2 March 2012 Optimization of continuous tube motion and step-and-shoot motion in digital breast tomosynthesis systems with patient motion
Raymond J. Acciavatti, Andrew D. A. Maidment
Author Affiliations +
Abstract
In digital breast tomosynthesis (DBT), a reconstruction of the breast is generated from projections acquired over a limited range of x-ray tube angles. There are two principal schemes for acquiring projections, continuous tube motion and step-and-shoot motion. Although continuous tube motion has the benefit of reducing patient motion by lowering scan time, it has the drawback of introducing blurring artifacts due to focal spot motion. The purpose of this work is to determine the optimal scan time which minimizes this trade-off. To this end, the filtered backprojection reconstruction of a sinusoidal input is calculated. At various frequencies, the optimal scan time is determined by the value which maximizes the modulation of the reconstruction. Although prior authors have studied the dependency of the modulation on focal spot motion, this work is unique in also modeling patient motion. It is shown that because continuous tube motion and patient motion have competing influences on whether scan time should be long or short, the modulation is maximized by an intermediate scan time. This optimal scan time decreases with object velocity and increases with exposure time. To optimize step-and-shoot motion, we calculate the scan time for which the modulation attains the maximum value achievable in a comparable system with continuous tube motion. This scan time provides a threshold below which the benefits of step-and-shoot motion are justified. In conclusion, this work optimizes scan time in DBT systems with patient motion and either continuous tube motion or step-and-shoot motion by maximizing the modulation of the reconstruction.
© (2012) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Raymond J. Acciavatti and Andrew D. A. Maidment "Optimization of continuous tube motion and step-and-shoot motion in digital breast tomosynthesis systems with patient motion", Proc. SPIE 8313, Medical Imaging 2012: Physics of Medical Imaging, 831306 (2 March 2012); https://doi.org/10.1117/12.911016
Lens.org Logo
CITATIONS
Cited by 12 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Modulation

Sensors

X-rays

Digital breast tomosynthesis

Modulation transfer functions

Signal attenuation

Breast

Back to Top