Paper
5 March 2015 Evaluation of the optical interference in a combined measurement system used for assessment of tissue blood flow
Zahra Abdollahi, Panayiotis A. Kyriacou, Justin P. Phillips
Author Affiliations +
Abstract
A dual-wavelength pulse oximetry system combined with laser Doppler was developed for the assessment of perfusion. Red and infrared PPG and Doppler signals were recorded from a healthy volunteer in three studies at different measurement sites to investigate the interference between PPG and laser Doppler flowmetry (LDF). Good quality photoplethysmographic (PPG) and Doppler signals were detected simultaneously using this combined probe from the skin of the finger. The influence of the PPG light sources on LDF measurements was investigated; also the influence of the LDF light sources to the PPG measurements was studied. In the worst case, the apparent change in PPG amplitude when the LDF system was switched on was less than 8%, and the change in LDF flux amplitude when the PPG system was switched on was 14.7%.
© (2015) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Zahra Abdollahi, Panayiotis A. Kyriacou, and Justin P. Phillips "Evaluation of the optical interference in a combined measurement system used for assessment of tissue blood flow", Proc. SPIE 9315, Design and Quality for Biomedical Technologies VIII, 93150R (5 March 2015); https://doi.org/10.1117/12.2079760
Lens.org Logo
CITATIONS
Cited by 1 patent.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Doppler effect

Blood circulation

Tissue optics

Light sources

Photoplethysmography

Signal detection

Sensors