Spectral sensitivity of a soft x-ray streak camera (hv=0.1-1.5 keV) was absolutely calibrated by using x-ray diodes and x-ray film; coupled with a transmission grating spectrometer. Distortion of the output image due to time and space nonlineality of the streak camera system was corrected by computer aided processing. A new analyzing data-reduction method of the x-ray spectum from the multifiltered XRD signals is also described.
M. Nakai,K. Kondou,T. Urasaki,K. Sakurai,H. Nishimura,Y. Izawa, andC. Yamanaka
"Absolute Calibration of a Soft X-Ray Streak Camera", Proc. SPIE 0689, X-Ray Calibration: Techniques, Sources, and Detectors, (12 August 1986); https://doi.org/10.1117/12.936577