4 April 2022 Quantitative assessment of motion effects in dual-source dual energy CT and dual-source photon-counting detector CT
Author Affiliations +
Abstract
Dual-source CT can provide high temporal resolution single energy imaging, but this is sacrificed in dual-energy mode (DS-DE). The current study evaluates a dual-source photon-counting-detector (DS-PCD) CT capable of high temporal resolution (66 ms) dual-energy imaging. A rod simulating coronary artery motion was scanned on both DS-DE and DS-PCD CT scanners and image quality was evaluated. Spatial registration—quantified by dice coefficient—between the high and low energy images was better with DS-PCD than DS-DE. Furthermore, the rod had a more circular appearance and its diameter was more accurate in iodine maps generated by DS-PCD whereas DS-DE suffered from motion artifacts.
Conference Presentation
© (2022) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Zaki Ahmed, Kishore Rajendran, Hao Gong, Cynthia McCollough, and Shuai Leng "Quantitative assessment of motion effects in dual-source dual energy CT and dual-source photon-counting detector CT", Proc. SPIE 12031, Medical Imaging 2022: Physics of Medical Imaging, 120311P (4 April 2022); https://doi.org/10.1117/12.2611030