber of the sample stage is larger and longer, and the image detector was renewed for the Astro-H calibration. For the focal plane detectors, charge-coupled device (CCD) and proportional counter (PC) are used. The FoV of the CCD is () at the distance of 5.6 m. The size of the PC is ( at 5.6 m). The CCD is used for investigating the image quality, whereas the PC is used for the EA. The details of the raster scan and beam facility are described in Refs. 1718.–19. The coordinate system is defined as in Fig. 2. The -axis is defined to align with the x-ray beam direction. The -axis is perpendicular to the ground, with the positive direction being upward. The -axis is on the horizontal plane, and so the three axes form a right-hand coordinate system. The EA and the image quality are measured with raster scan at each quadrant. The four quadrants in each telescope are named Q1, Q2, Q3, and Q4. In this paper, a quadrant coupled with the measured configuration of the coordinates is referred to as , where is the quadrant number of 1 to 4 and is either or for the axis; for example, means the Q1 with the positive -axis configuration. A dummy frame with the same shape of the thermal shield is installed at the measurement. The thermal shield is a forward-end component of the SXT that plays a role in the thermal shielding to the space. A structure of the frame obstruct a large fraction of the photons at the end sectors and less at the others. The dummy frame has no film or mesh. The mesh works as a neutral density filter in x-rays, whereas the film with the constant thickness uniformly absorbs incoming x-rays. The reduction of the EA due to the film or the mesh is not included in our measurements but was then modeled in our calculation. The errors quoted in this paper are at confidence level for one parameter of interest. 3.Effective AreaThe EA is given by where and are the scanning speed in the direction and the scanning pitch in the direction, respectively, is the photon count rate of the incident beam, and is the number of the collected photons from all the raster scan paths. In the typical setting, and . The beam size is , and the divergence of the beam is about 15". The aperture of the focal plane detector of the PC is a () circle. Roughly 3% of the reflected photons fall outside this diameter circle, irrespective of the energy (see the later section about the image quality calibration), a factor which is not corrected in the tables and figures presented in this section.3.1.Optical Axis DistributionFirst, we determined the optical axis of the SXTs for the x-ray calibration with Ti-K 4.5 keV. We measured the EA for all the quadrants at off-axis angles of , , 0, from the approximate optical axis of the whole telescope with optical parallel beams. The peaks fitted with the three-point data of and directions are defined as the optical axis on the respective coordinate axis. Figure 3 shows the results for the quadrant axis and the whole-telescope axis. The distribution of the measured optical axes among the quadrants is wide. The distribution in the SXT-S is narrower than that in the SXT-I. 3.2.On-Axis Effective AreaWe measured the EA of the quadrants of the SXT-I and SXT-S at the whole-telescope axis (on-axis) with seven energies (Al-K 1.5 keV, Ti-K 4.5 keV, 8.0 keV, 9.4 keV, 11.1 keV, 12.9 keV, and Mo-K 17.5 keV). Figure 4 and Table 3 summarize the results. For comparison, those of the Suzaku XRT-I are also included. The characteristics of the EA are found to vary significantly from quadrant to quadrant. For example, of SXT-I and for SXT-S have a relatively low EA, irrespective of x-ray energies. Table 3EA of all the quadrants measured at whole-telescope axis in unit of cm2.
