Paper
29 September 2004 Testing of mid-infrared detector arrays for FORCAST
Joseph D. Adams, Terry L. Herter, Luke D. Keller, George E. Gull, Bruce Pirger, Justin Schoenwald, Marc Berthoud
Author Affiliations +
Abstract
This paper presents results on performance testing of mid-infrared detector arrays for the Faint Object Infrared Camera for the SOFIA Telescope (FORCAST). FORCAST is a two-channel camera that utilizes a Si:As blocked impurity band (BIB) 256 x 256 detector array for imaging through discrete filters at 5 - 25 microns, and a Si:Sb BIB 256 x 256 detector array for imaging at 25 - 40 microns, over a 3.2' x 3.2' field of view, under high thermal background conditions. DRS Technologies has designed and fabricated several Si:As BIB and Si:Sb BIB engineering grade detector arrays which we test as candidate arrays for FORCAST. We present their initial laboratory test performance results.
© (2004) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Joseph D. Adams, Terry L. Herter, Luke D. Keller, George E. Gull, Bruce Pirger, Justin Schoenwald, and Marc Berthoud "Testing of mid-infrared detector arrays for FORCAST", Proc. SPIE 5499, Optical and Infrared Detectors for Astronomy, (29 September 2004); https://doi.org/10.1117/12.551490
Lens.org Logo
CITATIONS
Cited by 7 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Detector arrays

Electronics

Multiplexers

Cameras

Telescopes

Infrared sensors

RELATED CONTENT

IRC: an infrared camera on board the IRIS
Proceedings of SPIE (August 21 1998)
MAGIC: a new near-infrared camera for Calar Alto
Proceedings of SPIE (October 20 1993)
Mid-infrared camera without lens (MIRACLE) for SPICA
Proceedings of SPIE (August 05 2010)
MANIAC a new mid and near infrared array...
Proceedings of SPIE (October 23 1997)

Back to Top