Paper
9 July 1999 Design of an autostereoscopic endoscope for abdominal surgery
Christopher W. Connor, Neil A. Dodgson, John E. Moore, Tim D. Wilkinson, Adrian R. L. Travis
Author Affiliations +
Proceedings Volume 3595, Biomedical Diagnostic, Guidance, and Surgical-Assist Systems; (1999) https://doi.org/10.1117/12.351526
Event: BiOS '99 International Biomedical Optics Symposium, 1999, San Jose, CA, United States
Abstract
An imaging system has been demonstrated which captures a 3D scene with six parallax images using only one stationary CCD. The parallax effect is achieved either by imaging through a pinhole which is translated in front of the CCD or by imaging through a solid-state ferroelectric shutter which simulates a translatable pinhole. A single set of six compatible images is referred to as a single autostereoscopic image; the images so captured are presented on an autostereoscopic display. An autostereoscopic display system provides the perception of depth by presenting several images on the display over a range of viewing angles. When viewed at the correct distance from the display, a different image is perceived by each eye. With suitable images, this results in the illusion of depth without using stereoscopic eyewear. Improvements in the imaging/display rate and miniaturization of the camera assembly will produce a system compatible with use in laparoscopic surgery.
© (1999) COPYRIGHT Society of Photo-Optical Instrumentation Engineers (SPIE). Downloading of the abstract is permitted for personal use only.
Christopher W. Connor, Neil A. Dodgson, John E. Moore, Tim D. Wilkinson, and Adrian R. L. Travis "Design of an autostereoscopic endoscope for abdominal surgery", Proc. SPIE 3595, Biomedical Diagnostic, Guidance, and Surgical-Assist Systems, (9 July 1999); https://doi.org/10.1117/12.351526
Advertisement
Advertisement
RIGHTS & PERMISSIONS
Get copyright permission  Get copyright permission on Copyright Marketplace
KEYWORDS
Charge-coupled devices

Camera shutters

Surgery

Image segmentation

Autostereoscopic displays

LCDs

Laparoscopy

RELATED CONTENT

Dense 3D organ modeling from a laparoscopic video
Proceedings of SPIE (April 20 2021)
Video stereo-laparoscopy system
Proceedings of SPIE (February 02 2006)