We find that the EA is of the ideal value irrespective of the energy (Fig. 4). The ideal value is calculated with the design parameters and the reflectivity with the roughness of 4 Å. We adopted the Debye–Waller model for the reflectivity calculation and coded the formula by the Debye–Waller. During the calculation, we refer to the optical constant of the HENKE table. The total EA of the SXT-S is slightly larger than that of the SXT-I by . The loss of is not due to the test configuration on ground but due to deficiency of the telescope modeling. Misaki et al.20 discussed the cause of the deficiency for the ASTRO-E XRTs. The cause of the loss will be in part due to the nonperfect shape of each reflector in which some fraction of the reflected x-rays with a large angle due to the figure error hits the back of the inner reflector. The similar loss is also occurred with nonperfect positioning of the reflector. The loss of in the SXT EA can be qualitatively explained with a similar cause with the ASTRO-E XRTs. Compared with the XRT-Is of Suzaku, the EA of the SXTs is larger by factors of and at 1.5 and 8.0 keV, respectively. This improvement is mainly due to the scale up of the diameter from 40 to 45 cm and of the focal length from 4.75 to 5.6 m. Considering the EA of a telescope, the distribution of the quadrant axes is important. We measured the EA at each quadrant axis at 4.5 keV and show the results in Table 4. The sum of the EA of the quadrant is almost the same as the EA of the full telescope. This result implies that the optical axes of the four quadrants in SXT-I and SXT-S are well co-aligned. Table 4EA of the quadrants measured at each quadrant axis in unit of cm2 at the energy of 4.5 keV. The rows of “ratio QT/full” list the ratio of the EA at each quadrant axis to that at full-telescope axis.
3.3.Off-Axis Effective Area (SXT-I)The SXT-I is for the SXI detector, which has a large detector-limited FoV (). We measured the angular response at wide off-axis angles. First, we measured the EA at off-axis angles from to , scanning the full-telescope, at 4.5 keV. The measurement points of two directions ( and ), where the telescope is accordingly tilted to incident x-rays, are shown in Fig. 5. The data are fitted with a Lorentzian function with a full-width at half-maximum (FWHM) of . The residual in the best-fit results is found to be a few percent irrespective of the tilted direction. Next, we measured the EA at wide off-axis angles up to , scanning the quadrant at the multiple energies (1.5, 4.5, 8.0, 9.4, 11.1, 12.9, and 17.5 keV). Figure 6 shows the result. The data are found to be well fitted with the model of a Lorentzian function but not with a Gaussian model. The residual to the best-fit model is typically 5% at smaller off-axis angles and energies, whereas it increases toward larger off-axis angles or larger energies [Fig. 6(c)]. The SXT FoV is in FWHM at 1.5 keV, decreasing with higher x-ray energies, and is at 17.5 keV. The vignetting width in full width at 90% maximum of the EA is within and at 1.5 and 17.5 keV, respectively. Although the vignetting measurement was made only for one quadrant and one direction, the vignetting of the whole telescope must be in the similar shape because the tilt directions are intermediate between and , or between steeper and flatter angular responses. 3.4.Off-Axis Effective Area (SXT-S)We measured angular responses at a small off-axis angle at various tilt directions for the SXT-S. It is because the focal-plane detector SXS is so small that the FoV of the SXS system is limited by the detector size. First, we measured the EA at off-axis angles from to , scanning the full-telescope, at 4.5 keV. Figure 7(a) shows the results for the two tilt directions ( and ). We then fitted the data with a Lorentzian model and found the best-fit Lorentzian width to be in FWHM, which is the same value as that for the SXT-I (Fig. 5). The residual in fitting is . Next, we measured the EA at off-axis angle for four tilting points at energy bands of 1.5, 4.5, 8.0, 9.4, and 11.1 keV [Fig. 7(b)]. The residual to the best-fit model is typically a few percent, with an increasing trend toward larger energies. Finally, we measured the EA at and off-axis angles for eight tilting points at 4.5 keV [Fig. 7(c)]. The residual is . From these results, we conclude that the narrow vignetting in the SXS FoV has a small scattering of 1% to 2%, regardless of the tilt directions and x-ray energies. 