Paper
18 October 1996 CdZnTe strip detectors for astrophysical arc second imaging and spectroscopy: detector performance and radiation effects
Lyle M. Bartlett, Carl Michael Stahle, David M. Palmer, Louis M. Barbier, Scott D. Barthelmy, F. B. Birsa, Neil A. Gehrels, John F. Krizmanic, Peter Kurczynski, James L. Odom, Ann M. Parsons, Carol Sappington, Peter K. Shu, Bonnard J. Teegarden, Jack Tueller
Author Affiliations +
Abstract
CdZnTe strip detectors have been fabricated and tested to show the ability for arc second imaging and spectroscopy. Two dimensional CdZnTe strip detectors with 100 micron pitch have been fabricated and wire bonded to readout electronics to demonstrate the ability to localize 22 to 122 keV photons to less than 100 microns. Good spectral resolution has also been achieved. The uniformity and relative efficiency of the strip detector are discussed. Radiation damage effects by intermediate energy neutrons and low energy protons on the surface and bulk performance of CdZnTe devices have been investigated and are presented. Activation and annealing of radiation effects have been seen and are discussed.
© (1996) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Lyle M. Bartlett, Carl Michael Stahle, David M. Palmer, Louis M. Barbier, Scott D. Barthelmy, F. B. Birsa, Neil A. Gehrels, John F. Krizmanic, Peter Kurczynski, James L. Odom, Ann M. Parsons, Carol Sappington, Peter K. Shu, Bonnard J. Teegarden, and Jack Tueller "CdZnTe strip detectors for astrophysical arc second imaging and spectroscopy: detector performance and radiation effects", Proc. SPIE 2806, Gamma-Ray and Cosmic-Ray Detectors, Techniques, and Missions, (18 October 1996); https://doi.org/10.1117/12.254023
Lens.org Logo
CITATIONS
Cited by 12 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Sensors

Imaging spectroscopy

Electronics

Radiation effects

Gamma radiation

Spectroscopy

Sensor performance

RELATED CONTENT

High-dynamic-range MCP structures
Proceedings of SPIE (October 01 1991)
Microsatellite gamma-ray spectroscopy experiment
Proceedings of SPIE (October 01 1999)
Performance of RENA 3 IC with position sensitive solid state...
Proceedings of SPIE (September 04 2008)
Novel x ray silicon detector for 2D imaging and high...
Proceedings of SPIE (October 19 1999)

Back to Top