Paper
18 November 1989 Calibration Of 10-Channel Grating Polychromator
K. Kawahata, M. Sakamoto, J. Fujita, H. Matsuo, K. Sakai
Author Affiliations +
Proceedings Volume 1039, 13th Intl Conf on Infrared and Millimeter Waves; (1989) https://doi.org/10.1117/12.978382
Event: 13th International Conference on Infrared and Millimeter Waves, 1987, Honolulu, HI, United States
Abstract
A high temperature (T = 500°C) radiation source has been developed for the calibration of electron cyclotron emission measuring system on the JIPP T-IIU tokamak. The source has a heated area of (1)150 mm, and has the same emissivity as that of the microwave absorber Eccosorbe. The grating polychromator has been calibrated using the radiation source developed. The obtained electron temperatures agree with those by Thomson scattering within 10%.
© (1989) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
K. Kawahata, M. Sakamoto, J. Fujita, H. Matsuo, and K. Sakai "Calibration Of 10-Channel Grating Polychromator", Proc. SPIE 1039, 13th Intl Conf on Infrared and Millimeter Waves, (18 November 1989); https://doi.org/10.1117/12.978382
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Calibration

Electrochemical etching

Temperature metrology

Microwave radiation

Thomson scattering

Absorption

Plasma

Back to Top