Paper
26 December 1979 Large Area X-Ray Image Receptor With Electrophoretic Display
J . D. Kingsley, A . R . Sears, K. H. Yang
Author Affiliations +
Abstract
The feasibility of an x-ray imaging device based on photoconductive control of an electrophoretic display has been explored. Contrast ratios of 18 to 1 and resolutions of 8 line pairs/mm have been measured. The radiation to reflectance transfer characteristic is non-linear, with substantially less sensitivity to small exposures. Radiation exposures of 2.4.mR have produced satisfactory x-ray images. Excellent quality images have been obtained but cell stability and linearity need to be improved.
© (1979) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
J . D. Kingsley, A . R . Sears, and K. H. Yang "Large Area X-Ray Image Receptor With Electrophoretic Display", Proc. SPIE 0206, Recent and Future Developments in Medical Imaging II, (26 December 1979); https://doi.org/10.1117/12.958196
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Electrodes

Particles

Photoresistors

Selenium

Reflectivity

Imaging systems

X-rays

Back to Top