4.Image QualityIn this section, we present the measured results of the imaging capability of the Astro-H SXT-I and SXT-S. 4.1.On-AxisThe detailed results are described in Refs. 21 and 22. To summarize, the obtained HPDs of the full-telescope in the 1.5 to 17.5 keV band are to and to for the SXT-I and SXT-S, respectively. The angular resolution is almost independent of the x-ray energy but is marginally worse at higher energies. The trend is especially notable for the quadrant of the SXT-I. More detailed measurement has been carried out with the spot scan in order to examine this trend further, as reported in Refs. 21 and 22. By contrast, the angular resolution of the XRT-Is of Suzaku was to (HPD). The significant improvement in the angular resolutions from Suzaku to Astro-H comes primarily from the improvement in precision in the shape of the reflecting surface and reflector positions.5 The image quality with the Suzaku XRT-Is showed a large scatter from telescope to telescope due to a gravitation effect.23 However, we find from the measurement of the rotation of the telescope of that the gravitation effect causes no change in the optical axes or the image quality of the Astro-H SXTs. This improvement may be due to the fact that the reflectors are glued to the alignment bars in the Astro-H SXTs, whereas they were not in the Suzaku SXTs. 4.2.X-Ray Scattering into the SXS FoVBecause of the narrow detector-limited FoV of the SXS and the moderate angular resolution of the SXT-S, a significant fraction of the photons collected by the SXT-S does not fall onto the SXS even when the target source is located at the center of the SXS FoV. Figure 8 shows a SXT-S/SXS image of a point-like source located at the center of the FoV of the SXS. We estimated the EA for the SXS and its ratio to the EA with the circular FoV of a diameter of (ISAS’s PC ), using the on-axis images of the SXT-S taken with the ISAS’s CCD, and list them in Table 5 at six energy bands. We find the ratio to be and to hardly depend on the energy (91% to 88% at 1.5 to 17.5 keV). Table 5On-axis EA for the SXS and its ratio to the circular FoV of ϕ7.4′ in diameter.
The fact that a significant fraction of the collected photons misses the SXS detector-limited FoV means that a significant amount of the flux from a source just outside the detector’s FoV. We estimated the amount of the contamination, from sources at three off-axis angles of , , and . For the off-axis image, the source is shifted from the on-axis in the perpendicular direction to the SXS, and for the and , the sources are shifted in the diagonal direction, as demonstrated in Fig. 9. In the cases of and off-axis sources at 4.5 keV, we placed the off-axis source at four different positions, rotated for 90 deg, and measured the EA for each configuration. Table 6 shows the resultant EAs with the off-axis sources. Table 6EA of SXS for sources outside its FoV at 1.5/4.5/8.0 keV.
The 4.5-keV EA due to sources outside the detector FoV is 7.4 to 10.5 and 2.2 to at off-axis angles of and , respectively, and hence to slightly depend on the off-axis angle. These EAs correspond to and of the on-axis EA, respectively. The ratio of the EA to that of on-axis sources does not depend on the energy with errors . The EA for a highly off-axis source at an off-axis angle of is of the on-axis EA. 4.3.Large Off-Axis in the SXI FoVThe SXI has a large detector-limited FoV of , and thus, the images at off-axis angles more than the half width of the SXI FoV should be investigated. Figure 10 shows images at off-axis angles between and , where the positive is defined as in the direction of (DET-X, DET-Y) = (, ). The EA and image quality (HPD) for the off-axis sources with the off-angles are tabulated in Table 7. We find that the image shrinks along the tilting direction and its opposite, the EA becomes smaller, and HPD becomes larger, as the off-axis angle increases, presumably because of the vignetting effect. Above of off-axis angle, a fan-shaped stray-light structure appears in the opposite side of the tilting direction to the image center. The surface brightness of the stray-light structure is of that of image center. The details are described in the next section. Table 7SXT-I EA (ϕ7.4′) and image quality (HPD) for off-axis sources.
