Abstract
The present paper deals with theoretical basis of coulometric method used for measurement of chloride ions concentration in water solutions. Static transfer equations were derivated for generator electrodes circuit, as well as for equation describing the kinetics of phenomena taking place in detector electrodes circuit. Basing on assumptions resulting from mathematical models, chloride ions concentration meter was designed and built. This model was used for experimental verification of derivated transfer equations. Results of performed measurements and their analysis confirmed presented theoretical considerations.
© (2001) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Witold Holiczer "Coulometric method of the chloride ion concentration measurement: mathematical model of the measuring transducer and its experimental verification", Proc. SPIE 4516, Optoelectronic and Electronic Sensors IV, (8 August 2001); https://doi.org/10.1117/12.435917
Lens.org Logo
CITATIONS
Cited by 1 scholarly publication.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Ions

Electrodes

Sensors

Silver

Mathematical modeling

Transducers

Signal detection

Back to Top