Abstract
An AGEMA 870 thermal imaging system has been used to measure the temperature of a film cooled panel in a hot duct, ventilated by flowing air at atmospheric pressure. To avoid detecting radiation from these gases, it proved necessary to view the panel through a narrow bandpass filter centred on 3.95im. The thermography results failed to agree with spot thermocouple measurements until changes were made to both the experimental and calculation procedures. The paper describes these changes and the calculation algorithms used to calibrate the system, account for non-uniform ambient radiation and eliminate the distortion produced by the wide angle lens.
© (1990) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
T. J. Meddings and Christopher W. Wilson "Application of a commercially available thermal imager to a film-cooled panel temperature measurement", Proc. SPIE 1320, Infrared Technology and Applications, (1 October 1990); https://doi.org/10.1117/12.22331