5.Stray LightAlthough the FoV of the SXTs is in FWHM (Secs. 3.2 and 3.3), x-rays from outside of the FoV can reach the focal plane without experiencing the double reflection as they should be in the ideal Wolter-I type optics, because the thin-foil-nested mirrors used in the SXTs do not completely block the light paths of those kinds. The component makes a ghost image and is referred to as “stray light.” In order to cut off the stray light as much as possible, a “precollimator” is installed on the top of each SXT.8 The precollimator is composed of a thin and conical blade attached in front of the each primary reflector. The blade effectively blocks the light pass with an incident angle of about or larger on the plane (or radial) direction. The precollimator significantly reduces the light pass of the nondouble reflection in the mirror modules. In order to calibrate the stray lights, we, in advance, examined the light pass of the strays using the ray-tracing code. We then pick up three cases and made measurements as given as follows. The first and second is picked up for the SXT-I/SXI system while the third is for the SXT-S/SXS. 5.1.On-Axis in the SXI FoVFigure 11 is an image of an on-axis point-like source with the FoV of . The image is a mosaic of four maps taken with the CCD used in the ISAS beamline. The image clearly shows a ring-like stray-light structure with the radius of around the image center. This stray is constructed by the photons reflected by only the innermost secondary reflectors and is identified as the unique stray pass at on-axis. Its surface brightness is very low and of that of the image center. 5.2.Off-Axis in the SXI Detector-Limited FoVDue to the time limitation, we carried out the measurement at two representative off-axis angles of only. The stray light at the other off-axis angles is then deduced via ray-tracing simulation, which is calibrated with the off-axis angle data. First, we took images, irradiating x-rays on only one quadrant at off-axis angles of at 1.5 keV. In this measurement, the detector was placed at a distance of 3.733 m from the telescope ( of the focal length) in order to reduce the number of mosaic mapping, and images were taken at three or four offset positions from the telescope’s on-axis focus in order to cover the entire SXI FoV. Figures 12(a) and 12(b) show the resultant images at the off-axis angles of and , respectively. In the off-axis image, we identify some stray light components [e.g., the secondary-mirror single reflection, the direct component (no reflection), the backside-of-the-mirror reflection, and the precollimator reflection]. In the off-axis image, the backside-of-the-mirror reflection and precollimator reflection are marginally visible. We also estimate the EAs at the circular regions indicated in Fig. 12 (1S, 2S, 1B, and 2B), and list them in Table 8. Among them, the 2S region is found to be brightest with the EA of , which is of the on-axis EA of the full telescope. This region is dominated by the secondary-mirror single reflection component. Table 8EA of the SXT-I at an off-axis angle of ±30′ at 1.5 keV at several regions (see Fig. 12 for region IDs).
Next, we constructed an image, combining four images to cover the SXI FoV, each of which was illuminated for the entire telescope aperture at an off-axis angle of [Fig. 12(c)]. The illuminating x-ray beam is tilted toward one of the boundaries of the quadrants. The detector was placed at the focal point (5.6 m from the telescope). We find in the resultant image only the direct component within the FoV. The other stray-light components are absent in the image, presumably because there are no reflectors or precollimators at the quadrant boundary. 