22 October 1999 Performance of the x-ray facility for calibration of the JEM-X instrument aboard the INTEGRAL satellite
Carlo Pelliciari, Vittore Carassiti, M. De Paoli Vitali, Federico Evangelisti, Filippo Frontera, Giovanni Pareschi
Author Affiliations +
Abstract
We report preliminary results on the performance of the x- ray apparatus built to calibrate the x-ray monitor JEM-X aboard the INTEGRAL satellite for gamma-ray astronomy. JEM-X is based on a position sensitive xenon detector operative from 3 to 60 keV. It will make use of coded mask to get imaging capabilities. The x-ray apparatus allows to scan the x-ray detector with a monochromatic beam obtained with the use of a double crystal diffractometer. The current energy range of the beam can be chosen from about 10 keV to 120 keV while its direction is fixed independently of the photon energy.
© (1999) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Carlo Pelliciari, Vittore Carassiti, M. De Paoli Vitali, Federico Evangelisti, Filippo Frontera, and Giovanni Pareschi "Performance of the x-ray facility for calibration of the JEM-X instrument aboard the INTEGRAL satellite", Proc. SPIE 3765, EUV, X-Ray, and Gamma-Ray Instrumentation for Astronomy X, (22 October 1999); https://doi.org/10.1117/12.366567