14 April 2000 Comparison of two methods for evaluating image quality of chest radiographs
C. Herrmann, P Sund, A Tingberg, S. Keddache, L.G. Mansson, A. Almen, S. Mattsson
Author Affiliations +
Abstract
The Imix radiography system (Oy Imix Ab, Finland) consists of an intensifying screen, optics, and a CCD camera. An upgrade of this system (Imix 2000) with a red-emitting screen and new optics has recently been released. The image quality of Imix (original version), Imix 2000, and two storage-phosphor systems, Fuji FCR 9501 and Agfa ADC70 was evaluated in physical terms (DQE) and with visual grading of the visibility of anatomical structures in clinical images (141 kV). PA chest images of 50 healthy volunteers were evaluated by experienced radiologists. All images were evaluated on Siemens Simomed monitors, using the European Quality Criteria. The maximum DQE values for Imix, Imix 2000 Agfa and Fuji were 11%, 14%, 17% and 19%, respectively (141 kV, 5 (mu) Gy). Using the visual grading, the observers rated the systems in the following descending order: Fuji, Imix 2000, Agfa, and Imix. Thus, the upgrade to Imix 2000 resulted in higher DQE values and a significant improvement in clinical image quality. The visual grading agrees reasonably well with the DQE results; however, Imix 2000 received a better score than what could be expected from the DQE measurements.
© (2000) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
C. Herrmann, P Sund, A Tingberg, S. Keddache, L.G. Mansson, A. Almen, and S. Mattsson "Comparison of two methods for evaluating image quality of chest radiographs", Proc. SPIE 3981, Medical Imaging 2000: Image Perception and Performance, (14 April 2000); https://doi.org/10.1117/12.383116
Lens.org Logo
CITATIONS
Cited by 38 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Image quality

Visualization

Signal to noise ratio

Sensors

Modulation transfer functions

Chest imaging

Chest

Back to Top