5.3.Strays in the SXS Detector-Limited FoVWe studied the stray-light structures, illuminating the entire aperture of the SXT-S. We used both the CCD and the PC of the beamline, placing them at the on-axis focus, and used Al-K (1.5 keV) and Ti-K (4.5 keV). Figure 13 summarizes the EAs at off-axis angles of , , , , , and , where, for the CCD results, we limited the photon integration region of the CCD to the same as the FoV of the PC. The results with the two kinds of detectors are found to be generally in good agreement. The EA of the SXS FoV at the off-axis angle of is at 1.5 keV. The EA of the stray light at large off-axis angles () is times smaller than the on-axis one (). This implies that we can observe targets almost stray-light free; the contribution of stray lights is smaller than with the SXS even if there is a contamination source that is 10 times brighter than the target at off-axis. For comparison, the EAs of stray lights for Suzaku are and within the PC and SXS FoVs, respectively, at any of the off-axis angles , , . at 1.5 keV. The stray-light level of the ASTRO-H SXTs from to is smaller than that of Suzaku XRT-Is by a factor of more than 10. The reason of the reduction of stray lights from Suzaku to ASTRO-H is due to the small improvement in the aluminum substrate of the reflectors. We understand that the residual component of stray light with Suzaku XRT-Is is mostly the backside-of-the-mirror reflection after the precollimator has been introduced. To reduce the backside-of-the-mirror reflection, the scratch direction of the reflector substrate to the x-ray incident has been changed from parallel (Suzaku) to perpendicular (ASTRO-H). The surface with perpendicular scratches makes the reflectivity decrease by several orders of magnitude from that with parallel scratches. This method was adopted in the precollimator substrate of Suzaku.24 Note that we changed the height of the precollimator from Suzaku to ASTRO-H. It is to adjust the difference of the detector-limited FoV. The detector-limited FoV is and square for the Suzaku XRT-Is and the ASTRO-H SXT-I. To obstruct the stray light pass to the far edge of the SXI detector, we need to increase the height of the SXT-I precollimator. 6.SummaryWe have presented our results of the ground calibration of the Astro-H SXTs at the ISAS/JAXA 30 m x-ray beam facility. The measurements were carried out by means of the raster scan method with a narrow x-ray pencil beam with a divergence of . We find that the EA, image quality, and stray-light contamination for both on-axis and off-axis sources meet requirements for both the SXT-I and the SXT-S. All the parameters with the SXT-S are slightly better than those with the SXT-I. We also have compared the performance of the SXTs with that of the Suzaku XRTs. The angular resolution of the SXTs ( to in HPD) was confirmed as significantly improved from the Suzaku XRTs (). The SXT-S stray lights in the SXS FoV are reduced by an order or more than the Suzaku XRT-S’. AcknowledgmentsThis work was supported by JSPS KAKENHI under Grant Nos. 25870744, 25105516, 23540280, 26800113, 15J10520, and 17K05403. ReferencesT. Takahashi et al.,
“The ASTRO-H mission,”
Proc. SPIE, 7732 77320Z
(2010). https://doi.org/10.1117/12.857875 PSISDG 0277-786X Google Scholar
T. Takahashi et al.,
“The ASTRO-H x-ray observatory,”
Proc. SPIE, 8443 84431Z
(2012). https://doi.org/10.1117/12.926190 PSISDG 0277-786X Google Scholar
T. Takahashi et al.,
“The ASTRO-H x-ray astronomy satellite,”
Proc. SPIE, 9144 914425
(2014). https://doi.org/10.1117/12.2055681 PSISDG 0277-786X Google Scholar
T. Takahashi et al.,
“The ASTRO-H (Hitomi) x-ray astronomy satellite,”
Proc. SPIE, 9905 99050U
(2016). https://doi.org/10.1117/12.2232379 PSISDG 0277-786X Google Scholar
Y. Soong et al.,
“ASTRO-H soft x-ray telescope (SXT),”
Proc. SPIE, 9144 914428
(2014). https://doi.org/10.1117/12.2056804 PSISDG 0277-786X Google Scholar
T. Okajima et al.,
“First peek of ASTRO-H soft x-ray telescope (SXT) in-orbit performance,”
Proc. SPIE, 9905 99050Z
(2016). https://doi.org/10.1117/12.2231705 PSISDG 0277-786X Google Scholar
P. J. Serlemitsos et al.,
“The x-ray telescope onboard Suzaku,”
Publ. Astron. Soc. Jpn., 59 S9
