15 July 2015 Development of a movable diffuse reflectance spectroscopy system for clinical study of esophageal precancer
Author Affiliations +
Abstract
We constructed a movable imaging spectrograph-based system and a contact probe consisting of fibers with several source-to-detection separations (SDS) to measure spatially-resolved diffuse reflectance spectra from superficial tissue. The goal is to estimate optical properties of the mucosa in the oral cavity and investigate correlations between the estimated properties and precancerous changes in the mucosa. A previously developed GPU-based iterative curve-fitting inverse Monte Carlo model was used to extract optical properties from measured spectra. We validated the system with two-layer tissue phantoms, took in-vivo measurements on the buccal mucosa of three normal volunteers, and extracted optical parameters.
© (2015) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Yi-Hsien Hsiao, Gen-Hao Tien, Min-Jie Chuang, Fang-Wei Hsu, Hong-Po Hsieh, and Kung-Bin Sung "Development of a movable diffuse reflectance spectroscopy system for clinical study of esophageal precancer", Proc. SPIE 9537, Clinical and Biomedical Spectroscopy and Imaging IV, 95371Q (15 July 2015); https://doi.org/10.1117/12.2183737
Lens.org Logo
CITATIONS
Cited by 2 scholarly publications.
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Calibration

Tissue optics

Imaging systems

Monte Carlo methods

In vivo imaging

Diffuse reflectance spectroscopy

Reflectivity

Back to Top