16 September 2005 Stability of high-pressure xenon gamma-ray spectrometers
Author Affiliations +
Abstract
The stability of high-pressure xenon gamma-ray specrometers has been measured at Mirmar Sensor. Stability was measured by tracking the 137Cs peak of the entire spectroscopy system including detector, preamplifier, amplifier, and multichannle analyzer over 6 days in a laboratory setting. Overall system gain drift is about 0.016%. Resolution drifts are about 0.15%. The overall energy resolution is between 1.7 and 1.8% at 662 keV.
© (2005) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Albert Beyerle "Stability of high-pressure xenon gamma-ray spectrometers", Proc. SPIE 5922, Hard X-Ray and Gamma-Ray Detector Physics VII, 59220C (16 September 2005); https://doi.org/10.1117/12.622628