11 February 2011 Endoscopic device for functional imaging of the retina
Simon Barriga, Sweyta Lohani, Bret Martell, Peter Soliz, Dan Ts'o
Author Affiliations +
Proceedings Volume 7885, Ophthalmic Technologies XXI; 78851S (2011) https://doi.org/10.1117/12.878797
Event: SPIE BiOS, 2011, San Francisco, California, United States
Abstract
Non-invasive imaging of retinal function based on the recording of spatially distributed reflectance changes evoked by visual stimuli has to-date been performed primarily using modified commercial fundus cameras. We have constructed a prototype retinal functional imager, using a commercial endoscope (Storz) for the frontend optics, and a low-cost back-end that includes the needed dichroic beam splitter to separate the stimulus path from the imaging path. This device has been tested to demonstrate its performance for the delivery of adequate near infrared (NIR) illumination, intensity of the visual stimulus and reflectance return in the imaging path. The current device was found to be capable of imaging reflectance changes of 0.1%, similar to that observable using the modified commercial fundus camera approach. The visual stimulus (a 505nm spot of 0.5secs) was used with an interrogation illumination of 780nm, and a sequence of imaged captured. At each pixel, the imaged signal was subtracted and normalized by the baseline reflectance, so that the measurement was ΔR/R. The typical retinal activity signal observed had a ΔR/R of 0.3-1.0%. The noise levels were measured when no stimulus was applied and found to vary between ± 0.05%. Functional imaging has been suggested as a means to provide objective information on retina function that may be a preclinical indicator of ocular diseases, such as age-related macular degeneration (AMD), glaucoma, and diabetic retinopathy. The endoscopic approach promises to yield a significantly more economical retinal functional imaging device that would be clinically important.
© (2011) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Simon Barriga, Sweyta Lohani, Bret Martell, Peter Soliz, and Dan Ts'o "Endoscopic device for functional imaging of the retina", Proc. SPIE 7885, Ophthalmic Technologies XXI, 78851S (11 February 2011); https://doi.org/10.1117/12.878797
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Retina

Reflectivity

Imaging systems

Functional imaging

Visualization

Imaging devices

Near infrared

Back to Top