–S21
(2007). https://doi.org/10.1093/pasj/59.sp1.S9 Google Scholar
H. Mori et al.,
“The pre-collimator for the ASTRO-H x-ray telescopes: shielding from stray lights,”
Proc. SPIE, 8443 84435B
(2012). https://doi.org/10.1117/12.925902 PSISDG 0277-786X Google Scholar
Y. Tawara et al.,
“Development of ultra-thin thermal shield for ASTRO-H x-ray telescopes,”
Proc. SPIE, 8147 814704
(2011). https://doi.org/10.1117/12.893368 PSISDG 0277-786X Google Scholar
Y. Soong et al.,
“ASTRO-H soft x-ray telescope (SXT),”
Proc. SPIE, 8147 814702
(2011). https://doi.org/10.1117/12.894305 PSISDG 0277-786X Google Scholar
P. Gondoin et al.,
“Calibration of the first XXM flight mirror module: II. Effective area,”
Proc. SPIE, 3444 290
–301
(1998). https://doi.org/10.1117/12.331244 PSISDG 0277-786X Google Scholar
H. Kunieda et al.,
“Thirty-meter x-ray pencil beam line at the institute of space and astronautical science,”
Jpn. J. Appl. Phys., 32 4805
–4813
(1993). https://doi.org/10.1143/JJAP.32.4805 Google Scholar
D. E. Zissa,
“AXAF-1 high-resolution mirror assembly image model and comparison with x-ray ground-test image,”
Proc. SPIE, 3766 36
–50
(1999). https://doi.org/10.1117/12.363656 PSISDG 0277-786X Google Scholar
Y. Tsusaka et al.,
“Characterization of the advanced satellite for cosmology and astrophysics x-ray telescope: preflight calibration and ray tracing,”
Appl. Opt., 34 4848
–4856
(1995). https://doi.org/10.1364/AO.34.004848 APOPAI 0003-6935 Google Scholar
Y. Maeda et al.,
“Soft x-ray calibration for the NeXT x-ray telescope,”
Proc. SPIE, 7011 70112O
(2008). https://doi.org/10.1117/12.788652 PSISDG 0277-786X Google Scholar
T. Okajima et al.,
“First measurement of the ASTRO-H soft x-ray telescope performance,”
Proc. SPIE, 8443 844320
(2012). https://doi.org/10.1117/12.926443 PSISDG 0277-786X Google Scholar
T. Hayashi et al.,
“Upgrade of the 30-m x-ray pencil beam line at the Institute of Space and Astronautical Science,”
J. Astron. Telesc. Instrum. Syst., 1 044004
(2015). https://doi.org/10.1117/1.JATIS.1.4.044004 Google Scholar
K. Ichihara et al.,
“A ground calibration of the engineering model of the SXT onboard ASTRO-H using the ISAS 30m pencil beam facility,”
Proc. SPIE, 8443 844357
(2012). https://doi.org/10.1117/12.925860 PSISDG 0277-786X Google Scholar
T. Hayashi et al.,
“Upgrade of the thirty-meter x-ray pencil beam line at the Institute of Space and Astronautical Science,”
Proc. SPIE, 9144 91444A
(2014). https://doi.org/10.1117/12.2055741 PSISDG 0277-786X Google Scholar
K. Misaki et al.,
“x-ray telescope onboard Astro-E. III. Guidelines to performance improvements and optimization of the ray-tracing simulator,”
Appl. Opt., 44 916
–940
(2005). https://doi.org/10.1364/AO.44.000916 APOPAI 0003-6935 Google Scholar
T. Sato et al.,
“Revealing a detailed performance of the soft x-ray telescopes of the ASTRO-H mission,”
Proc. SPIE, 9144 914459
(2014). https://doi.org/10.1117/12.2055622 PSISDG 0277-786X Google Scholar
T. Sato et al.,
“Examining the angular resolution of the ASTRO-H’s soft x-ray telescopes,”
J. Astron. Telesc. Instrum. Syst., 2
(4), 044001
(2016). https://doi.org/10.1117/1.JATIS.2.4.044001 Google Scholar
A. Itoh et al.,
“Ground-based x-ray calibration of the Astro-E2 x-ray telescope: I. With pencil beam,”
Proc. SPIE, 5488 93
–102
(2004). https://doi.org/10.1117/12.551700 PSISDG 0277-786X Google Scholar
H. Mori et al.,
“Pre-collimator of the Astro-E2 x-ray telescopes for stray-light reduction,”
Publ. Astron. Soc. Jpn., 57 245
–257
(2005). https://doi.org/10.1093/pasj/57.1.245 Google Scholar
BiographyRyo Iizuka is a researcher of the Astro-H project team at the Institute of Space and Astronautical Science (ISAS) of Japan Aerospace Exploration Agency (JAXA). He received his MS and PhD degrees in astronomy from the University of Tokyo, Japan, in 2003 and 2006, respectively. His current main research interests are x-ray telescopes and x-ray polarimeters. He is not a member of SPIE. |