Paper
26 September 2013 SmartCAM: an adaptive clinical SPECT camera
Christopher Dumas, Adam Bernstein, Alonzo Espinoza, Donovan Morgan, Kevin Lewis, Matt Nipper, Harrison H. Barrett, Matthew A. Kupinski, Lars R. Furenlid
Author Affiliations +
Abstract
An adaptive pinhole aperture that fits a GE MaxiCam Single-Photon-Emission Computed Tomography (SPECT) system has been designed, built, and is undergoing testing. The purpose of an adaptive aperture is to allow the imaging system to make adjustments to the aperture while imaging data are being acquired. Our adaptive pinhole aperture can alter several imaging parameters, including field of view, resolution, sensitivity, and magnification. The dynamic nature of such an aperture allows for imaging of specific regions of interest based on initial measurements of the patient. Ideally, this mode of data collection will improve the understanding of a patient’s condition, and will facilitate better diagnosis and treatment. The aperture was constructed using aluminum and a low melting point, high-stopping-power metal alloy called Cerrobend. The aperture utilizes a rotating disk for the selection of a pinhole configuration; as the aluminum disk rotates, different pinholes move into view of the camera face and allow the passage of gamma rays through that particular pinhole. By controlling the angular position of the disk, the optical characteristics of the aperture can be modified, allowing the system to acquire data from controlled regions of interest. First testing was performed with a small radioactive source to prove the functionality of the aperture.
© (2013) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Christopher Dumas, Adam Bernstein, Alonzo Espinoza, Donovan Morgan, Kevin Lewis, Matt Nipper, Harrison H. Barrett, Matthew A. Kupinski, and Lars R. Furenlid "SmartCAM: an adaptive clinical SPECT camera", Proc. SPIE 8853, Medical Applications of Radiation Detectors III, 885307 (26 September 2013); https://doi.org/10.1117/12.2030090
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Cameras

Imaging systems

Single photon emission computed tomography

Aluminum

Data acquisition

Gamma radiation

Image resolution

RELATED CONTENT

Molecular imaging in the College of Optical Sciences an...
Proceedings of SPIE (October 20 2014)
Neutron computed tomography a tool for the testing of...
Proceedings of SPIE (February 27 1997)
A prototype instrument for adaptive SPECT imaging
Proceedings of SPIE (March 28 2007)
Optical moire leveraging analysis
Proceedings of SPIE (January 06 1994)

Back to Top